Renal Technician Program · Module 5 of 12

Reverse Osmosis

How the membrane works, the readings that matter, rejection and recovery, recognizing patterns, and responding to alarms.

PublishedNailathalaGipatikPepalwal: Audience: Dialysis water treatment technicians and trainees · chief technicians, head nurses, and medical directors who supervise the water system Read timeOras ng pagbasaOras sa pagbasaOras ning pamamasa:
A technician reading flow meters on the front of a reverse osmosis unit.
Renal Care Academy›Water Treatment System Technician›Module 5 of 12 Study series for dialysis water treatment technicians
🛠️

Study series for dialysis water treatment technicians

This module is study material for dialysis water treatment technicians. It explains the reasons behind safe practice; it does not replace your unit’s approved procedures, the manufacturer’s instructions for your equipment, or the limits your facility has adopted. Limits quoted here come from the international standard and are given so the numbers make sense; your facility’s written limits and action levels are the ones you work to. Every case is fictional.

After this module you should be able to

Pre-testBefore you read, answer the 5 self-assessment questions and note your score. After the module, select New set of questions and compare. The difference is what you learned; the questions you still miss are what to reread.

The big picture, in plain words

In plain wordsThink of a patient whose blood pressure is taken at every visit. One reading tells the doctor little. A year of readings that creep upward, visit after visit, tells a story. The reverse osmosis unit is like that patient. It removes most of what is still dissolved in the water: salts, metals, and nearly all bacteria and endotoxin. You judge its health with a handful of readings.

The picture has a limit. A patient can tell you how he feels. A membrane cannot, and it changes slowly, without symptoms, until it fails. Your work is to read the numbers the way a careful doctor reads a chart: always against last week and last month, never alone.

Figure 1 · Key terms for this module

MembraneA rolled sheet with pores too small for most dissolved substances.
Product waterThe clean water that passes through. Also called permeate.
Reject waterThe concentrated stream carried away. Also called concentrate.
Percent rejectionHow much of the dissolved salt the membrane stopped.
Percent recoveryHow much of the feed water became product water.
FoulingDirt, organic matter, or bacteria coating the membrane.
ScalingA crust of minerals on the membrane.

Case: The numbers that drifted

It is the last Saturday of the month, after the second shift. Ms. Grace Tolentino, 35, the water treatment technician, sits down with the reverse osmosis log to prepare the monthly summary for the chief technician, Mr. Arnel Bautista.

She notices that product water conductivity has crept up from 8 to 19 microsiemens per centimeter over the month, while the feed water has stayed near 400. No alarm has sounded, because the alarm is set at 30. Product flow has also fallen a little over the same weeks.

Before reading on, decide:

Keep your answers in mind. The lesson gives you what you need, and the case review near the end shows one sound response.

How reverse osmosis works

Separate salty water from pure water with a fine membrane, and water moves on its own toward the salty side. This is osmosis. Reverse osmosis uses a pump to push the other way: pressure forces water out of the salty side, through the membrane, and leaves the salts behind.

Cutaway of a spiral-wound reverse osmosis membrane showing feed, product, and reject.
Figure 2. Inside a spiral-wound membrane. Cutaway of a spiral-wound reverse osmosis membrane showing feed, product, and reject.

Figure 3 · Inside the reverse osmosis unit

  1. FeedPretreated water enters.
  2. PumpRaises the pressure against the membrane.
  3. MembraneWater passes; most salts, bacteria, and endotoxin do not.
  4. ProductClean water goes on to the loop.
  5. RejectConcentrated water goes to drain, or part of it is recycled.
Engineering diagram of a dual-pass reverse osmosis system.
Figure 4. Some systems pass the product water through a second set of membranes (dual pass) for higher purity. The principle at each pass is the same.

A membrane is damaged by chlorine, plugged by scale and dirt, and fouled by bacteria. That is why the pretreatment stages matter so much. Temperature matters as well, because water moves through the membrane more slowly when it is cold: output falls by roughly 3 percent for each degree Celsius the feed water cools. The practical point is that a unit that keeps up in summer may fall short on a cold morning.

The readings that matter and what moves them

A typical unit shows the readings below. Record all of them at the same time each day, with the system running steadily, so that each day can be compared fairly with the last.

ReadingWhat it tells youIf it risesIf it falls
Feed pressure (before the pump)Whether pretreatment is supplying enough waterRarely a problemClogged prefilter, low supply, or a tank in backwash
Pump (membrane) pressureThe push against the membraneMembrane fouling or scaling, colder water, or a reject valve closed down too farPump wear, a leak, or low feed
Product flowHow much clean water is madeWarmer water; or a damaged membrane letting too much throughFouling, scaling, colder water, low pressure
Reject flowHow much water carries the salts awayLess recovery; water wastedRisk of scaling; check the reject valve setting
Feed conductivityThe mineral load coming inSource change, or the softener passing brineSource change
Product conductivityThe mineral load getting throughMembrane damage, a leaking seal, a higher feed load, warmer waterUsually good; confirm that the sensor is working
Feed temperatureExplains changes in flowMore flow, slightly lower rejectionLess flow
RememberAlways look at the temperature before you blame the membrane. A cold morning lowers product flow on a perfectly healthy unit.

