Renal Technician Program · Module 9 of 12

Calibration and Test Equipment

Calibration and tolerance, as-found and as-left values, reference instruments and their care, and the instrument register.

PublishedNailathalaGipatikPepalwal: Audience: Biomedical equipment technicians working in or for hemodialysis centers · chief technicians, head nurses, and medical directors Read timeOras ng pagbasaOras sa pagbasaOras ning pamamasa:
A technician comparing a handheld reference meter reading with a machine display.
Renal Care Academy›Hemodialysis Center Biomedical Technician›Module 9 of 12 Study series for dialysis biomedical technicians
🛠️

Study series for dialysis biomedical technicians

This module is study material for biomedical technicians in dialysis. It explains how machines work and why maintenance steps exist; it is not a service manual. Repairs, calibrations, and safety tests are done only by people trained and authorized for that machine model, using the manufacturer’s service manual, approved parts, and calibrated test equipment. Every case is fictional.

After this module you should be able to

Pre-testBefore you read, answer the 5 self-assessment questions and note your score. After the module, select New set of questions and compare. The difference is what you learned; the questions you still miss are what to reread.

The big picture, in plain words

In plain wordsThink of a tailor measuring cloth for a barong. Her tape measure has stretched a little over the years. If she checks it against a steel ruler and finds it a few millimeters long over a meter, she can still cut a good shirt, because a few millimeters do not matter for a sleeve. If the tape is off by a few centimeters, the shirt will not fit. Calibration answers the tailor’s question about the machine: is what it says true? You answer it by comparing the machine with an instrument you trust more, and correcting the difference if it is too large.

The picture has a limit. The tailor’s steel ruler is close enough to perfect for cloth. Your reference meter is not perfect at the scale you work at, and its own error can be a large share of the difference you are trying to judge. Two details therefore separate careful calibration from guesswork: knowing how large a difference is allowed, and knowing how far your own instrument can be trusted.

Figure 1 · Key terms for this module

CalibrationComparing with a trusted reference and adjusting to match.
VerificationComparing without adjusting, to confirm a reading is within tolerance.
ToleranceThe allowed difference.
AccuracyHow close an instrument’s reading is to the true value.
As found, as leftThe reading before and after adjustment.
TraceableLinked by documented calibrations to a national or international standard.
Standard solutionA liquid of precisely known value used to check a meter.

Case: The meter nobody checked

It is Wednesday, just after one in the afternoon. Your center calibrates dialysate conductivity on all twelve machines every six months, using one handheld reference meter. The chief technician, Mr. Ramon Cruz, finished the last round two weeks ago.

A visiting service engineer, Mr. Danilo Reyes, 46, borrows the meter and shows you its label. The meter’s own calibration certificate expired fourteen months ago. He checks it in fresh standard solution. The meter reads 3 percent low. On the floor, the nurses are connecting the afternoon shift.

Before reading on, decide:

Keep your answers in mind. The lesson gives you what you need, and the case review near the end shows one sound response.

What calibration is

A drifting sensor gives no warning. The number on the screen looks just as confident when it is wrong. Calibration brings that number back to the truth, one parameter at a time, in a fixed order.

Six steps of calibrating one parameter.
Figure 2. Calibrating one parameter. Six steps of calibrating one parameter.
Five calibrated parameters with the consequence of each being wrong.
Figure 3. Five things calibrated. Five calibrated parameters with the consequence of each being wrong.

Figure 4 · Calibrating one parameter

  1. StabilizeLet the machine reach steady operation.
  2. Measure as foundRead the machine and the reference at the same point and time.
  3. CompareIs the difference within the manual’s tolerance?
  4. AdjustIf not, adjust by the manual’s procedure.
  5. Measure as leftRead both again and confirm.
  6. RecordBoth sets of readings and the reference meter’s identity.
ParameterTypical referenceWhy it matters to the patient
Dialysate conductivityCalibrated conductivity meterThe patient receives the wrong amount of sodium and other salts
Dialysate temperatureCalibrated thermometerChilling if too cold, or damage to red cells if too hot
Arterial and venous pressureCalibrated pressure meterAlarms that are missed, or alarms that sound for nothing
Dialysate flow and blood pump speedMeasuring cylinder and timer, or flow meterLess blood cleaning than prescribed
Ultrafiltration volumeMeasuring cylinder or calibrated scaleToo much or too little fluid removed

Notice the order: you read the machine before you touch it. The “as found” value is not paperwork for its own sake. If a machine was far out of tolerance when you found it, patients treated since the last visit may have been affected, and the clinical team needs to know. Once you adjust, that evidence is gone.

Diagram of the parts of a hemodialysis prescription, such as session length, blood flow, dialysate flow, and dialyzer.
Figure 5. Each prescribed setting is only delivered if the machine’s sensors and pumps are accurate. Calibration connects the prescription to what the patient actually receives.

