Renal Technician Program · For trainers

Trainer and Preceptor Resource

Program structure, completion criteria, and printable skills checklists for the Hemodialysis Center Biomedical Technician series.

PublishedNailathalaGipatikPepalwal: Audience: Head nurses, chief technicians, educators, and preceptors Read timeOras ng pagbasaOras sa pagbasaOras ning pamamasa:
Renal Care Academy›Hemodialysis Center Biomedical Technician›Trainer and preceptor resource

Using the series to train biomedical staff

Use the twelve modules as guided study alongside the manufacturer’s training for each machine model and supervised bench work. A practical sequence is one module a week, each followed by its checklist on a machine that is out of clinical service.

Figure 1 · Suggested sequence

  1. Open the manualFind the flow diagram and the maintenance schedule for the model.
  2. Study the moduleRead, work the case, and take the self-assessment.
  3. DemonstrateA senior technician performs the task and explains each step.
  4. PracticeThe learner performs it under observation on an out-of-service machine.
  5. Sign offRecord competence per model on the facility’s form.

Proposed completion criteria

These are proposals for the facility to adapt.

Written examination. The end-of-course examination draws 30 questions at random from the question banks of all modules. A proposed pass mark is 80 percent, with missed questions reviewed against their modules before a second sitting. For a supervised sitting, the learner takes it on a unit device with the preceptor present and the preceptor records the score.

Skills checklists for every module

Print this page for a full set. Each checklist starts on a new sheet. Steps marked Critical must all be satisfactory before sign-off; a missed critical step is repeated with the preceptor. Replace any step that does not match your equipment or approved procedure.

Module 1: The Biomedical Technician’s Role and Safe Working

Safe preparation for service and formal release. Materials: an out-of-clinical-use machine, the service manual, tags, a lockout device, protective equipment, and blank service and release forms.

Observable stepSatisfactoryNeeds practiceNot observed
Confirms with the charge nurse that the machine is free and records its identity and assigned area.
CriticalApplies a visible out-of-service tag with date, reason, and name.
CriticalConfirms the machine has completed disinfection before opening it.
CriticalSwitches off, unplugs, closes the water supply, and applies lockout.
Wears gloves and eye protection and uses static protection when handling boards.
Locates the correct service manual and revision for the model.
CriticalStates which tests are required after the simulated repair, by the manual.
CriticalCompletes the service record and obtains acceptance from the charge nurse before removing the tag.

Learner ____________________   Preceptor ____________________   Date __________

Module 2: Electrical and Electronic Fundamentals

Multimeter basics on a de-energized training board. Materials: a multimeter with a valid calibration label, a training board or an unplugged out-of-service machine, the service manual, a wrist strap, and an antistatic mat.

Observable stepSatisfactoryNeeds practiceNot observed
CriticalConfirms the equipment is unplugged and locked out before resistance or continuity checks.
CriticalSelects function and range before connecting the probes.
Tests the meter on a known source.
CriticalChecks a fuse for continuity and confirms its rating against the manual.
Measures a heater element or valve coil resistance and compares with the manual.
Inspects the plug, cord, and inlet and states what would reject each.
Uses a wrist strap and handles a circuit board by its edges.
Identifies five sensors and five actuators on the machine using the flow diagram.

Learner ____________________   Preceptor ____________________   Date __________

Module 3: Dialysate Side I: Water Inlet, Heating, and Degassing

Temperature verification against a reference thermometer. Materials: an out-of-service machine, its service manual, a calibrated reference thermometer with test adapter, and a service record.

Observable stepSatisfactoryNeeds practiceNot observed
CriticalChecks the reference thermometer’s identity and calibration due date.
Brings the machine to steady dialysis conditions.
Measures at the point the manual specifies and waits for a stable reading.
CriticalRecords as-found display and reference values.
CriticalStates the tolerance and decides correctly whether adjustment is needed.
CriticalDemonstrates the high-temperature alarm and bypass by the manual’s method.
Inspects the inlet hose, filter, and connections.
Records the result with the instrument’s identity.