Two calculations every technician should know

Percent rejection = (feed conductivity − product conductivity) ÷ feed conductivity × 100
Example: feed 400, product 8. (400 − 8) ÷ 400 × 100 = 98%.
Percent recovery = product flow ÷ feed flow × 100
Example: product 6 liters per minute from a feed of 10 liters per minute = 60%.

Conductivity measures how easily water carries electricity, and that depends on the salts dissolved in it. So conductivity is a convenient stand-in for the amount of dissolved material. Rejection tells you how well the membrane is separating. A healthy membrane usually rejects well over 90 percent; your system’s manual gives the expected figure and the alarm point. Recovery tells you how hard the membrane is being worked. Push recovery too high and the salts concentrate on the reject side until they form scale.

Formulas and worked examples for percent rejection and percent recovery.
Figure 5. Rejection and recovery. Formulas and worked examples for percent rejection and percent recovery.
Four trend sketches for a reverse osmosis unit with likely causes.
Figure 6. Reading the changes. Four trend sketches for a reverse osmosis unit with likely causes.

Figure 7 · Reading the changes

Rejection fallingMembrane damage (often from chlorine), a leaking seal, or scaling. Report it even if no alarm has sounded.
Product flow fallingFouling or scaling, colder feed water, or low pump pressure.
Feed-side pressure drop risingThe membrane or its prefilter is plugging.
Sudden change after serviceSomething was put back together incorrectly.
RememberConductivity does not detect chlorine, chloramine, bacteria, or endotoxin. A perfect rejection figure never replaces the chlorine test or the monthly cultures.

Recognizing the pattern

A single reading can mislead you. The pattern across several readings points to the cause, the way a set of symptoms points to a diagnosis.

PatternMost likely causeWhat usually follows
Product flow falling slowly, pump pressure rising, rejection steadyFouling by particles or biological growthCheck the prefilters and pretreatment; membrane cleaning by the supplier
Product flow falling, rejection fallingScaling from hardnessCheck softener performance; cleaning with a descaling agent by the supplier
Product flow rising, rejection fallingMembrane damage, classically from chlorineCheck the chlorine logs; membrane replacement
Sudden jump in product conductivity after serviceA seal or connector not seatedCall the service provider back
Everything changes together on one dayThe feed water changedCheck the source; compare feed conductivity and temperature

Membranes are cleaned or replaced by trained service staff, with chemicals and procedures specific to each membrane. Your part is to notice early, report with figures, and make sure the cause upstream is fixed. Otherwise the new membrane goes the same way as the old one.

Start-up, shutdown, and alarms

Figure 8 · A routine start-up

  1. Check pretreatmentTanks in service, timers correct, valves in their normal positions.
  2. Open the feedConfirm that the feed pressure is adequate.
  3. Start the unitMany units flush to drain first; product is diverted until its quality is acceptable.
  4. Wait for steady readingsDo not test or record during the first minutes.
  5. Record and compareToday’s readings against yesterday and last month.

Many units flush themselves at intervals while idle, so that water does not stand against the membrane. A unit left switched off for days grows bacteria and may need disinfection before use. After any long shutdown, follow your procedure before the water goes to patients.

AlarmWhat the unit usually doesFirst checks
Low feed pressureStops the pump to protect itSupply, prefilter pressure drop, a tank in backwash or regeneration, a closed valve
High product conductivityDiverts product to drain, or alarmsFeed conductivity, softener status, recent service, the rejection trend
High temperatureStops to protect the membraneFeed temperature; a pump running against a closed valve
Low product flow or low pressure in the loopMachines alarm for low waterDemand at that moment, temperature, the product flow trend
Motor overloadPump tripsDo not reset it again and again; report it for an electrical check
Stop and reportReport any alarm that affects water quality or supply to the charge nurse at once, with the stations that are running. Resetting an alarm is not a response.
Engineering diagram of a dual-pass reverse osmosis system.
Figure 9. In a dual-pass system each pass has its own readings. Record both, and calculate the rejection for each pass.

Real situations: problems and solutions

These are fictional but typical. For each one, cover the answer, decide what you would do, then compare. Where a step depends on local rules, your unit’s procedure and your supervisor’s direction come first.

