Renal Technician Program · Module 2 of 12

Electrical and Electronic Fundamentals

Voltage, current, and resistance in plain terms; power, fuses, and earth; the sensors and actuators in a machine; and safe use of a multimeter.

PublishedNailathalaGipatikPepalwal: ReferencesMga SanggunianMga TinubdanReng Reperensya: 1 Audience: Biomedical equipment technicians working in or for hemodialysis centers · chief technicians, head nurses, and medical directors Read timeOras ng pagbasaOras sa pagbasaOras ning pamamasa:
A technician holding multimeter probes over a training circuit board on an antistatic mat, wrist strap on.
Renal Care Academy›Hemodialysis Center Biomedical Technician›Module 2 of 12 Study series for dialysis biomedical technicians
🛠️

Study series for dialysis biomedical technicians

This module is study material for biomedical technicians in dialysis. It explains how machines work and why maintenance steps exist; it is not a service manual. Repairs, calibrations, and safety tests are done only by people trained and authorized for that machine model, using the manufacturer’s service manual, approved parts, and calibrated test equipment. Every case is fictional.

After this module you should be able to

Pre-testBefore you read, answer the 5 self-assessment questions and note your score. After the module, select New set of questions and compare. The difference is what you learned; the questions you still miss are what to reread.

The big picture, in plain words

In plain wordsThink of an aircon in a sari-sari store. A thermostat feels the room, a small board decides, and the compressor runs until the room is cool. Every function of a dialysis machine is a loop like that: a sensor measures something, a control board decides, and an actuator, meaning a pump, valve, or heater, does something about it. Once you can see those loops, the inside of the machine stops being a box of wires and becomes a set of small systems you can understand one at a time.

The picture has a limit. If the aircon’s thermostat fails, the store gets warm and someone complains. A dialysis machine cannot wait for a complaint. So it adds what the aircon lacks: a second, independent sensor that watches the loop and forces a safe state if the loop goes wrong.

Figure 1 · One control loop

  1. SensorMeasures temperature, pressure, conductivity, flow, or position.
  2. Control boardCompares the reading with the target.
  3. ActuatorA heater, pump, or valve changes the process.
  4. ResultThe reading moves toward the target.
  5. Protective sensorIndependently checks the result and can force a safe state.

Figure 2 · Key terms for this module

VoltageThe electrical push, measured in volts.
CurrentThe flow of electricity, measured in amperes.
ResistanceOpposition to flow, measured in ohms.
FuseA deliberate weak link that melts when current is too high.
Protective earthThe wire that carries fault current safely away from the metal case.
SensorA device that turns a physical quantity into an electrical signal.
ActuatorA device that turns an electrical signal into movement, flow, or heat.
ContinuityAn unbroken electrical path.

Case: The fuse that keeps blowing

On a Saturday evening, after the last shift, machine 8 blows its main fuse during heat disinfection. The junior technician on duty, Mr. Jomar Aquino, replaces it. It blows again. He then fits a fuse of a higher rating “so it will stop tripping.” The machine completes disinfection.

On Monday morning, the chief technician, Mr. Ramon Cruz, opens the workshop and notices that the heater area of machine 8 smells of hot plastic.

Before reading on, decide:

Keep your answers in mind. The lesson gives you what you need, and the case review near the end shows one sound response.

Voltage, current, and resistance

Picture the water pipes of a house. Voltage is like the water pressure. Current is like the flow in the pipe. Resistance is like the narrowness of the pipe. More pressure pushes more flow, and a narrower pipe lets less through. The three are tied together by one rule.

Voltage = current × resistance (Ohm’s law)
Power = voltage × current
Illustration: a heater element of 30 ohms on a 230-volt supply draws about 230 ÷ 30 = 7.7 amperes and gives about 230 × 7.7 = 1,770 watts of heat.

Most electrical faults are one of four kinds. Each changes resistance in a particular way, and each shows itself in a particular way.

FaultIn electrical termsWhat you see
Open circuitA break: resistance is infinite, and no current flowsA pump or valve that does nothing; a sensor reading stuck at the end of its scale
Short circuitAn unintended low-resistance path, with very high currentA blown fuse, a tripped breaker, a burning smell
High-resistance jointA loose or corroded connectionIntermittent faults, heat at the connector, readings that jump
Leakage to earthCurrent finding a path to the case, often through moistureTripped earth-leakage protection; a failed electrical safety test

Dialysis machines keep water and electricity in one cabinet, and moisture is behind a large share of their electrical faults. The sequence is slow and predictable. A small leak drips on a connector. The metal corrodes. Resistance rises, the joint heats, and eventually current leaks to the case. When you fix an electrical fault, look for water.

The power path, fuses, and protective earth

Figure 3 · From the wall to the electronics

  1. OutletLive, neutral, and protective earth.
  2. Power cord and plugThe earth pin is the longest, so it connects first.
  3. Mains switch and fusesCut power and protect against overload.
  4. Power supplyConverts mains voltage to the low voltages the boards use.
  5. High-power partsThe heater and some pumps run at mains voltage.
  6. Low-voltage partsBoards, sensors, and small valves.

