Nephrology Nursing · Role subtrack · Module 9 of 12

Performance Testing and Acceptance

Total cell volume and its baseline, the integrity test, inspection, independent acceptance gates, and repeats without gaming.

PublishedNailathalaGipatikPepalwal: ReferencesMga SanggunianMga TinubdanReng Reperensya: 3 Audience: Reprocessing operators and dialysis technicians in Philippine hemodialysis units · biomedical technicians who service reprocessors · head nurses and medical directors who supervise reuse programs Read timeOras ng pagbasaOras sa pagbasaOras ning pamamasa:
A reprocessor screen showing a total cell volume result and a pressure-hold test curve, with a technician pointing at the result and a dialyzer mounted beside it.
Renal Care Academy›Nephrology Nursing Track›By Role›Dialyzer Reprocessing Technician›Module 9 of 12 Nephrology Nursing subtrack for the dialyzer reuse room
♻️

Nephrology Nursing subtrack for the dialyzer reuse room

This module is study material for dialyzer reprocessing. It explains the reasoning behind each step; it is not a procedure. Concentrations, contact times, storage limits and test thresholds come from the germicide and dialyzer instructions for use and your unit’s validated procedure, never from this page. The medical director decides who is eligible for reuse; the nurse decides whether a dialyzer is used on a patient. Every case, name and identifier is fictional. Pending clinical review.

After this module you should be able to

Classroom 6 hSkills lab 4 hSupervised placement 48 h
Pre-testBefore you read, answer the 5 self-assessment questions and note your score. After the module, select New set of questions and compare. The difference is what you learned; the questions you still miss are what to reread.

The big picture, in plain words

In plain wordsBuying a used car, you check that the engine runs, that the brakes work, and that the papers are in order. A strong engine does not make up for brakes that fail. Every check must pass on its own.

The picture has a limit. A buyer can choose to accept a small fault. A technician cannot trade one result against another, because each test looks at a different danger. The volume test asks whether enough fibers are open to deliver the dose. The integrity test asks whether any fiber is broken, which would let blood and dialysis fluid mix. A dialyzer with perfect volume and a broken fiber is a dangerous dialyzer.

Figure 1 · Key terms for this module

Total cell volume (TCV)The volume of the blood compartment, measured to estimate how many fibers are open.
BaselineThe TCV measured on this dialyzer before its first use, under the program’s method.
Retained TCVMeasured TCV as a percentage of baseline.
Integrity testA pressure-hold test of the membrane: the blood side is pressurized and any leak shows as a fall in pressure.
Acceptance gateA condition that must pass on its own before release.
DispositionThe recorded decision for a dialyzer: release, reprocess, or discard.

Your job is to measure honestly, apply every gate independently, and let one failure stop the dialyzer.

Case: Seventy-nine and eighty

It is 4:40 on a Friday afternoon at a unit in Davao. Ms. Grace Alcantara is finishing the day’s run. The unit floor for retained TCV is 80%, from the Philippine Health Insurance Corporation (PhilHealth) quality circular and the program.1

Dialyzer RU-0731 has a baseline of 100 mL and a measured volume of 79 mL. Dialyzer RU-0732 has a baseline of 100 mL and a measured volume of 86 mL, but its pressure-hold test failed. Dialyzer RU-0733 shows a measured volume of 104 mL against a baseline of 98 mL. A colleague suggests running RU-0731 again “until it passes,” and says RU-0732 is “fine, the volume is great.”

Before reading on, decide:

Keep your answers in mind. The lesson gives you what you need, and the case review near the end shows one sound response.

Measuring against a defensible baseline

A dialyzer on a reprocessor with fluid displaced from the blood compartment into a measuring scale, showing baseline 100 milliliters and measured 82 milliliters, retained 82 percent.
Figure 2. Total cell volume is measured against this dialyzer’s own baseline, not a catalog number.

TCV is measured by displacing the fluid in the blood compartment and measuring its volume. Each dialyzer’s first measurement, before first use, is its baseline. Every later result is compared with that baseline, never with a catalog value, because individual dialyzers of the same model differ.

Retained TCV (%) = measured TCV ÷ baseline TCV × 100  ·  Example for training only: 82 mL ÷ 100 mL × 100 = 82%.

The PhilHealth quality circular and the Association for the Advancement of Medical Instrumentation (AAMI) standard use a floor of at least 80% retained.1,2 A dialyzer at 80% has lost a fifth of its blood compartment. It meets the floor, and it may still clean less than it did when new (Module 2).