Deciding whether a reading is within tolerance

Every calibration procedure gives a tolerance: the largest acceptable difference between the machine and the reference. Your reference instrument is not perfect either; its certificate states its accuracy. So any difference you measure carries the reference’s own uncertainty with it. Work through the illustration below slowly, because it shows why a reading that passes can still deserve attention.

Illustration. The manual allows ±0.2 mS/cm for dialysate conductivity. Your reference meter is accurate to ±0.05 mS/cm. The machine shows 14.0 and the reference reads 14.15. The difference is 0.15: within tolerance, but close. The true difference could be anywhere from 0.10 to 0.20. Recording the as-found value lets the next visit see whether it is drifting.
SituationDecision
Well within toleranceRecord it; do not adjust
Within tolerance but close to the edge, or moving the same way at each visitRecord it, and adjust if the manual permits; look for the cause of the drift, such as scale
Outside toleranceRecord the as-found value; clean or descale if relevant; calibrate; record the as-left value; report
Far outside toleranceAs above, and tell the chief technician and the charge nurse the same day, because patients may have been affected
Cannot be brought within toleranceTake it out of service; repair it; then calibrate
RememberA reference instrument should be several times more accurate than the tolerance you are checking. If your meter’s own accuracy is as wide as the tolerance, it cannot tell you whether the machine passes or fails.

Who checks the checker?

Calibration moves the machine toward the reference. If the reference is wrong, you move every machine toward the same wrong value, and they then agree with each other perfectly. Nothing in the unit will reveal it.

Chain linking a machine sensor to a national standard through calibrated references.
Figure 6. Who checks the checker. Chain linking a machine sensor to a national standard through calibrated references.

Reference instruments therefore have their own calibration, done by an accredited laboratory or the manufacturer at a stated interval, with a certificate that links them to national standards. This chain is called traceability.

Figure 7 · Rules for test equipment

LabelEach instrument carries its identity and its calibration due date.
Check before useNever use an instrument past its due date.
Verify between calibrationsFor example, check a conductivity meter in standard solution.
Record which instrumentWrite it on every service record.
If one is found wrongReview every machine it was used on since it was last known to be good.

Looking after reference instruments

A reference instrument is accurate on the day the laboratory signs its certificate. What happens to it between that day and the next certificate is in your hands. The common mistakes in the table below are mistakes of handling, not faults of the instrument.

InstrumentDaily careCheck between calibrationsCommon mistakes
Conductivity meterRinse the cell with clean water and then with sample; keep it free of scaleStandard solution of a value near dialysate, on the day of useOld or contaminated standard solution; wrong reference temperature; air bubble in the cell
Reference thermometerKeep the probe clean and straightCompare with a second thermometer or an ice-point check, as its instructions allowReading before the probe has settled
Pressure meterKeep fluid out of it, and use a protectorZero check open to air; comparison with a second meterLeaking test tubing
Measuring cylinder or scaleKeep it clean and dry, on a level surfaceKnown weights for the scaleReading the cylinder from above; a scale not zeroed
Electrical safety analyzerKeep the leads undamagedSelf-check by its manualUsing it past its calibration date
MultimeterKeep the leads and probes intactKnown source checkWrong function selected

A standard solution is only as good as the way it is kept. It has an expiry date and a stated value at a stated temperature. Water evaporates from an open bottle and leaves the salts behind, so the solution becomes stronger than its label. A probe dipped into the stock carries rinse water and residue into it. So pour out what you need; never dip a probe into the stock bottle or pour used solution back. Close the bottle at once.

Figure 8 · The instrument register

IdentityEach instrument numbered and labeled.
CertificateDate, laboratory, and next due date.
Label on the instrumentShowing the due date.
Use logWhich machines it was used on, and when.
ChecksStandard solution and zero checks recorded.
Out-of-tolerance eventsWhat was found and which machines were reviewed.

A service record that stands up

Figure 9 · What every service record contains

The machineModel, serial number, asset number, location, operating hours.
The reasonScheduled visit or reported fault, with the exact message.
The findingsAs-found readings and what was observed.
The workSteps performed, parts replaced with part numbers, software version.
The testsAs-left readings, alarm and safety tests, with pass or fail.
The instrumentsIdentity and calibration due date of each.
The peopleTechnician’s name and signature, date, and who accepted the machine back.

A service record answers the question someone will ask months from now: what was done to this machine, by whom, and with what?

Keep records for the life of the machine, organized per machine so that its whole history can be read in order. Patterns show up only in a history: the same pump replaced three times, or conductivity drifting the same way at every visit.

Illustrative quality dashboard with metric tiles and small trend lines.
Figure 10. Simple counts from the service records, such as visits done on time, repeat faults, and days out of service, can be charted for the unit’s quality meeting.