Not enough water at peak

What you see
On Monday mornings at 6:30, when every station starts at once, three or four machines alarm for low water pressure.
What is likely going on
Demand is more than the system can deliver at that moment. The feed water may be colder, the membrane may be fouled, or a station may have been added.
What to do
Record product flow and feed temperature at the time of the alarms, and report. Ask the charge nurse whether starts can be staggered while the cause is found.
Why it matters
Capacity problems show first at peak demand.

Rejection fell after the breakthrough

What you see
A week after a chlorine breakthrough, percent rejection is three points lower than before it and has not recovered.
What is likely going on
Chlorine has probably damaged the membrane.
What to do
Report it with the before-and-after figures. Expect that the membrane will need to be replaced.
Why it matters
Each stage protects the next one. A carbon failure becomes a membrane failure.

The reject line is hot

What you see
At 9:15 during the first shift, the reverse osmosis pump housing is hot to the touch and much louder than usual.
What is likely going on
The pump may be running against a closed valve, or without enough feed water.
What to do
Follow the shutdown procedure. Check the valve positions and the feed pressure, and report before you restart.
Why it matters
A pump that runs dry fails within minutes.

Low flow on a cold morning

What you see
On a cool January morning, product flow is a tenth lower than yesterday. Feed water temperature has dropped four degrees overnight.
What is likely going on
Cold water passes through the membrane more slowly. A fall of this size is about what the temperature change predicts.
What to do
Record both readings and the temperature. If flow recovers as the water warms through the day, there is no fault. If it does not, report it.
Why it matters
Checking the temperature first prevents a needless service call, and it teaches you how your own system normally behaves.
Field notes
  • Calculate rejection yourself now and then. Do not rely only on the display.
  • Compare today with last month, not only with yesterday.
  • Conductivity tells you nothing about chlorine or bacteria.
  • Walk the loop with the drawing once a year and mark everything that has changed.

Back to the case: The numbers that drifted

The rejection, then and now. At the start of the month rejection was (400 − 8) ÷ 400 = 98 percent. Now it is (400 − 19) ÷ 400 = about 95 percent. The membrane is letting through more than twice as much salt as it did four weeks ago.

Is there anything to report? Yes. The alarm has not sounded, but the trend is clear, and a clear trend is reportable. Tell Mr. Bautista, or whoever is responsible for the system, with the figures in front of you, and record that you told him. An alarm marks the point where the problem is already large; a trend lets the unit act while it is still small.

What might be happening to the membrane. Falling rejection together with falling flow suggests the membrane is scaling or fouling. Falling rejection on its own, especially after a chlorine breakthrough, suggests the membrane has been damaged. Finding the cause is for the responsible person and the service provider. Your part was to notice it three weeks before the alarm would have.

Skills checklist: Reverse osmosis readings, calculations, and alarm response

Materials: the unit’s display and gauges, the log with at least one month of readings, a calculator, and the alarm list from the manual. Your preceptor rates each step as satisfactory, needs practice, or not observed. Steps marked Critical must all be satisfactory before sign-off. This checklist supports supervised learning; your unit’s own procedure and competency form are the authority. A printable version of every checklist is on the trainer page.

Placement activity. Plot one month of product flow, pump pressure, product conductivity, and temperature and present the pattern to your preceptor.

Check your understanding

Each visit draws 5 questions at random from this module’s 8-question bank. Choose the single best answer for each, then select Check answers; an explanation appears under the best answer. Select New set of questions for a different mix. A good target is 4 of 5 on two different sets. Nothing you select is sent anywhere; scoring happens in your browser.

Question 1 of 8What passes through a reverse osmosis membrane most easily?
Question 2 of 8Feed conductivity is 500 and product conductivity is 25. What is the percent rejection?
Question 3 of 8Feed flow is 20 liters per minute and product flow is 12. What is the percent recovery?
Question 4 of 8Rejection has fallen steadily for three weeks but no alarm has sounded. What should you do?
Question 5 of 8Which of these does conductivity NOT detect?
Question 6 of 8Product flow drops on an unusually cold morning with no other change. What is the most likely reason?
Question 7 of 8Product flow is rising and rejection is falling. What is the most likely cause?
Question 8 of 8The low feed pressure alarm stops the pump. What is checked first?

Read next on this site

Glossary & abbreviationsTalahulugan at mga daglatTalaan sa mga pulong ug daglatTalatinigan ampo reng daglat terms used on this page

Terms in plain words

Permeate
Another word for product water.
Concentrate
Another word for reject water.
Dual pass
Two sets of membranes in series; product from the first feeds the second.
Microsiemens per centimeter
The unit of conductivity for very pure water.
Dr. William Gregory M. Rivero, MD

William Gregory Rivero, MD, FPCP, DPSN

Internal Medicine · Nephrology · Nutrition · Philippines · PRC 0105184

Educational study material for dialysis water treatment staff. Does not replace the medical director’s direction, approved unit procedures, manufacturer instructions, the standard edition your facility follows, DOH licensing requirements, or supervised training.