SituationWhat it meansAction
Baseline missingNo denominator; the result means nothingHold; never substitute a catalog value
Baseline implausible for the modelPossible measurement or recording errorHold and investigate
Measured above baselinePossible measurement error, air, or record mismatchFlag for review; do not cap or accept automatically
Measured below the floorToo many fibers closedFails this gate; follow the procedure for discard
RememberCompare every result with this dialyzer’s own baseline. A missing baseline is a hold, not a guess.

The integrity test

In a pressure-hold test, the reprocessor pressurizes the blood compartment, usually with air, and watches the pressure. An intact membrane holds it. A broken fiber lets gas escape to the other side, and the pressure falls faster than the allowed rate. The pass criterion comes from the reprocessor and the dialyzer model, as validated in your procedure.

A broken fiber is dangerous in both directions. Dialysis fluid and anything in it can enter the blood, and blood can leak into the dialysis fluid. Volume and integrity measure different things, so one can be perfect while the other fails. In a Centers for Disease Control and Prevention (CDC) study of three centers whose reprocessed dialyzers were all integrity-tested before use, the rate of pyrogenic reactions was 0.7 per 1,000 treatments despite high bacterial counts in the dialysis fluid.3 The study did not test the integrity check itself, but it shows the setting in which membrane integrity protects patients.

Stop and reportA failed integrity test stops the dialyzer, whatever its volume. Never override it, never repeat it until it passes without the procedure allowing it.

Inspecting the processed dialyzer

Look forWhy it rejects the dialyzer
Cracks or leaks in the housing or portsLoss of containment; contamination
Visible clot, dark streaks or discolored fibers more than the procedure allowsBlocked surface; trapped residue
Blood or protein in the headers after cleaningCleaning incomplete
Missing or damaged caps; damaged O-ringsGermicide can leak out and contaminants can enter
Label damaged or unreadableIdentity lost (Module 7)

Inspect in good light, at a set point in the procedure, and record what you saw. “Looked fine” written at the end of the day is not an inspection.

Every gate must pass

A row of padlocks labeled identity, eligibility, uses left, inspection, volume, integrity, processing record, germicide, storage time and residual test; one red padlock keeps the door to release closed.
Figure 3. Release needs every gate. One failed gate keeps the door shut no matter how many others pass.

Before a processed dialyzer can be stored for release, every acceptance gate must pass. Each is independent. A favorable result on one cannot compensate for a failure on another.

GateQuestionIf it fails or is missing
IdentityIs this the labeled patient’s dialyzer, with a matching record?Quarantine
EligibilityIs the patient currently eligible?Hold
Uses leftIs the next use within the limit?Discard at the limit
InspectionAny rejecting finding?Discard or reprocess per procedure
VolumeRetained TCV at or above the floor, against a valid baseline?Discard below the floor; hold if baseline invalid
IntegrityPressure held within the criterion?Discard or follow procedure; never override
Processing recordEvery step and test recorded, with the run linked to its water, germicide and reprocessor checks?Hold until resolved, or discard

After this, the germicide, contact time, storage and residual-test gates follow (Module 10).

RememberTen gates, ten separate proofs. A passing gate cannot vote for a failing one.

Repeating a test without gaming it

Measurements can go wrong for innocent reasons: air in the line, a loose connection, a sensor fault. Your procedure may allow one repeat for a defined reason. What it never allows is repeating a test until a passing number appears and keeping only that one.

Figure 4 · Legitimate repeats

Defined reasonA recorded cause, such as a loose connection, that the procedure names.
Every result keptThe first result stays in the record beside the repeat.
Limited numberNo more repeats than the procedure allows.
Pattern escalatedSeveral inconsistent results on different dialyzers suggest an equipment problem; report it to the biomedical technician (Module 11).

A failed dialyzer stays unavailable until its disposition follows the program. A later finding that the reprocessor was faulty does not erase the dialyzer’s recorded failure; it adds a question about every dialyzer processed on that reprocessor (Module 11).

RememberEvery result belongs in the record. A test repeated until it passes proves only that the test can be beaten.

Back to the case: Seventy-nine and eighty

The three results. RU-0731: 79 ÷ 100 × 100 = 79%, below the 80% floor. It fails the volume gate. RU-0732: 86%, which meets the floor, but its integrity test failed. RU-0733: 104 ÷ 98 × 100 = 106%, above baseline. That is not a super-dialyzer; it suggests a measurement or record error and must be flagged for review, not accepted.

Can RU-0732’s volume make up for the failed pressure test? No. The two tests look at different dangers. A broken fiber can mix blood and dialysis fluid regardless of how many fibers are open. RU-0732 follows the procedure for an integrity failure.

