Nephrology Nursing · Role subtrack · Module 8 of 12

Receiving, Rinsing and Cleaning

Time to processing, rinsing and reverse ultrafiltration, header cleaning, cleaning agents, and what the approved procedure controls.

PublishedNailathalaGipatikPepalwal: ReferencesMga SanggunianMga TinubdanReng Reperensya: 3 Audience: Reprocessing operators and dialysis technicians in Philippine hemodialysis units · biomedical technicians who service reprocessors · head nurses and medical directors who supervise reuse programs Read timeOras ng pagbasaOras sa pagbasaOras ning pamamasa:
A reprocessing technician connecting a used dialyzer to an automated reprocessor, with a refrigerator of waiting dialyzers and a temperature log beside it.
Renal Care Academy›Nephrology Nursing Track›By Role›Dialyzer Reprocessing Technician›Module 8 of 12 Nephrology Nursing subtrack for the dialyzer reuse room
♻️

Nephrology Nursing subtrack for the dialyzer reuse room

This module is study material for dialyzer reprocessing. It explains the reasoning behind each step; it is not a procedure. Concentrations, contact times, storage limits and test thresholds come from the germicide and dialyzer instructions for use and your unit’s validated procedure, never from this page. The medical director decides who is eligible for reuse; the nurse decides whether a dialyzer is used on a patient. Every case, name and identifier is fictional. Pending clinical review.

After this module you should be able to

Classroom 6 hSkills lab 6 hSupervised placement 48 h
Pre-testBefore you read, answer the 5 self-assessment questions and note your score. After the module, select New set of questions and compare. The difference is what you learned; the questions you still miss are what to reread.

The big picture, in plain words

In plain wordsWashing a pot right after cooking is easy. Leave it overnight and the food dries on, and you scrub for twenty minutes and still miss some. Leave it in the sun and it starts to smell.

The picture has a limit. Food on a pot is visible, and you can scrub it. Blood and protein left inside ten thousand fibers cannot be seen or scrubbed; they can only be rinsed and lifted out by water and cleaning agents moving through the right paths. Bacteria in that residue multiply with time and warmth. So the clock starts at the end of treatment, and the cleaning follows a method that has been shown to work on this dialyzer model.

Figure 1 · Key terms for this module

ReceivingAccepting a used dialyzer into the reuse room, with identity, eligibility and time checked.
HeaderThe cap at each end of the dialyzer where blood enters or leaves the fibers.
RinseFlushing water through the blood compartment to remove blood.
Reverse ultrafiltrationPushing water from the dialysate side into the fibers to lift deposits from the inner wall.
Cleaning agentA chemical used before disinfection to remove protein and blood residue.
Validated procedureA method shown, for this dialyzer and reprocessor, to achieve the result it claims.

Your work is to start the cleaning in time, follow the validated steps exactly, and stop when a dialyzer cannot be cleaned within what the procedure allows.

Case: Three hours in the hallway

At 3:15 on a Wednesday afternoon, Ms. Nina Aquino finds a used dialyzer in a tray on a cart in the hallway outside the reuse room. Its label is clear, and the patient is eligible. The run sheet shows the treatment ended at 12:05. The hallway is warm. The refrigerator in the reuse room has space, and its temperature log was last signed at 7:00.

The dialyzer is on its third use. The patient is a 70-year-old woman whose new dialyzer would cost the unit another one from its yearly allowance.

Before reading on, decide:

Keep your answers in mind. The lesson gives you what you need, and the case review near the end shows one sound response.

Receiving a used dialyzer

Six numbered steps of dialyzer reprocessing, from rinsing through testing, disinfection, and labeling.
Figure 2. The reprocessing chain. Each step leaves a record, and a dialyzer with a missing record is not used.

Receiving is the first gate inside the room. A dialyzer that fails it is not processed.

CheckPassHold or reject
IdentityLabel readable, matches the receiving logMissing, unreadable or mismatched label
EligibilityPatient on the current eligibility listNot listed, status changed, or unknown
Use countNext use within the limitLimit reached
Time since end of treatmentWithin the limit your procedure sets, or refrigerated within itOver the limit without refrigeration
ConditionHousing intact, ports cappedCracks, leaks, gross clot, unusual color
RememberReceiving is a gate, not a drop-off. Record the time the dialyzer arrived and what you checked.

