Renal Technician Program · Module 3 of 12

Dialysate Side I: Water Inlet, Heating, and Degassing

How the machine regulates, warms, and degasses incoming water, how temperature is verified, and how to troubleshoot these stages.

PublishedNailathalaGipatikPepalwal: ReferencesMga SanggunianMga TinubdanReng Reperensya: 1 Audience: Biomedical equipment technicians working in or for hemodialysis centers · chief technicians, head nurses, and medical directors Read timeOras ng pagbasaOras sa pagbasaOras ning pamamasa:
A technician holding a reference thermometer probe at a dialysate sample point on a machine.
Renal Care Academy›Hemodialysis Center Biomedical Technician›Module 3 of 12 Study series for dialysis biomedical technicians
🛠️

Study series for dialysis biomedical technicians

This module is study material for biomedical technicians in dialysis. It explains how machines work and why maintenance steps exist; it is not a service manual. Repairs, calibrations, and safety tests are done only by people trained and authorized for that machine model, using the manufacturer’s service manual, approved parts, and calibrated test equipment. Every case is fictional.

After this module you should be able to

Pre-testBefore you read, answer the 5 self-assessment questions and note your score. After the module, select New set of questions and compare. The difference is what you learned; the questions you still miss are what to reread.

The big picture, in plain words

In plain wordsThink of preparing water for a baby’s bath. You open the tap to a steady flow, warm the water to body temperature, and let it settle before the baby goes in. Before the machine can mix dialysate, it does the same: it steadies the water pressure, warms the water to body temperature, and pulls out the dissolved air. These steps are unglamorous, and they cause a large share of everyday alarms. A machine that warms slowly, or that shows bubbles in its lines, is telling you about scale, a weak pump, or the water supply.

The picture has a limit. At a bath, you test the water with your elbow and adjust as you go. The machine cannot feel. It trusts a temperature sensor, and it needs a second, independent sensor to catch the first one when it drifts. Your reference thermometer is the elbow the machine does not have.

Figure 1 · Water preparation inside the machine

  1. InletTreated water enters through a filter and a pressure regulator.
  2. Heat exchangerOutgoing used dialysate gives its warmth to the incoming water.
  3. HeaterRaises the water to the set temperature.
  4. DegassingA pump lowers the pressure so that dissolved air comes out and is vented.
  5. Ready for mixingWarm, bubble-free water goes on to the concentrate pumps.

Figure 2 · Key terms for this module

Pressure regulatorA valve that keeps inlet pressure steady.
Heat exchangerA device that passes heat from outgoing fluid to incoming water without mixing them.
DegassingRemoving dissolved air.
Control sensorThe sensor the machine uses to regulate.
Protective sensorAn independent sensor that triggers an alarm and a safe state.
Reference thermometerA calibrated thermometer used to check the machine.

Case: Slow to warm on cold mornings

It is a cool week in January. At the five-thirty start, machine 4 takes twenty minutes longer than its neighbors to pass its start-up checks. Twice this week it has alarmed for low temperature in the first hour of treatment. The charge nurse, Ms. Teresita Ramos, asks the biomedical technician, Mr. Noel Mendoza, to look at it.

Mr. Mendoza pulls the records. Machine 4 was last descaled three months ago. The other machines are descaled monthly. The water room also reports that the feed water is several degrees colder than last month.

Before reading on, decide:

Keep your answers in mind. The lesson gives you what you need, and the case review near the end shows one sound response.

The water inlet

The machine expects water within a stated pressure and temperature range. You will find the figures in the instructions for use. Outside that range, the machine alarms or cannot keep up.

PartJobWhat goes wrong
Inlet hose and connectorBrings water from the wall outletKinks; biofilm in old hoses; leaks at the connector
Inlet filter or strainerStops particlesPlugs gradually; low-water alarms at peak demand
Inlet valveAdmits water when the machine calls for itSticks; leaks through when closed
Pressure regulatorHolds a steady pressure for the hydraulicsDrifts; the diaphragm wears; unstable flow downstream
Heat exchangerPre-warms incoming water with heat from used dialysateScale lowers efficiency; an internal leak would cross-connect fresh and used fluid

When several machines alarm for low water at the same moment, the cause is in the supply they share. When only one does, look at its own hose, filter, and regulator. That one question, one machine or several, saves hours.

