Nephrology Nursing · Role subtrack · Module 5 of 12

Water for Reprocessing

Why water quality matters even more in the reuse room, the DOH monitoring schedule, ISO limits and action levels, and sampling.

PublishedNailathalaGipatikPepalwal: ReferencesMga SanggunianMga TinubdanReng Reperensya: 8 Audience: Reprocessing operators and dialysis technicians in Philippine hemodialysis units · biomedical technicians who service reprocessors · head nurses and medical directors who supervise reuse programs Read timeOras ng pagbasaOras sa pagbasaOras ning pamamasa:
A reprocessing technician filling a sterile sample bottle from a labeled water outlet in the reuse room, with the water treatment loop visible through a window.
Renal Care Academy›Nephrology Nursing Track›By Role›Dialyzer Reprocessing Technician›Module 5 of 12 Nephrology Nursing subtrack for the dialyzer reuse room
♻️

Nephrology Nursing subtrack for the dialyzer reuse room

This module is study material for dialyzer reprocessing. It explains the reasoning behind each step; it is not a procedure. Concentrations, contact times, storage limits and test thresholds come from the germicide and dialyzer instructions for use and your unit’s validated procedure, never from this page. The medical director decides who is eligible for reuse; the nurse decides whether a dialyzer is used on a patient. Every case, name and identifier is fictional. Pending clinical review.

After this module you should be able to

Classroom 6 hSkills lab 6 hSupervised placement 40 h
Pre-testBefore you read, answer the 5 self-assessment questions and note your score. After the module, select New set of questions and compare. The difference is what you learned; the questions you still miss are what to reread.

The big picture, in plain words

In plain wordsWhen you wash a baby’s bottle, you rinse it with clean water, not with water from a puddle. The bottle can be spotless and still make the baby sick if the last rinse was dirty.

The picture has a limit. A bottle’s rinse water touches the outside of what the baby swallows. Reprocessing water fills the blood compartment of the dialyzer, the inside of the fibers where the patient’s blood will flow next treatment. Bacteria and their toxins left there meet the blood directly. So the reuse outlet is treated as a point of use, sampled and protected like a dialysis station.

Figure 1 · Key terms for this module

Product waterWater leaving the treatment system, usually after reverse osmosis.
Point of useAn outlet where water enters a process that touches patients, such as a dialysis machine or the reprocessor.
EndotoxinA toxin from the walls of certain bacteria. It can cause fever and chills even when the bacteria are dead.
Colony-forming units (CFU)A count of live bacteria grown from a water sample.
Total dissolved solids (TDS)A measure of how much dissolved mineral and salt remains in the water.
Action levelA value below the limit at which you must investigate and correct.

Your role is not to run the water plant. It is to know where the reuse outlet sits in the system, to sample and record correctly when assigned, to read the logs before you start, and to stop when the evidence says the water is not fit.

Case: Monday morning

At 6:00 on a Monday, Mr. Arnold Mendoza opens the reuse room at a unit in Laguna. Before starting the reprocessor, he checks the water log. The weekly product-water TDS reading taken on Saturday is 14 mg/L. The action limit written on the log is 10 mg/L. Nobody has signed the corrective-action box.

Twenty used dialyzers from Saturday’s last shift are in the refrigerator waiting to be processed. The water technician starts work at 8:00.

Before reading on, decide:

Keep your answers in mind. The lesson gives you what you need, and the case review near the end shows one sound response.

Why reprocessing water matters

Water is used to rinse the blood compartment, to clean by reverse ultrafiltration, to dilute germicide, and to rinse germicide out before treatment. Each time, it fills the inside of the fibers. The reprocessing standard of the Association for the Advancement of Medical Instrumentation (AAMI) expects reprocessing water to meet the quality required for dialysis water.1

Outbreak investigations show what happens when it does not. In a Centers for Disease Control and Prevention (CDC) investigation of one United States center, water used to prepare dialysis fluid and to reprocess dialyzers exceeded the allowed bacteria and endotoxin levels; 13 episodes of bacteremia or pyrogenic reaction followed. They were traced to the water and to one machine type that was not heat-disinfected as its maker advised, not to reuse itself, but the same water was filling reprocessed dialyzers.2 In Hanoi, 303 dialysis patients had pyrogenic reactions over nine months; the bacteria found were largely waterborne, and investigators named suboptimal water or reprocessing as the likely cause.3 Most dialysis outbreaks in a CDC review were preventable by adequate water treatment, disinfection and adherence to reprocessing protocols.4

The most severe case was in Caruaru, Brazil, in 1996. Water from a reservoir reached a dialysis center without proper treatment. It carried microcystins, toxins made by blue-green algae. Of 130 patients, 101 developed acute liver failure, and 50 died.5 That center did not reuse dialyzers; the lesson is that water reaches the blood, and anything in it reaches the patient.

RememberWater that fills the fibers reaches the next patient’s blood. The reuse outlet is a point of use.