Time and temperature before processing

A clock beside a used dialyzer: warm room with bacteria multiplying on one side, and a refrigerator with a temperature log on the other, with a hold label for a dialyzer that exceeded the time allowed.
Figure 3. Time is part of cleanliness. Your procedure sets how long a used dialyzer may wait, and where.

After treatment, the fibers hold blood, protein and whatever bacteria entered the circuit. At room temperature, bacteria multiply; dried protein sticks to the membrane. Both make cleaning less effective and raise endotoxin. That is why programs set a maximum time to processing and require refrigeration for dialyzers that will wait longer. The exact time and temperature come from your procedure, which follows the Association for the Advancement of Medical Instrumentation (AAMI) standard and the reprocessor maker.1

The refrigerator is part of the system. A refrigerator for used dialyzers holds only used dialyzers, never food or medicines, and its temperature is logged. The United States Food and Drug Administration (FDA) advises that dialyzers should not be reused when reprocessing or required refrigeration facilities are not working.2

Stop and reportA used dialyzer that has passed the time limit without refrigeration is not processed. Follow the procedure, usually discard, and report how it happened.

Rinsing, reverse ultrafiltration and the headers

Cutaway of a dialyzer during reverse ultrafiltration: clean water pushed from the dialysate side into the fibers, lifting protein and clot out of the blood compartment, with the header being cleaned separately.
Figure 4. Reverse ultrafiltration pushes water backward through the membrane to lift deposits from the inside of the fibers.

Cleaning works from the outside in. First, water flushes the blood out of the blood compartment. Then reverse ultrafiltration pushes water backward through the membrane, from the dialysate side into the fibers, lifting protein and small clots off the inner wall. A cleaning agent, if the program uses one, dissolves what remains. Last, the headers are cleaned, because clot often collects there where blood enters the fibers.

StepWhat it removesCommon error
Initial rinseLoose bloodStarting with a warm, dried dialyzer
Reverse ultrafiltrationProtein and small clots on the inner fiber wallPressure outside the reprocessor’s set range
Cleaning agentRemaining proteinWrong agent, wrong strength, or skipped rinse afterward
Header cleaningClot at the fiber openingsRemoving the header without authorization or damaging the O-ring; splash to the face

Header removal and cleaning carry a splash risk, so the face shield stays down (Module 4). Some dialyzer models must not have their headers removed; the instructions for use (IFU) says which.

RememberRinse before the residue dries. Clean in the order the procedure sets. A step skipped is a step unproven.

What cleaning agents do to the membrane

Cleaning agents are chosen for a dialyzer model by the program. They can change the membrane itself. In the Hemodialysis (HEMO) Study (HEMO) analysis, the clearance of beta-2-microglobulin with reuse depended on both the membrane material and the reprocessing technique; with some combinations it fell, and with others it rose substantially.3 A membrane that becomes more open can lose more albumin into the dialysis fluid. This is why a program cannot swap one agent for another, or add an extra cleaning step, without validation and the medical director’s approval.

Agents must also be compatible with the germicide that follows. Some combinations react dangerously; Module 6 explains why bleach must never meet acid germicides outside a validated sequence.

RememberA stronger clean is not a better clean. Use only the agent, strength and sequence your validated procedure names.

Following the approved procedure

This module describes the logic of cleaning. The steps you perform come from your unit’s approved procedure, written for its own reprocessor, dialyzer models and agents. This track does not publish a recipe, because a recipe that works for one system can fail on another.

Figure 5 · Before you start a processing run

ProcedureCurrent version at hand, signed by the medical director.
ReprocessorDaily checks and calibration recorded (Module 11).
WaterLog reviewed, no unsigned failures (Module 5).
Agents and germicideCorrect products, in date, batch potency recorded (Module 6).
DialyzerReceived, identified, eligible, within time (this module).
Stop and reportIf any start-of-run check is missing, do not start. A run that begins on an unproven system makes every dialyzer in it unproven.

Back to the case: Three hours in the hallway

What is the problem? The dialyzer has waited more than three hours at warm room temperature since the end of treatment. Bacteria may have multiplied and protein may have dried onto the membrane. Unless your procedure allows that time without refrigeration, it is outside the limit.

What does the procedure require? Usually that such a dialyzer is not processed and is discarded, with a record. The decision follows the written procedure; if it is unclear, the head nurse or medical director’s designate decides. The cost of one dialyzer and the yearly allowance do not change the safety decision; Module 3 explained that a payer allowance never justifies using a doubtful dialyzer.