Learner ____________________   Preceptor ____________________   Date __________

Module 4: Dialysate Side II: Proportioning, Conductivity, and Bypass

Conductivity verification and bypass proof. Materials: an out-of-service machine with the unit’s concentrates, its service manual, a reference conductivity meter, standard solution, and a service record.

Observable stepSatisfactoryNeeds practiceNot observed
CriticalChecks the reference meter’s calibration date and verifies it against standard solution.
CriticalConfirms the meter and the machine use the same reference temperature and units.
CriticalConfirms the machine’s configured concentrate system matches the concentrates connected.
Samples by the manual’s method at steady state.
Records as-found readings for each conductivity cell.
CriticalStates the tolerance and decides correctly whether calibration is needed.
CriticalDemonstrates alarm and bypass for an out-of-limit condition by the manual’s method.
Records as-left readings with the meter’s identity.

Learner ____________________   Preceptor ____________________   Date __________

Module 5: Fluid Balance and Ultrafiltration Control

Ultrafiltration accuracy check with a measuring cylinder. Materials: an out-of-service machine, its service manual, the test connector, a measuring cylinder or calibrated scale, a timer, and a service record.

Observable stepSatisfactoryNeeds practiceNot observed
States which fluid balance design the machine uses.
Removes air from the loop before testing.
CriticalPerforms the tightness test by the manual and interprets the result.
Sets the ultrafiltration rate and time the manual specifies.
CriticalCollects and reads the volume correctly at eye level.
CriticalCalculates the percent error and compares it with the tolerance.
Calibrates or reports for repair if outside tolerance, then repeats the check.
Records as-found and as-left values and the measuring device used.
Inspects and replaces worn dialyzer connector seals.

Learner ____________________   Preceptor ____________________   Date __________

Module 6: The Blood Side: Pump, Pressures, and Heparin Pump

Pump occlusion and pressure verification. Materials: an out-of-service machine, its service manual, the unit’s tubing set, a calibrated pressure meter, a measuring cylinder, a timer, and the unit’s heparin syringes.

Observable stepSatisfactoryNeeds practiceNot observed
CriticalConfirms the tubing size setting matches the unit’s tubing set.
CriticalPerforms the occlusion test by the manual and adjusts if required.
Verifies rotor speed or flow delivery at the points the manual specifies.
Inspects rollers, springs, and guides for wear.
CriticalConfirms the cover switch stops the pump.
CriticalVerifies arterial and venous pressure readings against the reference at zero and across the range.
CriticalConfirms the heparin pump’s syringe configuration and checks delivery accuracy.
Confirms the hand crank is present and works.
Records as-found and as-left values with instrument identities.

Learner ____________________   Preceptor ____________________   Date __________

Module 7: Protective Systems and Alarm Testing

Functional tests of the air detector, clamp, and blood leak detector. Materials: an out-of-service machine, its service manual, a training tubing set, the manufacturer’s test fixtures or solutions, a calibrated pressure meter, and a service record.

Observable stepSatisfactoryNeeds practiceNot observed
Reads the test method and pass criteria in the manual before starting.
CriticalSets up with training tubing and confirms no patient connection.
CriticalPerforms the air detector test and confirms pump stop, clamp closure, and alarm.
CriticalPerforms the venous clamp holding test and records the pressure held.
CriticalCleans and tests the blood leak detector by the manual.
Confirms cover and door switches stop the pump.
Confirms the audible alarm and the power-failure alarm.
CriticalKeeps the machine out of service if any test fails, and states the next steps.
Records each result with instruments used.

Learner ____________________   Preceptor ____________________   Date __________

Module 8: The Preventive Maintenance Program

Building a maintenance calendar and running a visit. Materials: the equipment inventory, service manuals, a year planner, one maintenance kit, and a blank visit checklist.