From the plant to the reuse outlet

Water path from feed water through pretreatment, reverse osmosis, storage and the distribution loop to the reprocessing outlet, with sampling points for chemical, microbiology and endotoxin marked.
Figure 2. The reuse outlet is a point of use. It is sampled and protected like a dialysis station.
StageWhat it doesWhat can go wrong for the reuse room
Pretreatment (filters, softener, carbon)Removes particles, hardness and chlorine or chloramineExhausted carbon lets chlorine through, which can damage membranes and patients
Reverse osmosisRemoves most dissolved solids and bacteriaA failing membrane raises TDS and conductivity
Storage and distribution loopDelivers product water to points of useBiofilm grows in dead legs and stagnant pipes
Reprocessing outlet and hoseFeeds the reprocessor or the manual rinseUnlabeled outlet, old hose, outlet left running into a sink, contamination at the tip

The Malaysian 2026 standards, a current regional comparison, require labeled water outlets in the reprocessing area.6 Biofilm, a slimy layer of bacteria on pipe walls, is the usual source of persistent contamination. It shelters bacteria from disinfectant and sheds endotoxin into the water.

RememberKnow which loop feeds your outlet, and report anything that changes it: a new hose, a dead leg, a leak.

The monitoring schedule, limits and action levels

The DOH water order sets the minimum monitoring for licensed facilities.7

TestWhereHow often (DOH minimum)
MicrobiologyRaw water, product water and point of useBefore operation, then at least monthly
Chemical analysisPoint of useBefore operation, then at least every six months
Total dissolved solidsProduct waterWeekly; corrective action above 10 mg/L

The order requires stopping and compliant repeat testing when water fails, and makes the medical director accountable for water quality.7 Sampling and transport follow the accredited laboratory’s instructions.

A vertical gauge with green, amber and red bands: below the action level, at or above the action level with an investigate-and-correct label, and above the maximum limit with a stop-and-escalate label.
Figure 3. An action level warns before a limit is broken. Act at the amber line, not the red one.

The international standard for dialysis water, International Organization for Standardization (ISO) 23500-3, sets a maximum total viable microbial count of 100 CFU/mL and endotoxin of 0.25 endotoxin units (EU) per mL, with action levels usually set at half those values.8 Check these figures against the edition your facility follows. An action level is a warning: the water still meets the limit, but something is changing, and the time to find it is now.

RememberAct at the action level, not at the limit. A limit broken means patients may already have been exposed.

Collecting a sample that tells the truth

A sample is only as good as its collection. A contaminated bottle reports a false failure; a sample taken after flushing a heavily used outlet for a long time may report a false pass. Follow the laboratory’s instructions for flushing, disinfecting the outlet and bottle handling, because they differ between laboratories and the DOH order allows some local variation.7

Figure 4 · Sampling the reuse outlet

  1. PrepareSterile bottle from the laboratory, label filled in before sampling, hand hygiene, clean gloves.
  2. OutletFollow the laboratory’s instructions for disinfecting and flushing the outlet.
  3. CollectOpen the bottle only at the moment of filling. Do not touch the inside of the cap or the outlet tip.
  4. Close and labelOutlet name, date, time, collector.
  5. SendCold chain and timing as the laboratory requires.
  6. RecordLog the sample and later the result, and who reviewed it.
RememberLabel the bottle before you open it. A result without a location and time is useless.

When a result fails

Figure 5 · What you do with an abnormal water result

At or above an action levelReport the same day to the water technician and the head nurse. Increase monitoring as the program directs.
Above a limitStop reprocessing if the program requires, report to the medical director’s designate, and follow the corrective-action plan.
A result you cannot read or is missingTreat it as not proven. Ask before you start.
Patients with fever or chills during treatmentReport at once; it may be water, the dialyzer or the machine (Module 11).
Stop and reportIf the water log shows an unsigned failure, do not start the reprocessor. Ask the person responsible for water before any dialyzer is processed.

Back to the case: Monday morning

What does 14 mg/L tell you? The reverse osmosis system is passing more dissolved solids than it should. A membrane may be failing, or pretreatment may be exhausted. It does not tell you whether bacteria or endotoxin have risen, and it does not by itself say the water is dangerous today. It does say that the corrective action required above 10 mg/L has not been documented.7

May he start? Not until the person responsible for water confirms what was done and whether the water is fit for use. The twenty dialyzers can stay refrigerated within the time your procedure allows; if that time will run out, they follow the procedure for overdue dialyzers (Module 8), and patients receive new dialyzers.

Who must know? The water technician, the head nurse, and the medical director’s designate. He records the time he found the unsigned result, who he told, and the decision. That record protects the patients and him.

Skills checklist: Water sampling and log review

Practice with a sterile bottle on a training outlet, then review a set of synthetic water logs. Your preceptor rates each step as satisfactory, needs practice, or not observed. Steps marked Critical must all be satisfactory before sign-off. This checklist supports supervised learning; your unit’s own procedure and competency form are the authority. A printable version of every checklist is on the trainer page.

Placement activity. Collect one water sample from the reuse outlet under supervision, and follow it to its result and the review signature.