Preparing the water: heating and degassing

Water enters the machine cold and at a pressure that rises and falls with the plant. A regulator steadies the pressure, and a heater brings the water to roughly 35 to 37 degrees Celsius. Temperature matters to the patient in both directions. Dialysate that is too hot damages red blood cells. Dialysate that is too cold makes the patient shiver. For this reason machines measure temperature with more than one sensor, and they stop dialysate from reaching the dialyzer if it leaves the allowed range, with protection required before it exceeds about 41 degrees Celsius.1

Schematic of water inlet, heating, and degassing in a dialysis machine.
Figure 3. From inlet to degassed warm water. Schematic of water inlet, heating, and degassing in a dialysis machine.
Three effects of air bubbles in dialysate.
Figure 4. Why bubbles matter. Three effects of air bubbles in dialysate.
Clean heater element beside a scaled one.
Figure 5. A scaled heater. Clean heater element beside a scaled one.

Cold water also holds dissolved air. You have seen this at home: bubbles form on the inside of a pot long before the water boils. In the machine, those bubbles collect on conductivity sensors and give false readings. They disturb the balancing system, and they cover part of the dialyzer membrane. The degassing stage holds the water under low pressure so that the air comes out of solution and is vented before the concentrates are added.

Figure 6 · When heating or degassing goes wrong

Heater scaled or failingSlow warm-up and temperature alarms.
Temperature sensor driftedThe displayed temperature differs from a reference thermometer.
Degassing pump weak or leakingBubbles in the dialysate lines, unstable conductivity, and fluid balance errors.

Verifying dialysate temperature

The machine’s display shows what its own sensor believes. Verification asks a calibrated instrument the same question, under conditions where the answer can be trusted.

Figure 7 · A temperature verification

  1. ReferenceA calibrated thermometer with a valid certificate.
  2. Steady stateThe machine in dialysis mode, with stable flow and temperature.
  3. Sample pointAs the manual specifies, usually at the dialyzer connectors with a test adapter.
  4. WaitUntil the reference reading stops changing.
  5. CompareThe reference against the display, at one or more set temperatures as the manual states.
  6. DecideWithin tolerance, record. Outside, adjust by the manual, then repeat.
FindingLikely meaning
Display and reference agreeThe control sensor is accurate
Display reads higher than the referenceDialysate is cooler than shown; patients may feel cold
Display reads lower than the referenceDialysate is warmer than shown; check the protective sensor urgently
Control and protective sensors disagree with each otherOne has drifted; the machine may alarm intermittently
Reading swings up and downUnstable flow, air, or failing heater control

Notice which direction is dangerous. A display that reads low hides dialysate that is too warm, and warm dialysate damages red blood cells. Next, prove the protection: by the manual’s method, show that a temperature beyond the limit produces an alarm and bypass. Typical set temperatures are around 35 to 37 degrees Celsius, and the standard requires protection before dialysate exceeds about 41 degrees Celsius.1

Troubleshooting inlet, heating, and degassing

Work from the simple and the shared toward the specific and the expensive. Most of the “check first” column below needs no tools.

SymptomCheck firstThen
Low water or inlet pressure alarmIs it one machine or several? Is the hose kinked? Is the wall outlet fully open?Inlet filter, inlet valve, pressure regulator
Slow to reach temperatureInlet water temperature; descaling historyHeater resistance, heat exchanger scale, power to the heater
Low temperature alarms during treatmentSame as above; the dialysate flow settingSensor verification
High temperature alarmIs flow reduced or stopped? A heater with no flow overheats quicklyHeater control, sensors, thermal cut-out
Patients feel cold, display normalVerify with a reference thermometerSensor calibration
Bubbles in the dialysate lines; unstable conductivityInlet connections drawing air; water unusually coldDegassing pump pressure, restrictor, vent
Fluid balance errors with bubbles presentDegassingFluid balance test after repair
RememberScale is the common thread. It insulates heaters, coats sensors, blocks restrictors, and jams valves. Before you replace any part on a machine with repeated heating or degassing trouble, check its descaling history.

Real situations: problems and solutions

These are fictional but typical. For each one, cover the answer, decide what you would do, then compare. Where a step depends on local rules, your unit’s procedure and your supervisor’s direction come first.

Dialysate feels cool

What you see
Three patients on the same machine this week have asked the nurse for a blanket. The display shows a normal temperature.
What is likely going on
The temperature sensor may have drifted, so the display is wrong.
What to do
Take the machine out of service and verify the temperature with a reference thermometer, by the manual’s method.
Why it matters
Patients are an independent sensor. When they and the display disagree, check the display.

