Nephrology Nursing · Role subtrack · For trainers

Trainer and Preceptor Resource

Program structure, completion criteria, and printable skills checklists for the Dialyzer Reprocessing Technician series.

PublishedNailathalaGipatikPepalwal: Audience: Head nurses, chief technicians, educators, and preceptors Read timeOras ng pagbasaOras sa pagbasaOras ning pamamasa:
Renal Care Academy›Nephrology Nursing Track›By Role›Dialyzer Reprocessing Technician›Trainer and preceptor resource

Running the track in your unit

Preparation. Obtain the approved reuse program, the reprocessor operating manual, the germicide and dialyzer instructions for use, the safety data sheets, and blank logs. Use training dialyzers and synthetic labels (for example, RU-0412) until workplace eligibility is confirmed. Never use patient identifiers in classroom material.

Figure 1 · Suggested sequence

  1. Weeks 1 and 2Modules 1 to 4; induction, safety orientation, observation in the reuse room.
  2. Weeks 3 and 4Modules 5 to 7; water sampling lab, germicide handling lab, receiving and labeling practice.
  3. Weeks 5 and 6Modules 8 and 9; cleaning and testing on training dialyzers, then supervised cycles.
  4. Weeks 7 and 8Module 10; storage, rinse, residual testing and two-person handoffs.
  5. Weeks 9 and 10Module 11; calibration walk-through, fault and quarantine drills, power and water failure drill.
  6. Weeks 11 to 13 or laterModule 12; audit exercise, capstone, remaining supervised hours, practical assessment.

Record each observation with the learner, date, equipment model, standard operating procedure (SOP) version, simulated or real, critical items met, and the assessor’s signature, in the facility’s own system. Reassess after a change of device or germicide, a serious incident, a long absence, and yearly.

Proposed completion criteria

Completing the Academy track is a learning record. It does not authorize the release of a dialyzer.

Written examination. The end-of-course examination draws 40 questions at random from the question banks of all modules. A proposed pass mark is 80 percent, with missed questions reviewed against their modules before a second sitting. For a supervised sitting, the learner takes it on a unit device with the preceptor present and the preceptor records the score.

Skills checklists for every module

Print this page for a full set. Each checklist starts on a new sheet. Steps marked Critical must all be satisfactory before sign-off; a missed critical step is repeated with the preceptor. Replace any step that does not match your equipment or approved procedure.

Module 1: The Reuse Room and Your Place in It

Responsibility map and escalation drill. Complete this with your preceptor, using a training dialyzer with a deliberately incomplete record.

Observable stepSatisfactoryNeeds practiceNot observed
CriticalNames the owner of each decision: eligibility, use on the patient, processing, equipment fitness, contamination response.
CriticalIdentifies the missing gate in the record and states that the dialyzer is unproven.
CriticalApplies a quarantine label and moves the dialyzer to the quarantine shelf.
CriticalStates that the patient receives a new dialyzer while the question is answered.
CriticalRefuses, politely and clearly, a request to release outside authorization.
Names the person to call when no authorized operator is present.
Writes a short factual report: what, when, where it may come from, what was done.
States two tasks outside the operator’s scope.

Learner ____________________   Preceptor ____________________   Date __________

Module 2: The Dialyzer and Why Reuse Exists

Reading a dialyzer and its history. Use a cut-open training dialyzer and a set of synthetic processing records.

Observable stepSatisfactoryNeeds practiceNot observed
CriticalIdentifies the blood ports, the dialysate ports, the headers and the fiber bundle.
Explains diffusion and ultrafiltration in one sentence each.
CriticalRecognizes clot in a header and dark streaks in a fiber bundle.
CriticalCalculates retained volume for three records and identifies a falling trend that still passes.
States what the outcome evidence on reuse does and does not show.
Names the people who receive a trend report and what they may decide.

Learner ____________________   Preceptor ____________________   Date __________

Module 3: The Rules: Philippine Governance and Other Countries

Rule-finding and use-count drill. Use printed copies of your unit’s SOP, the relevant issuances, and five synthetic dialyzer records with mixed counting systems.