Check your understanding

Each visit draws 5 questions at random from this module’s 12-question bank. Choose the single best answer for each, then select Check answers; an explanation appears under the best answer. Select New set of questions for a different mix. A good target is 4 of 5 on two different sets. Nothing you select is sent anywhere; scoring happens in your browser.

Question 1 of 12Why must reprocessing water meet dialysis water quality?
Question 2 of 12What can cause fever and chills even when bacteria are dead?
Question 3 of 12What happened at Caruaru, Brazil, in 1996?
Question 4 of 12Under the DOH order, how often is product-water TDS measured?
Question 5 of 12Under the DOH order, how often at minimum is microbiology tested at the point of use?
Question 6 of 12What is an action level?
Question 7 of 12A water sample bottle was opened before it was labeled, and the outlet name was forgotten. What is the result worth?
Question 8 of 12Where does persistent bacterial contamination of a water loop usually come from?
Question 9 of 12You find an unsigned failed result in the water log at the start of a shift. What do you do?
Question 10 of 12In the Hanoi report, what were the bacteria found in patients mostly like?
Question 11 of 12According to a CDC review of dialysis outbreaks, most were:
Question 12 of 12Who is accountable for water quality under the DOH order?

Read next on this site

Glossary & abbreviationsTalahulugan at mga daglatTalaan sa mga pulong ug daglatTalatinigan ampo reng daglat terms used on this page

Abbreviations

AAMI
Association for the Advancement of Medical Instrumentation.
CDC
Centers for Disease Control and Prevention.
CFU
Colony-forming units.
DOH
Department of Health — Philippines.
EU
Endotoxin units.
ISO
International Organization for Standardization.
TDS
Total dissolved solids.

Terms in plain words

Product water
Water leaving the treatment system, usually after reverse osmosis.
Point of use
An outlet where water enters a process that touches patients.
Endotoxin
A toxin from bacterial walls that causes fever and chills.
Biofilm
A slimy layer of bacteria on pipe walls that resists disinfection.
Action level
A value below the limit that triggers investigation and correction.
Dead leg
A pipe section where water does not flow, so bacteria can grow.
Reverse osmosis
Purification by forcing water through a membrane that holds back dissolved solids.
ReferencesMga SanggunianMga TinubdanReng Reperensya 8 sources
  1. Association for the Advancement of Medical Instrumentation (AAMI). (2008). Reprocessing of hemodialyzers (ANSI/AAMI RD47:2008/(R)2013).
  2. Jackson, B. M., Beck-Sague, C. M., Bland, L. A., Arduino, M. J., Meyer, L., & Jarvis, W. R. (1994). Outbreak of pyrogenic reactions and gram-negative bacteremia in a hemodialysis center. American Journal of Nephrology, 14(2), 85–89. https://doi.org/10.1159/000168694
  3. Archibald, L. K., Khoi, N. N., Jarvis, W. R., Reller, L. B., Cam, P. D., Thu, T. A., & Hung, N. V. (2006). Pyrogenic reactions in hemodialysis patients, Hanoi, Vietnam. Infection Control and Hospital Epidemiology, 27(4), 424–426. https://doi.org/10.1086/503347
  4. Roth, V. R., & Jarvis, W. R. (2000). Outbreaks of infection and/or pyrogenic reactions in dialysis patients. Seminars in Dialysis, 13(2), 92–96. https://doi.org/10.1046/j.1525-139x.2000.00027.x
  5. Jochimsen, E. M., Carmichael, W. W., An, J. S., Cardo, D. M., Cookson, S. T., Holmes, C. E., Antunes, M. B., de Melo Filho, D. A., Lyra, T. M., Barreto, V. S., Azevedo, S. M., & Jarvis, W. R. (1998). Liver failure and death after exposure to microcystins at a hemodialysis center in Brazil. New England Journal of Medicine, 338(13), 873–878. https://doi.org/10.1056/NEJM199803263381304
  6. Ministry of Health Malaysia. (2026). National haemodialysis quality standards (3rd ed.). https://www.moh.gov.my/images/04-penerbitan/penerbitan-klinikal/perkhidmatan-perubatan/National_Haemodialysis_Standards_2026_3rd_Edition.pdf
  7. Department of Health (DOH). (n.d.). Guidelines on the monitoring of water used for dialysis. https://elibrary.judiciary.gov.ph/thebookshelf/showdocs/10/69036
  8. International Organization for Standardization (ISO). (2024). Preparation and quality management of fluids for haemodialysis and related therapies. Part 3: Water for haemodialysis and related therapies (ISO 23500-3:2024).
Dr. William Gregory M. Rivero, MD

William Gregory Rivero, MD, FPCP, DPSN

Internal Medicine · Nephrology · Nutrition · Philippines · PRC 0105184

Educational study material. Completing it is an Academy learning record only. It is not a DOH, PSN, PhilHealth, TESDA or PRC credential, it does not authorize anyone to release a dialyzer, and it does not replace the unit’s approved reuse program, the device and germicide instructions for use, or supervised practice.

QR code — scan to save Dr. Rivero's contact info

Scan and save