Slow to reach temperature

What you see
During the rainy season, one machine takes far longer than the others to warm up at the six o’clock start.
What is likely going on
The heater may be scaled or failing, or the incoming water may be colder than the design allows.
What to do
Check the descaling history, run a descale by the manual if one is due, and measure heater performance as the manual describes.
Why it matters
Scale works like insulation around the heater. It also flakes off into valves and sensors.

Low water alarm on one station

What you see
At shift start, only the machine at station 3 alarms for low water. The machines on either side of it are running normally.
What is likely going on
The supply is adequate, because the neighbors are fine. Something particular to station 3 restricts flow: a kinked hose, a half-closed outlet valve, or a plugged inlet filter.
What to do
Check the hose and the wall valve, then the machine’s inlet filter. Tell the water technician if the outlet itself is at fault.
Why it matters
“One machine or several?” is the question that separates supply problems from machine problems.
Field notes
  • Check the descaling history before you replace any heating or degassing part.
  • Verify temperature with a reference thermometer, never by touch.
  • A display that reads lower than the true temperature is the dangerous direction, because the dialysate is hotter than it shows.
  • Look at the inlet hoses and filters at every visit.

Back to the case: Slow to warm on cold mornings

Two things that would make it slower to warm. Three months without descaling, so its heater and heat exchanger are probably coated with scale. And colder feed water, which exposes that weakness: a clean heater copes, a scaled one falls behind.

What to check first, and how. Start with the simple things. Confirm the descaling record, run a descale by the manual, and inspect the inlet filter. Then measure the heater by the manual, and verify temperature against a reference thermometer.

What to verify before it returns to use. Record the as-found and as-left temperatures. Prove the high-temperature alarm. Run disinfection and confirm a negative residual test. Put the machine back on the monthly descaling schedule, and find out why it dropped off the schedule in the first place. Then tell Ms. Ramos what you found and what was done.

Skills checklist: Temperature verification against a reference thermometer

Materials: an out-of-service machine, its service manual, a calibrated reference thermometer with test adapter, and a service record. Your preceptor rates each step as satisfactory, needs practice, or not observed. Steps marked Critical must all be satisfactory before sign-off. This checklist supports supervised learning; your unit’s own procedure and competency form are the authority. A printable version of every checklist is on the trainer page.

Placement activity. Compare the descaling records of all machines in the unit and list any that are off schedule.

Check your understanding

Each visit draws 5 questions at random from this module’s 8-question bank. Choose the single best answer for each, then select Check answers; an explanation appears under the best answer. Select New set of questions for a different mix. A good target is 4 of 5 on two different sets. Nothing you select is sent anywhere; scoring happens in your browser.

Question 1 of 8Why is water degassed before concentrates are added?
Question 2 of 8Why must dialysate never be allowed to overheat?
Question 3 of 8What does the heat exchanger do?
Question 4 of 8Six machines alarm for low water at shift start. Where is the cause most likely?
Question 5 of 8The display reads lower than a reference thermometer. Why is this the more worrying direction?
Question 6 of 8A heater has power but water is not flowing past it. What happens?
Question 7 of 8One machine warms slowly while its neighbors are normal. What do you check before replacing parts?
Question 8 of 8Bubbles in the dialysate lines and unstable conductivity suggest a problem with which stage?

Read next on this site

Glossary & abbreviationsTalahulugan at mga daglatTalaan sa mga pulong ug daglatTalatinigan ampo reng daglat terms used on this page

Terms in plain words

Strainer
A coarse filter at the water inlet.
Restrictor
A narrow passage used to create low pressure for degassing.
Steady state
The condition when readings have stopped changing.
ReferencesMga SanggunianMga TinubdanReng Reperensya 1 source
  1. International Electrotechnical Commission (IEC). (2018). Medical electrical equipment. Part 2-16: Particular requirements for basic safety and essential performance of haemodialysis, haemodiafiltration and haemofiltration equipment (IEC 60601-2-16:2018).
Dr. William Gregory M. Rivero, MD

William Gregory Rivero, MD, FPCP, DPSN

Internal Medicine · Nephrology · Nutrition · Philippines · PRC 0105184

Educational study material for biomedical technicians. Does not replace the manufacturer’s service manual and training, approved facility procedures, applicable electrical and medical device regulations, or supervised practice.

QR code — scan to save Dr. Rivero's contact info

Scan and save