Observable stepSatisfactoryNeeds practiceNot observed
Names the source that answers each of five questions (licensing, water, eligibility, payment, product use).
CriticalConverts each record to a use number and identifies the one that would reach a sixth use.
CriticalDistinguishes a processing cycle from a use.
CriticalApplies the stricter of an instructions for use (IFU) limit and the payer ceiling.
CriticalHolds a dialyzer whose infection eligibility is unknown or conflicting.
Checks the issue date and current status of one rule.

Learner ____________________   Preceptor ____________________   Date __________

Module 4: Infection Prevention and Protecting Yourself

Protective equipment and exposure drill. Practice in the reuse room with your preceptor, using water in place of blood and germicide.

Observable stepSatisfactoryNeeds practiceNot observed
CriticalPuts on gown, gloves, mask and face shield in the order the procedure sets, and removes them without contaminating hands or face.
CriticalKeeps the face shield down for the whole splash-risk task.
CriticalChanges gloves and performs hand hygiene between dirty and clean tasks.
Identifies the clean and dirty areas and the one-way path in the room.
CriticalLocates the eye wash, the spill kit and the safety data sheet (SDS) binder, and checks the eye wash flow.
CriticalDemonstrates eye flushing with lids held open.
CriticalStates who to report an exposure to and how quickly.
CriticalStates what to do with a dialyzer when a patient’s infection status changes.

Learner ____________________   Preceptor ____________________   Date __________

Module 5: Water for Reprocessing

Water sampling and log review. Practice with a sterile bottle on a training outlet, then review a set of synthetic water logs.

Observable stepSatisfactoryNeeds practiceNot observed
CriticalLabels the bottle with outlet, date, time and collector before opening it.
Follows the laboratory’s outlet preparation steps.
CriticalFills the bottle without touching the inside of the cap or the outlet tip.
CriticalReviews the water log before starting the reprocessor.
CriticalIdentifies a result at an action level and one above a limit.
CriticalStops and asks when the log shows an unsigned failure.
Records who was told and when.

Learner ____________________   Preceptor ____________________   Date __________

Module 6: Germicides and Chemical Safety

Germicide preparation and spill drill. Practice with water and a dye in place of germicide, then with the real product under supervision.

Observable stepSatisfactoryNeeds practiceNot observed
CriticalChecks product name, expiry and container integrity before use.
Calculates the volume of concentrate with C1V1 = C2V2 and has the arithmetic checked.
CriticalPrepares the solution in the order and with the mixing method the IFU sets.
CriticalPerforms and records a potency test on the batch before use.
CriticalWears chemical-compatible gloves, eye and face protection and a gown.
Locates the SDS and reads the first-aid and protection sections aloud.
CriticalContains and cleans a simulated spill with the spill kit.
CriticalStates why bleach must never be mixed with acid germicides.

Learner ____________________   Preceptor ____________________   Date __________

Module 7: Identity and Chain of Custody

Labeling and hand-over drill. Use synthetic patients with similar names and training dialyzers.

Observable stepSatisfactoryNeeds practiceNot observed
CriticalLabels a new dialyzer with two identifiers and the required fields before first use.
CriticalChecks a received dialyzer against the receiving log and eligibility list.
CriticalPicks a dialyzer from storage by label and two identifiers, not by position.
CriticalAsks for the full name and second identifier when given a surname only.
CriticalPerforms a two-person read-back at release.
CriticalQuarantines a dialyzer with an unreadable label.
States why a use count is never reset.

Learner ____________________   Preceptor ____________________   Date __________

Module 8: Receiving, Rinsing and Cleaning

Receiving and cleaning on a training dialyzer. Practice receiving and the cleaning steps on training dialyzers and the unit’s reprocessor under supervision.

Observable stepSatisfactoryNeeds practiceNot observed
CriticalChecks identity, eligibility, use count, time since treatment and condition at receiving, and records the time.
CriticalRejects a dialyzer outside the time limit and follows the procedure.
CriticalConfirms the start-of-run checks before starting the reprocessor.
CriticalPerforms the rinse and reverse ultrafiltration steps as the procedure sets.
CriticalCleans the header with the face shield down and without damaging seals, only if the IFU permits.
CriticalUses only the approved cleaning agent at the approved strength.
Records each step as it is done, not afterward.
Checks and records the used-dialyzer refrigerator temperature.

