Dialysate Side II: Proportioning, Conductivity, and Bypass

How dialysate is mixed and checked, how conductivity is measured and verified, what bypass does, and how to troubleshoot conductivity alarms.

PublishedNailathalaGipatikPepalwal: Audience: Biomedical equipment technicians working in or for hemodialysis centers · chief technicians, head nurses, and medical directors Read timeOras ng pagbasaOras sa pagbasaOras ning pamamasa:
A technician dipping a reference conductivity meter probe into a dialysate sample beside a machine display.
Renal Care Academy›Hemodialysis Center Biomedical Technician›Module 4 of 12 Study series for dialysis biomedical technicians
🛠️

Study series for dialysis biomedical technicians

This module is study material for biomedical technicians in dialysis. It explains how machines work and why maintenance steps exist; it is not a service manual. Repairs, calibrations, and safety tests are done only by people trained and authorized for that machine model, using the manufacturer’s service manual, approved parts, and calibrated test equipment. Every case is fictional.

After this module you should be able to

Pre-testBefore you read, answer the 5 self-assessment questions and note your score. After the module, select New set of questions and compare. The difference is what you learned; the questions you still miss are what to reread.

The big picture, in plain words

In plain wordsThink of the pharmacist at the corner botika who mixes a child’s syrup from a powder. She adds water to the line on the bottle, shakes it, and checks the label before handing it over. At the mixing stage, the machine is that pharmacist. It combines water with an acid concentrate and a bicarbonate concentrate in exact proportion, then checks its own work by measuring how well the mixture conducts electricity. If the check fails, it sends the fluid to the drain instead of to the patient. Your work is to make sure that check is true.

The picture has a limit. The pharmacist reads the label and knows which medicine is in the bottle. The machine cannot read labels. Its check sees only the total amount of dissolved salt, so it trusts that the right concentrates were connected and that it was configured for them.

Figure 1 · Key terms for this module

ProportioningMixing water and concentrates in a fixed ratio.
Concentrate systemThe particular ratio a machine is configured for.
Conductivity cellThe sensor that measures conductivity.
Temperature compensationCorrecting a conductivity reading to a standard temperature.
BypassDiverting dialysate to the drain when it is not safe.
Reference meterA calibrated conductivity meter used to check the machine.

Case: A machine from the other unit

On a Thursday, one of the unit’s machines breaks down. To cover the gap, a machine is borrowed from the network’s other center. It is the same model, but that center uses a different brand of concentrate.

On its first treatment, at eight in the morning, the conductivity will not stabilize, and the machine goes in and out of bypass. The nurse, Ms. Angelica Torres, finds the biomedical technician, Mr. Rodel Castillo, and asks him to “adjust the conductivity so it stops alarming.”

Before reading on, decide:

Keep your answers in mind. The lesson gives you what you need, and the case review near the end shows one sound response.

Proportioning and conductivity

Dialysate is made from three streams: treated water, acid concentrate, and bicarbonate concentrate. The machine mixes them in a fixed ratio, and that ratio depends on the concentrate system the machine is set up for. Ratio systems are not interchangeable. The concentrates must match the machine’s configuration.

Schematic of proportioning with control and protective conductivity cells.
Figure 2. Mixing and checking dialysate. Schematic of proportioning with control and protective conductivity cells.
Two different concentrates producing the same conductivity reading.
Figure 3. The total, not the recipe. Two different concentrates producing the same conductivity reading.

Figure 4 · Two ways machines proportion

Fixed-volume (volumetric)Pumps or chambers deliver set volumes of each stream. Conductivity is then a check on the result.
Conductivity feedback (servo)Pumps are adjusted until conductivity reaches the target. A second, independent sensor guards against a fault in the first.

Why conductivity? Pure water carries almost no electricity. Every dissolved salt adds charged particles, and each one makes the fluid carry current a little better. Final dialysate usually reads around 13 to 15 millisiemens per centimeter, depending on the prescription.

The weakness follows from the same physics. Conductivity measures only the total. It cannot tell which salts are present, so a wrong concentrate can still produce a believable number. That is why operators read the label, and why your calibration of the sensor matters: all of the machine’s confidence in its mix rests on that one measurement.

Side-by-side illustration of diffusion and convection across a dialyzer membrane.
Figure 5. What correct dialysate is for: its composition sets the direction and speed of exchange across the membrane. A machine that mixes wrongly changes the treatment.
RememberThe conductivity alarm protects against a wrong total. It cannot protect against the wrong recipe. Check sensor accuracy against a reference meter, not against the machine’s own display.

