1 · Total cell volume (TCV) retained
Modules 02, 09
Formula. Retained TCV (%) = measured ÷ baseline × 100. The baseline is this dialyzer’s own first measurement, never a catalog value. A missing, zero or negative baseline returns HOLD. A measured value above baseline is flagged for review, never capped or accepted automatically. Meeting the floor passes one gate only: identity, eligibility, uses left, inspection, integrity, the processing record, germicide, storage time and the residual test must all pass too.
2 · Use counter and acceptance gates
Modules 03, 09
Logic. Next use = completed uses + 1. The limit is the stricter of the payer ceiling and any IFU or facility limit. A next use above the limit is NOT ELIGIBLE, whatever the tests say. Any failed gate makes the dialyzer NOT ELIGIBLE; any gate not recorded returns HOLD. Uses are not reuses: five uses means the first treatment plus four reuses. A processing cycle that had to be repeated is not a use.
3 · Germicide dilution check (C1V1 = C2V2)
Module 06
Formula. Volume of concentrate V1 = (target C2 × final volume V2) ÷ stock C1; water = V2 − V1. No default values are given because concentrations are product-specific. The arithmetic predicts the strength; only the potency test proves it. In a CDC-investigated outbreak, poorly mixed germicide measured 1.4% at the top and 3.5% at the bottom of the same container (Beck-Sague et al., 1990). Mix in the order and manner the IFU sets, then test.
4 · Contact time and storage clock
Module 10
Formula. Earliest use = fill time + minimum contact time. Expiry = fill time + maximum storage. Before the earliest use the dialyzer is not yet disinfected; after expiry it is outside the proven window. Either way it is reprocessed or discarded as your procedure directs. A missing fill time returns HOLD.
5 · Dialysis water log check
Module 05
Rules shown. DOH AO 2013-0003: weekly product-water TDS with corrective action above 10 mg/L; microbiology at least monthly. The default microbiology and endotoxin values follow ISO 23500-3 dialysis water (maximum 100 CFU/mL and 0.25 EU/mL, action levels at half); check them against the edition your facility follows. Green = below action level; amber = at or above the action level (investigate and correct); red = above the maximum, or a test overdue (TDS older than 7 days, microbiology older than 31 days). The medical director owns water quality; this tool never clears water for use.
6 · Monthly reuse audit
Module 12
Formulas. Average uses per dialyzer = treatments ÷ dialyzers used. Discard rate by reason = discards for that reason ÷ processing cycles. Reactions per 1,000 treatments = reactions ÷ treatments × 1,000. Average uses is a result of good care, not a target: rewarding a higher number can lead staff to pass dialyzers that should fail. Read the numbers together; a rise in reactions is always investigated.