Learner ____________________   Preceptor ____________________   Date __________

Module 9: Performance Testing and Acceptance

Acceptance decisions on ten records. Use ten synthetic processing records and the Reprocessing Technician Toolkit, then observe acceptance testing on the unit’s reprocessor.

Observable stepSatisfactoryNeeds practiceNot observed
CriticalCalculates retained total cell volume (TCV) correctly for every record.
CriticalHolds a record with a missing or implausible baseline.
CriticalFlags a measured volume above baseline for review.
CriticalRejects a dialyzer with a failed integrity test regardless of volume.
CriticalIdentifies the record that would reach a sixth use.
Inspects a processed training dialyzer and records findings.
CriticalStates when a repeat test is legitimate and keeps every result.
Writes a disposition (release to storage, reprocess, discard, hold) for each record.

Learner ____________________   Preceptor ____________________   Date __________

Module 10: Disinfect, Store, Rinse and Hand Off

Storage, residual testing and handoff. Practice on training dialyzers filled with water, then observe and perform under supervision.

Observable stepSatisfactoryNeeds practiceNot observed
CriticalChecks that both compartments are full, without air, and all four ports are capped.
CriticalRecords the fill time to the minute on the label and in the log.
CriticalCalculates the earliest-use time and expiry from IFU values.
Places the dialyzer in the patient’s labeled position and performs the daily storage check.
CriticalUses the correct, in-date residual strip for the germicide.
CriticalReads the residual test at the exact time its instructions set.
CriticalTreats a doubtful color as a failed result.
CriticalCompletes a two-person verification, each person checking independently.

Learner ____________________   Preceptor ____________________   Date __________

Module 11: Machines, Faults and When Things Go Wrong

Fault, trace-back and continuity drill. A tabletop exercise with synthetic logs, then a walk-through at the reprocessor with the biomedical technician.

Observable stepSatisfactoryNeeds practiceNot observed
CriticalPerforms and records the reprocessor’s daily user checks with actual readings.
CriticalTags out a faulty reprocessor and refuses to use it until written return to service.
CriticalIdentifies the trace-back window from a set of logs.
CriticalLists and quarantines the dialyzers processed in the window.
CriticalKeeps the dialyzer and lines from a patient with a pyrogenic reaction.
Checks the processing, germicide, water and reprocessor records for a cluster.
CriticalStates the first actions for power, water, refrigeration and reprocessor failures.
CriticalDistinguishes equipment return to service from dialyzer release.

Learner ____________________   Preceptor ____________________   Date __________

Module 12: Records, Quality and the Capstone

Record audit and capstone. Audit twenty synthetic records, then work through the capstone case aloud with your assessor.

Observable stepSatisfactoryNeeds practiceNot observed
CriticalIdentifies every record with a missing or inconsistent gate.
CriticalCorrects a sample record error by the single-line method, without erasing.
Calculates average uses, discard rate by reason and reactions per 1,000 treatments from a monthly summary.
CriticalExplains why a rising average use number is not a quality target.
CriticalDisposes of a discarded training dialyzer and germicide as the procedure and SDS direct.
CriticalWorks the capstone from receiving to handoff, naming each gate and its proof.
CriticalRecognizes the use limit in the capstone and the trace-back window after the fault.
States who decides each unproven question in the capstone.

Learner ____________________   Preceptor ____________________   Date __________

Glossary & abbreviationsTalahulugan at mga daglatTalaan sa mga pulong ug daglatTalatinigan ampo reng daglat terms used on this page

Abbreviations

IFU
Instructions for use — the manufacturer’s written directions for a device or chemical
SDS
Safety data sheet.
SOP
Standard operating procedure.
TCV
Total cell volume — the volume of the blood compartment of a dialyzer, measured to estimate how many fibers are still open
Dr. William Gregory M. Rivero, MD

William Gregory Rivero, MD, FPCP, DPSN

Internal Medicine · Nephrology · Nutrition · Philippines · PRC 0105184

Educational study material. Completing it is an Academy learning record only. It is not a DOH, PSN, PhilHealth, TESDA or PRC credential, it does not authorize anyone to release a dialyzer, and it does not replace the unit’s approved reuse program, the device and germicide instructions for use, or supervised practice.

QR code — scan to save Dr. Rivero's contact info

Scan and save