How conductivity is measured and compensated

A conductivity cell passes a small alternating current between electrodes in the fluid. The more dissolved ions there are, the more easily current flows. Two things besides the salt content also change the reading, and both matter to you.

InfluenceEffectConsequence for your work
TemperatureConductivity rises by roughly 2 percent for each degree CelsiusMachines and meters correct the reading to a standard temperature. The machine and your reference meter must use the same reference temperature, or they will disagree while both are right
Deposits and air on the electrodesScale or bubbles reduce the contact area and lower the readingA scaled cell makes the machine add more concentrate than intended

Why does temperature matter so much? Warm ions move faster through water, so the same fluid carries current more easily when it is warm. The worked example shows how large that effect is.

Illustration. A dialysate reading 14.0 mS/cm at 25 °C will read about 14.0 × (1 + 0.02 × 12) = 17.4 mS/cm at 37 °C if no correction is applied. A disagreement of this size between two instruments nearly always means they are using different reference temperatures.

Machines have at least two cells in the final dialysate: one that controls or monitors the mix, and an independent one that protects. Some machines also measure after the first concentrate is added, to check each stage. The flow diagram shows where each cell sits.

Bypass: the safe state

When conductivity or temperature leaves its limits, a valve sends the dialysate straight to the drain, so it never reaches the dialyzer. Blood keeps circulating, but no exchange takes place. This is bypass. It is the machine’s way of saying, “I do not trust this fluid.”

Normal flow and bypass flow of dialysate compared.
Figure 6. Bypass: the safe state. Normal flow and bypass flow of dialysate compared.

Figure 7 · What happens in a conductivity alarm

  1. Sensor detectsThe reading is outside the alarm window.
  2. Bypass valve switchesDialysate goes to the drain, not to the dialyzer.
  3. Alarm soundsThe operator is alerted.
  4. Cause correctedFor example, an empty concentrate container is replaced.
  5. Reading stabilizesOnly then does dialysate return to the dialyzer.

When you test a machine, your work is to prove that this chain holds: that an out-of-range reading really does switch the valve. Use the method in the service manual.

Verifying conductivity with a reference meter

A verification is only as good as the reference you compare against. Before you trust your meter to judge the machine, prove the meter itself.

Figure 8 · A conductivity verification

  1. ReferenceA calibrated meter, checked that day against standard solution.
  2. Match settingsThe same reference temperature and units as the machine.
  3. Steady stateCorrect concentrates connected; the machine stable in dialysis mode.
  4. SampleFrom the point, and by the method, that the manual specifies; rinse the probe or cell with sample first.
  5. CompareThe reference against the display, for each cell the manual lists.
  6. DecideWithin tolerance, record. Outside, clean or descale, then calibrate by the manual and repeat.
Common errorResult
Meter not checked against standard solutionYou may adjust a correct machine to match a wrong meter
Different reference temperaturesA false disagreement of several percent
Sample allowed to cool or sitThe reading drifts; bicarbonate dialysate loses carbon dioxide
Probe not rinsed with sampleCarry-over from the last measurement
Measuring before the machine is stableMeaningless numbers
Wrong concentrate connected for the testThe machine appears out of calibration when it is not

Once the reading is verified, prove the protection. By the manual’s method, create an out-of-limit condition and confirm the alarm and bypass. Record the as-found and as-left values and the meter’s identity.

Stop and reportNever adjust conductivity calibration to silence an alarm. Calibration is changed only when a verified reference shows the sensor is wrong.

Troubleshooting conductivity alarms

Read the pattern before you touch the machine. Who is affected, when it started, and which way the reading is off usually point to the cause.

PatternLikely causeFirst checks
Several machines at onceShared concentrate supply or waterCentral tank level, batch, valve; water alarms
One machine, just after a container changePick-up tube out of the fluid, wrong product, or air drawn inContainer, label, pick-up tube and filter
One machine, low and unstableAir: poor degassing or a loose concentrate connectorBubbles in the lines; connectors; degassing
One machine, steadily off in one directionConcentrate pump wear, a scaled cell, or the wrong concentrate system configuredVerify with a reference; pump delivery test; configuration
Bicarbonate-stage conductivity low, final normal or highBicarbonate supply weak or cartridge exhaustedCartridge or batch; bicarbonate pump
Alarm appears after the machine was movedConfiguration does not match the concentrates at the new siteConfigured concentrate system against the labels
Drifts over weeks, then alarmsScale building on cells and in pumpsDescaling history

Real situations: problems and solutions

These are fictional but typical. For each one, cover the answer, decide what you would do, then compare. Where a step depends on local rules, your unit’s procedure and your supervisor’s direction come first.