How can the unit prevent it? A rule that used dialyzers go straight to the reuse room or its refrigerator, a log of time received, a hallway cart that is never a parking place, and a shift hand-over that includes the reuse room. Ms. Aquino reports the event so the cause can be fixed without blame.

Skills checklist: Receiving and cleaning on a training dialyzer

Practice receiving and the cleaning steps on training dialyzers and the unit’s reprocessor under supervision. Your preceptor rates each step as satisfactory, needs practice, or not observed. Steps marked Critical must all be satisfactory before sign-off. This checklist supports supervised learning; your unit’s own procedure and competency form are the authority. A printable version of every checklist is on the trainer page.

Placement activity. For one week, record the time from end of treatment to receiving for every dialyzer. Show your preceptor the longest five and their causes.

Check your understanding

Each visit draws 5 questions at random from this module’s 12-question bank. Choose the single best answer for each, then select Check answers; an explanation appears under the best answer. Select New set of questions for a different mix. A good target is 4 of 5 on two different sets. Nothing you select is sent anywhere; scoring happens in your browser.

Question 1 of 12Why does time before processing matter?
Question 2 of 12Where do the maximum time to processing and the refrigeration rules come from?
Question 3 of 12What does reverse ultrafiltration do?
Question 4 of 12Where does clot often collect in a used dialyzer?
Question 5 of 12A dialyzer has been in a warm hallway for three hours. Your procedure’s limit is shorter. What do you do?
Question 6 of 12What may a different cleaning agent do to a membrane?
Question 7 of 12A colleague suggests adding an extra bleach step to get dialyzers cleaner. What is the correct response?
Question 8 of 12What may be stored in the used-dialyzer refrigerator?
Question 9 of 12According to the FDA, what should happen when reprocessing or required refrigeration is not working?
Question 10 of 12Which check must be done before starting a processing run?
Question 11 of 12When should each cleaning step be recorded?
Question 12 of 12The IFU for a dialyzer model says the headers must not be removed. What do you do?

Read next on this site

Glossary & abbreviationsTalahulugan at mga daglatTalaan sa mga pulong ug daglatTalatinigan ampo reng daglat terms used on this page

Abbreviations

AAMI
Association for the Advancement of Medical Instrumentation.
HEMO
Hemodialysis (HEMO) Study — a United States randomized trial of dialysis dose and membrane flux
IFU
Instructions for use — the manufacturer’s written directions for a device or chemical

Terms in plain words

Receiving
Accepting a used dialyzer into the reuse room after checks.
Header
The cap at each end of the dialyzer where blood enters or leaves the fibers.
Reverse ultrafiltration
Pushing water from the dialysate side into the fibers to clean them.
Cleaning agent
A chemical that removes protein and blood residue before disinfection.
Validated procedure
A method shown to achieve its claimed result on that system.
O-ring
A rubber seal between the header and the dialyzer body.
ReferencesMga SanggunianMga TinubdanReng Reperensya 3 sources
  1. Association for the Advancement of Medical Instrumentation (AAMI). (2008). Reprocessing of hemodialyzers (ANSI/AAMI RD47:2008/(R)2013).
  2. U.S. Food and Drug Administration (FDA). (n.d.). Reopening dialysis clinics after restoration of power and water. https://www.fda.gov/medical-devices/emergency-situations-medical-devices/reopening-dialysis-clinics-after-restoration-power-and-water
  3. Cheung, A. K., Agodoa, L. Y., Daugirdas, J. T., Depner, T. A., Gotch, F. A., Greene, T., Levin, N. W., & Leypoldt, J. K. (1999). Effects of hemodialyzer reuse on clearances of urea and beta2-microglobulin. Journal of the American Society of Nephrology, 10(1), 117–127. https://doi.org/10.1681/ASN.V101117
Dr. William Gregory M. Rivero, MD

William Gregory Rivero, MD, FPCP, DPSN

Internal Medicine · Nephrology · Nutrition · Philippines · PRC 0105184

Educational study material. Completing it is an Academy learning record only. It is not a DOH, PSN, PhilHealth, TESDA or PRC credential, it does not authorize anyone to release a dialyzer, and it does not replace the unit’s approved reuse program, the device and germicide instructions for use, or supervised practice.

QR code — scan to save Dr. Rivero's contact info

Scan and save