When is a repeat legitimate? Only for a defined, recorded reason the procedure allows, with the first result kept. Running RU-0731 “until it passes” is gaming. If several dialyzers that day gave odd readings, Ms. Alcantara reports a possible equipment fault before more dialyzers are processed.

Skills checklist: Acceptance decisions on ten records

Use ten synthetic processing records and the Reprocessing Technician Toolkit, then observe acceptance testing on the unit’s reprocessor. Your preceptor rates each step as satisfactory, needs practice, or not observed. Steps marked Critical must all be satisfactory before sign-off. This checklist supports supervised learning; your unit’s own procedure and competency form are the authority. A printable version of every checklist is on the trainer page.

Placement activity. During supervised practice, handle at least one failed volume and one failed integrity result from start to disposition, observed by your assessor.

Check your understanding

Each visit draws 5 questions at random from this module’s 12-question bank. Choose the single best answer for each, then select Check answers; an explanation appears under the best answer. Select New set of questions for a different mix. A good target is 4 of 5 on two different sets. Nothing you select is sent anywhere; scoring happens in your browser.

Question 1 of 12Baseline 100 mL, measured 85 mL. What is the retained TCV?
Question 2 of 12Baseline 100 mL, measured 79 mL, floor 80%. What is the decision on volume?
Question 3 of 12A dialyzer has 90% retained volume but failed its integrity test. What happens?
Question 4 of 12A record has no baseline TCV. What do you do?
Question 5 of 12Measured volume is above the baseline. What is the correct response?
Question 6 of 12What does a pressure-hold test detect?
Question 7 of 12What did the PhilHealth quality circular set for retained TCV?
Question 8 of 12Your colleague repeats a failed volume test four times and records only the passing result. What is wrong?
Question 9 of 12Several dialyzers today gave inconsistent volume readings. What should you suspect and do?
Question 10 of 12Why is a catalog volume not used as the baseline?
Question 11 of 12A dialyzer passes volume, integrity and inspection, but its processing record has a blank where the reprocessor run number should be. Is it ready?
Question 12 of 12A dialyzer at exactly 80% retained TCV means:

Read next on this site

Glossary & abbreviationsTalahulugan at mga daglatTalaan sa mga pulong ug daglatTalatinigan ampo reng daglat terms used on this page

Abbreviations

AAMI
Association for the Advancement of Medical Instrumentation.
CDC
Centers for Disease Control and Prevention.
PhilHealth
Philippine Health Insurance Corporation.
TCV
Total cell volume — the volume of the blood compartment of a dialyzer, measured to estimate how many fibers are still open

Terms in plain words

Total cell volume
The volume of a dialyzer’s blood compartment.
Baseline
This dialyzer’s TCV measured before first use.
Retained TCV
Measured TCV as a percentage of baseline.
Integrity test
A pressure-hold test that detects a broken fiber or leak.
Acceptance gate
A condition that must pass on its own before release.
Disposition
The recorded decision: release, reprocess or discard.
Gaming
Repeating or selecting results to obtain a pass.
ReferencesMga SanggunianMga TinubdanReng Reperensya 3 sources
  1. Philippine Health Insurance Corporation (PhilHealth). (2024). PhilHealth Circular No. 2024-0029 [Quality standards for hemodialysis facilities]. https://www.philhealth.gov.ph/circulars/2024/PC2024-0029.pdf
  2. Association for the Advancement of Medical Instrumentation (AAMI). (2008). Reprocessing of hemodialyzers (ANSI/AAMI RD47:2008/(R)2013).
  3. Gordon, S. M., Oettinger, C. W., Bland, L. A., Oliver, J. C., Arduino, M. J., Aguero, S. M., McAllister, S. K., Favero, M. S., & Jarvis, W. R. (1992). Pyrogenic reactions in patients receiving conventional, high-efficiency, or high-flux hemodialysis treatments with bicarbonate dialysate containing high concentrations of bacteria and endotoxin. Journal of the American Society of Nephrology, 2(9), 1436–1444. https://doi.org/10.1681/ASN.V291436
Dr. William Gregory M. Rivero, MD

William Gregory Rivero, MD, FPCP, DPSN

Internal Medicine · Nephrology · Nutrition · Philippines · PRC 0105184

Educational study material. Completing it is an Academy learning record only. It is not a DOH, PSN, PhilHealth, TESDA or PRC credential, it does not authorize anyone to release a dialyzer, and it does not replace the unit’s approved reuse program, the device and germicide instructions for use, or supervised practice.

QR code — scan to save Dr. Rivero's contact info

Scan and save