The manufacturer chooses each fuse for three things: its current rating, its speed (fast or slow-blow), and its voltage rating. Replace a fuse only with the exact type in the manual. A fuse that blows once may simply have aged. A fuse that blows again has found a fault.

Mains power path into a dialysis machine with the protective earth bonded to the case.
Figure 4. Power path and protective earth. Mains power path into a dialysis machine with the protective earth bonded to the case.
Stop and reportNever fit a fuse of a higher rating, and never bridge a fuse. The fuse protects the wiring from fire and the patient from a fault. If it blows twice, the machine is out of service until the cause is found.

Protective earth is a separate idea from the fuse. The metal case is connected to earth. If a live wire ever touches the case, the current runs down the earth wire, the fuse blows, and the case never becomes dangerous to touch. Now consider a broken earth connection. The same fault makes the case live, and nothing trips. A person who touches it becomes the path to earth. This is why the earth path is tested on a schedule, and why Module 11 exists.1

Much of this can be found before any meter is connected. Look first.

Visual checkReject if
PlugCracked, pins bent or loose, or signs of heat
CordCut, crushed, taped, or stiff; strain relief pulled out
Inlet socket on the machineLoose or discolored
Fuse holdersWrong fuse fitted, or foil or wire bridging
Inside the cabinetWater marks, corrosion, or a burnt smell

The sensors and actuators in a dialysis machine

Every loop in the machine begins with a sensor and ends with an actuator. Learn each one by asking three questions: what does it measure or do, how does it work, and how does it fail. The way it works usually tells you the way it fails.

Block diagram of a control loop with a sensor, controller, actuator, and an independent protective sensor.
Figure 5. Sensor, controller, actuator. Block diagram of a control loop with a sensor, controller, actuator, and an independent protective sensor.
SensorMeasuresHow it works, brieflyHow it fails
Temperature sensorDialysate temperatureIts electrical resistance changes with temperatureDrift; a scale coating slows its response
Conductivity cellDialysate conductivityPasses a small current between electrodes in the fluidScale or air bubbles on the electrodes; drift
Pressure transducerBlood-side and dialysate-side pressuresA thin diaphragm flexes and changes an electrical signalDrift; a wet or blocked protector; diaphragm damage
Air detectorAir in the return lineUltrasound passes well through liquid and poorly through airDirt, misalignment, or the wrong tubing
Blood leak detectorBlood in used dialysateBlood absorbs light passed through the fluidA dirty window; bubbles
Flow and level sensorsDialysate flow; chamber levelsVarious: turbines, floats, opticalSticking; scale
Position switchesDoors, covers, pump rotorMagnetic or mechanicalMisalignment; deliberate bypass
ActuatorDoesHow it fails
Solenoid valvesOpen and close fluid paths, including bypassStick open or closed with scale; the coil burns out; the seal leaks
Gear and piston pumpsMove dialysate, concentrates, and ultrafiltrateWear lowers delivery; seals leak; air locks
Blood pump motorTurns the rollers at a set speedSpeed error; worn rollers or springs
HeaterWarms incoming waterScale; an open element; a failed thermal cut-out
Venous clampCloses the return line on alarmA weak spring or solenoid; does not fully occlude
Syringe (heparin) pumpPushes a syringe at a set rateWrong syringe size setting; drive slip
RememberMost machines have two sensors for each safety-critical quantity: one to control and one to protect. When you calibrate, verify both.

Using a multimeter safely

A multimeter answers one question at a time. Your work is to ask the right question, on a circuit in the right state, with a meter you know is working.

Multimeter with safe-use reminders.
Figure 6. Multimeter safely. Multimeter with safe-use reminders.

Figure 7 · Every measurement

  1. DecideWhat you expect to read, and whether the circuit should be live.
  2. SelectFunction and range, before you connect the probes.
  3. Check the meterTest it on a known source first.
  4. ConnectOne hand where possible; probes with insulated tips.
  5. Read and compareAgainst what you expected.
  6. RemoveProbes first, then change the function.
FunctionCircuit stateTypical use
AC voltageLiveIs mains reaching the power supply or the heater?
DC voltageLiveAre the power supply outputs correct at the test points in the manual?
Continuity (beeper)Power off and unpluggedIs a fuse, wire, or switch unbroken?
ResistancePower off and unpluggedA heater element, valve coil, or sensor, against the manual’s value
Stop and reportResistance and continuity are measured only with the machine unplugged. Live measurements are made only when the manual calls for them, by someone trained to work on live equipment.

Circuit boards are damaged by static electricity that you cannot feel. Wear a grounded wrist strap, work on an antistatic mat, hold boards by their edges, and carry them in antistatic bags. Boards are usually replaced as whole units, not repaired on site. Many are tied to a software version, so check the manual before you swap one, and record the board’s serial number and software version in the service record.

Real situations: problems and solutions

These are fictional but typical. For each one, cover the answer, decide what you would do, then compare. Where a step depends on local rules, your unit’s procedure and your supervisor’s direction come first.