Real situations: problems and solutions

These are fictional but typical. For each one, cover the answer, decide what you would do, then compare. Where a step depends on local rules, your unit’s procedure and your supervisor’s direction come first.

As found, far out

What you see
At a scheduled visit, before you adjust anything, a machine’s conductivity reads 4 percent different from the reference meter.
What is likely going on
Patients on this machine may have been receiving dialysate with the wrong salt content since the last visit.
What to do
Record the as-found value, correct it, and tell the chief technician and the charge nurse the same day.
Why it matters
The clinical team may need to review those patients. They can do that only if you tell them.

Two reference meters disagree

What you see
On a Monday check, the center’s two conductivity meters give readings on the same solution that differ by more than their stated accuracy.
What is likely going on
At least one of them is wrong.
What to do
Check both against fresh standard solution. Remove the one that fails, and review the machines it was used on.
Why it matters
A second meter is a cheap way to catch the first one drifting.

Standard solution left open

What you see
On Friday you notice that the bottle of conductivity standard has stood uncapped on the bench since Monday.
What is likely going on
Water has evaporated from it, so it is now more concentrated and reads higher than its label.
What to do
Discard it and open a fresh bottle. Recheck any calibration done with it.
Why it matters
A wrong standard makes a good meter wrong, and the meter then makes every machine wrong.
Field notes
  • Read the due date on the instrument before every use.
  • Record the as-found value before the as-left value.
  • Pour out standard solution; never dip a probe into the stock bottle.
  • If a reference is found to be wrong, review every machine it touched.

Back to the case: The meter nobody checked

What a meter that reads low means for the twelve machines. All twelve are wrong in the same direction. A meter that reads low makes a correct machine look high, so each was adjusted to match. If the meter reads 3 percent low, each machine adjusted to agree with it now delivers dialysate with conductivity about 3 percent higher than it displays. That means more sodium than prescribed, on all twelve machines, possibly for many months, because nobody knows when the meter began to drift.

What you do first, and whom you tell. Tell the chief technician, the charge nurse, and the medical director today. The clinical team decides what the error means for patients; that decision is theirs, not yours. Your work is the equipment. Obtain a reference meter with a valid certificate, verify every machine, record the as-found and as-left values, and correct those out of tolerance. The as-found values tell the clinical team how far off each machine was.

The control that would have stopped it. A register of test instruments with calibration due dates, a label on each instrument, and a rule that an overdue instrument is not used. It is simple and cheap. The due date was on the meter all along; nobody was required to read it.

Skills checklist: A calibration with a traceable record

Materials: an out-of-service machine, its service manual, a reference conductivity and temperature meter with a valid certificate, standard solution, and a blank service record. Your preceptor rates each step as satisfactory, needs practice, or not observed. Steps marked Critical must all be satisfactory before sign-off. This checklist supports supervised learning; your unit’s own procedure and competency form are the authority. A printable version of every checklist is on the trainer page.

Placement activity. Build or review your center’s test instrument register and maintenance calendar with a senior technician.

Check your understanding

Each visit draws 5 questions at random from this module’s 8-question bank. Choose the single best answer for each, then select Check answers; an explanation appears under the best answer. Select New set of questions for a different mix. A good target is 4 of 5 on two different sets. Nothing you select is sent anywhere; scoring happens in your browser.

Question 1 of 8What does calibration mean?
Question 2 of 8Why record the as-found value before adjusting?
Question 3 of 8A reference meter is past its calibration due date. What is correct?
Question 4 of 8What does traceable mean for a reference instrument?
Question 5 of 8A reference meter is discovered to have been reading wrong. What must be reviewed?
Question 6 of 8Which item belongs in every service record?
Question 7 of 8The tolerance is ±0.5 °C. The machine shows 37.0 and a reference thermometer reads 37.4. What is correct?
Question 8 of 8Why should you never dip a probe directly into the stock bottle of standard solution?

Read next on this site

Glossary & abbreviationsTalahulugan at mga daglatTalaan sa mga pulong ug daglatTalatinigan ampo reng daglat terms used on this page

Terms in plain words

Asset number
The facility’s own identification number for a piece of equipment.
Standard solution
A liquid of precisely known conductivity used to check a meter.
Accredited laboratory
A laboratory formally recognized as competent to calibrate instruments.
Dr. William Gregory M. Rivero, MD

William Gregory Rivero, MD, FPCP, DPSN

Internal Medicine · Nephrology · Nutrition · Philippines · PRC 0105184

Educational study material for biomedical technicians. Does not replace the manufacturer’s service manual and training, approved facility procedures, applicable electrical and medical device regulations, or supervised practice.

QR code — scan to save Dr. Rivero's contact info

Scan and save