Observable stepSatisfactoryNeeds practiceNot observed
Lists every machine with model, serial number, interval, and last visit date.
Calculates the number of visits due per year and spreads them across the calendar.
CriticalIdentifies any machine currently overdue and reports it.
CriticalChecks for open safety notices affecting each model.
Follows the visit order: prepare, receive, decontaminate, inspect, replace, verify, test, finish.
Asks the operators about problems before starting.
CriticalRecords as-found values before adjusting.
Updates the inventory with the visit date and next due date.
CriticalReleases the machine formally to the charge nurse.

Learner ____________________   Preceptor ____________________   Date __________

Module 9: Calibration and Test Equipment

A calibration with a traceable record. Materials: an out-of-service machine, its service manual, a reference conductivity and temperature meter with a valid certificate, standard solution, and a blank service record.

Observable stepSatisfactoryNeeds practiceNot observed
CriticalChecks the reference instrument’s identity and calibration due date before use.
Verifies the reference meter against standard solution.
Allows the machine to reach steady operation before measuring.
CriticalMeasures and records as-found conductivity and temperature.
CriticalStates the tolerance from the service manual and decides correctly whether adjustment is needed.
Adjusts by the manual’s procedure, if needed, and records as-left values.
CriticalCompletes the service record with machine identity, readings, instrument identity, and signature.
CriticalReports any as-found value far outside tolerance to the chief technician and charge nurse.

Learner ____________________   Preceptor ____________________   Date __________

Module 10: Disinfection, Descaling, and Machine Fluid Hygiene

Disinfection with residual test, hose check, and filter change. Materials: an out-of-service machine, its manual, approved agents, residual strips for those agents, protective equipment, a replacement dialysis fluid filter, and the machine’s record.

Observable stepSatisfactoryNeeds practiceNot observed
CriticalSelects the program and agent approved for the stated purpose.
Wears the protective equipment on the safety data sheet.
CriticalAllows the full cycle and rinse to complete.
CriticalPerforms the residual test with the correct strip at the stated outlet and time.
Inspects the inlet hose and concentrate connectors and states their replacement dates.
CriticalConfirms the drain hose has an air gap.
CriticalChecks the dialysis fluid filter’s due date and replaces it by the manual if due.
Records agent, program, times, residual result, and filter details.

Learner ____________________   Preceptor ____________________   Date __________

Module 11: Electrical Safety Testing and Facility Utilities

Electrical safety test with an analyzer. Materials: an out-of-service machine, its service manual, an electrical safety analyzer with a valid calibration, an outlet tester, and the test record.

Observable stepSatisfactoryNeeds practiceNot observed
CriticalCarries out the visual inspection in order and corrects or reports defects before testing.
CriticalChecks the analyzer’s calibration date and leads.
CriticalMeasures protective earth resistance while flexing the cord and compares with the limit.
CriticalPerforms the leakage or insulation tests by the manual’s method and compares with the limits.
Confirms the machine functions normally afterward.
CriticalTags the machine out of service if any test fails.
Records values, limits, analyzer identity, and the next due date, and labels the machine.
Checks the station outlet with an outlet tester and reports any fault to the facility electrician.
CriticalConfirms no extension leads or adapters are in use at the station.

Learner ____________________   Preceptor ____________________   Date __________

Module 12: Troubleshooting, Life Cycle, and Capstone

Fault-finding and return to service. Materials: an out-of-service machine with a simulated fault set by the preceptor, the service manual, test instruments, and blank service and release forms.

Observable stepSatisfactoryNeeds practiceNot observed
Collects the exact alarm text and the circumstances before touching the machine.
CriticalChecks simple and shared causes before opening the cabinet.
Uses the flow diagram and fault table to narrow the fault area.
Tests one idea at a time and states the reasoning.
CriticalRepairs by the manual with approved parts.
CriticalLists and performs the verification tests the repair requires.
CriticalDisinfects and passes a residual test if the fluid path was opened.
CriticalCompletes the service record and hands over in person to the charge nurse before removing the tag.

Learner ____________________   Preceptor ____________________   Date __________