Dead machine, live outlet

What you see
At 6:15 in the morning, machine 5 shows a blank screen when you switch it on. You plug a desk lamp into the same outlet, and the lamp lights.
What is likely going on
Power reaches the wall but not the electronics. The break lies somewhere between the plug and the power supply: the cord, the inlet, a fuse, the switch, or the power supply itself.
What to do
Unplug the machine. Check the cord and the inlet, then test the fuses for continuity and for the correct rating. If a fuse has blown, find out why before you replace it.
Why it matters
Working from the wall inward, one step at a time, finds the break quickly and keeps you away from live parts.

A valve that hums but does not move

What you see
During a bench test, a solenoid valve buzzes each time it is energized, but no fluid flows past it. The machine is due for descaling next week.
What is likely going on
The coil is receiving power, but the plunger is stuck. Scale is the usual cause.
What to do
Descale and inspect the valve by the manual. Replace it if it does not free. Then ask why scale reached it.
Why it matters
A stuck valve can mean that bypass does not work at the moment it is needed.

Board swapped, new alarms

What you see
On Monday, after a control board is replaced with one taken from a retired machine, the unit shows alarms that nobody on the floor has seen before.
What is likely going on
The donor board carries a different software version or a different configuration.
What to do
Take the machine out of service. Check the versions against the manual, and obtain the correct board or software from the manufacturer.
Why it matters
Boards are not interchangeable just because they fit. A board that fits but does not match can make a machine behave in ways no one expects.
Field notes
  • When you fix an electrical fault, look for water nearby.
  • Replace a fuse with the same type and the same rating, every time.
  • Select the meter function before you connect the probes.
  • Write board serial numbers and software versions in the service record.

Back to the case: The fuse that keeps blowing

What was the fuse telling you? That the heater circuit drew more current than the design allows, during heat disinfection, when the heater works hardest. The likely causes are a heater element breaking down, scale forcing the element to run hot, or moisture making a path to earth.

What is the danger of a higher-rated fuse? It lets the excess current keep flowing. The wiring and connectors then overheat, and that is the smell of hot plastic Mr. Cruz noticed. The next step is a fire or a live case.

What should have been done? Take the machine out of service after the second blown fuse. Fit only the specified fuse. Measure the heater element’s resistance and its insulation to earth by the manual, and look for water. Repair, run an electrical safety test, and record it all. Machine 8, still carrying the wrong fuse, comes out of service now. Mr. Cruz should also talk with Mr. Aquino calmly, so that the reason for the rule stays with him.

Skills checklist: Multimeter basics on a de-energized training board

Materials: a multimeter with a valid calibration label, a training board or an unplugged out-of-service machine, the service manual, a wrist strap, and an antistatic mat. Your preceptor rates each step as satisfactory, needs practice, or not observed. Steps marked Critical must all be satisfactory before sign-off. This checklist supports supervised learning; your unit’s own procedure and competency form are the authority. A printable version of every checklist is on the trainer page.

Placement activity. With a senior technician, find the control sensor and the protective sensor for temperature and for conductivity on one machine model.

Check your understanding

Each visit draws 5 questions at random from this module’s 8-question bank. Choose the single best answer for each, then select Check answers; an explanation appears under the best answer. Select New set of questions for a different mix. A good target is 4 of 5 on two different sets. Nothing you select is sent anywhere; scoring happens in your browser.

Question 1 of 8In the water picture of electricity, what is voltage most like?
Question 2 of 8A heater element of 23 ohms is connected to 230 volts. About how much current flows?
Question 3 of 8A fuse blows a second time after replacement. What is correct?
Question 4 of 8What does the protective earth connection do?
Question 5 of 8Which measurement must only be made with the machine unplugged?
Question 6 of 8A sensor reading is jumping and a connector feels warm. What kind of fault is likely?
Question 7 of 8Why do machines have two sensors for safety-critical quantities?
Question 8 of 8How should a circuit board be handled?

Read next on this site

Glossary & abbreviationsTalahulugan at mga daglatTalaan sa mga pulong ug daglatTalatinigan ampo reng daglat terms used on this page

Terms in plain words

Ampere
The unit of electrical current.
Ohm
The unit of electrical resistance.
Solenoid
A coil that moves a plunger when powered; used in valves and clamps.
Thermal cut-out
A switch that opens if a heater gets too hot.
Rating plate
The label on a machine giving model, serial number, and electrical ratings.
ReferencesMga SanggunianMga TinubdanReng Reperensya 1 source
  1. International Electrotechnical Commission (IEC). (2020). Medical electrical equipment. Part 1: General requirements for basic safety and essential performance (IEC 60601-1, Edition 3.2).
Dr. William Gregory M. Rivero, MD

William Gregory Rivero, MD, FPCP, DPSN

Internal Medicine · Nephrology · Nutrition · Philippines · PRC 0105184

Educational study material for biomedical technicians. Does not replace the manufacturer’s service manual and training, approved facility procedures, applicable electrical and medical device regulations, or supervised practice.

QR code — scan to save Dr. Rivero's contact info

Scan and save