Conductivity wanders

What you see
Through the morning shift, one machine’s conductivity reading drifts up and down by small amounts. Twice it crosses the limit and alarms.
What is likely going on
The usual causes are air from poor degassing, a concentrate pick-up wand that is drawing air, or scale on the conductivity cell.
What to do
Check the concentrate supply and connections first. Then look for bubbles in the dialysate lines. Then verify the cell against a reference meter.
Why it matters
Unstable conductivity means repeated bypass, and every bypass is lost treatment time.

Meter and machine disagree by a fifth

What you see
During a calibration round, your reference meter reads about a fifth higher than the machine on the same dialysate sample.
What is likely going on
A difference that large is almost never a calibration error. The meter is probably set to no temperature compensation, or to different units.
What to do
Before you touch the machine, check the meter’s compensation setting and units against the machine’s, and check the meter in standard solution.
Why it matters
Adjusting a correct machine to match a mis-set meter creates the very error you were trying to prevent.

Cartridge nearly empty

What you see
In the fourth hour of a long treatment, the bicarbonate-stage conductivity drifts slowly down and the machine alarms.
What is likely going on
A powder cartridge is running out, or channeling, so that water passes through without dissolving enough powder.
What to do
Tell the nurse; the cartridge is changed by the unit’s procedure. If the alarm recurs with fresh cartridges, check the bicarbonate pump and the cartridge holder.
Why it matters
Not every conductivity alarm is a machine fault. Rule out the supply first.
Field notes
  • Check your reference meter in standard solution on the day you use it.
  • Match the reference temperature and the units before you compare readings.
  • Never calibrate to silence an alarm.
  • When a machine moves to another site, check its concentrate configuration.

Back to the case: A machine from the other unit

What is the most likely explanation? The machine is configured for the concentrate system used at the other center. With this center’s concentrates, the fixed proportions produce a mixture the machine does not expect, so conductivity will not settle.

Why is adjusting the conductivity wrong? Adjusting the calibration to stop the alarm would teach the machine to accept dialysate of the wrong composition. That is exactly what the alarm exists to prevent. Mr. Castillo should explain this to Ms. Torres plainly. The machine comes off the floor, and the nurse arranges another machine for the patient. Mr. Castillo then checks the configured concentrate system against the labels in use here. He reconfigures it by the manual only if that is permitted and he is trained to do it. Then he verifies conductivity with a reference meter, proves bypass, disinfects, and records.

What to check on any machine that moves between sites. The concentrate configuration, the hepatitis isolation status, the software version, the maintenance status, and that the machine is entered in the receiving unit’s inventory.

Skills checklist: Conductivity verification and bypass proof

Materials: an out-of-service machine with the unit’s concentrates, its service manual, a reference conductivity meter, standard solution, and a service record. Your preceptor rates each step as satisfactory, needs practice, or not observed. Steps marked Critical must all be satisfactory before sign-off. This checklist supports supervised learning; your unit’s own procedure and competency form are the authority. A printable version of every checklist is on the trainer page.

Placement activity. List the concentrate system each machine in the unit is configured for and compare with the concentrates in the store.

Check your understanding

Each visit draws 5 questions at random from this module’s 8-question bank. Choose the single best answer for each, then select Check answers; an explanation appears under the best answer. Select New set of questions for a different mix. A good target is 4 of 5 on two different sets. Nothing you select is sent anywhere; scoring happens in your browser.

Question 1 of 8What does dialysate conductivity measure?
Question 2 of 8What is bypass?
Question 3 of 8Why can a wrong concentrate sometimes produce a normal conductivity reading?
Question 4 of 8Against what is a machine’s conductivity sensor verified?
Question 5 of 8By roughly how much does conductivity change per degree Celsius?
Question 6 of 8Your reference meter and the machine disagree by several percent, and both pass their own checks. What should you suspect first?
Question 7 of 8A nurse asks you to adjust conductivity so an alarm stops. What is correct?
Question 8 of 8A machine borrowed from another center will not stabilize conductivity. What do you check?

Read next on this site

Glossary & abbreviationsTalahulugan at mga daglatTalaan sa mga pulong ug daglatTalatinigan ampo reng daglat terms used on this page

Terms in plain words

Millisiemens per centimeter
The unit of conductivity used for dialysate.
Reference temperature
The temperature to which a conductivity reading is corrected, often 25 °C.
Pick-up tube
The tube that draws concentrate from its container.