Renal Technician Program · Module 3 of 12

Sediment Filtration and Softening

Filters, pressure drop, and backwash; how a softener works and regenerates; the hardness test; and troubleshooting.

PublishedNailathalaGipatikPepalwal: Audience: Dialysis water treatment technicians and trainees · chief technicians, head nurses, and medical directors who supervise the water system Read timeOras ng pagbasaOras sa pagbasaOras ning pamamasa:
A technician checking the salt level in a softener brine tank.
Renal Care Academy›Water Treatment System Technician›Module 3 of 12 Study series for dialysis water treatment technicians
🛠️

Study series for dialysis water treatment technicians

This module is study material for dialysis water treatment technicians. It explains the reasons behind safe practice; it does not replace your unit’s approved procedures, the manufacturer’s instructions for your equipment, or the limits your facility has adopted. Limits quoted here come from the international standard and are given so the numbers make sense; your facility’s written limits and action levels are the ones you work to. Every case is fictional.

After this module you should be able to

Pre-testBefore you read, answer the 5 self-assessment questions and note your score. After the module, select New set of questions and compare. The difference is what you learned; the questions you still miss are what to reread.

The big picture, in plain words

In plain wordsThink of the screen on a kitchen sink drain and the descaler you run through a kettle. Neither one cooks anything. But leave the screen off and the pipe clogs, and skip the descaler and the kettle stops heating. The first two stages of the water room are like that: a filter that catches dirt and a softener that removes the minerals that form scale.

The picture has a limit. A clogged sink is noticed the same day. A neglected filter or softener does its damage quietly, further down the line, to the carbon and the membrane that depend on it. Neither stage protects the patient directly. They protect the stages that do, and when they are neglected, everything downstream fails sooner.

Figure 1 · Key terms for this module

Multimedia filterA tank with layers of sand and gravel that trap particles.
Cartridge filterA replaceable filter element held in a housing.
Pressure dropThe pressure before the filter minus the pressure after it.
BackwashReverse flow that lifts and rinses the bed.
ResinBeads that exchange sodium for calcium and magnesium.
RegenerationRecharging the resin with brine.
HardnessDissolved calcium and magnesium.

Case: The softener that regenerated at noon

A brownout cut the power to the clinic for about two hours on Monday night. Nobody reset the softener’s timer, and it is now several hours wrong. On Tuesday at 12:10, in the middle of the second shift, the softener starts to regenerate.

A few minutes later the machines at stations 4 and 7 alarm for high conductivity. The charge nurse, Mr. Paolo Aquino, calls the water room. The technician, Ms. Rhea Domingo, 27, finds the reverse osmosis display showing product conductivity rising with each reading, and she can hear the softener valve cycling.

Before reading on, decide:

Keep your answers in mind. The lesson gives you what you need, and the case review near the end shows one sound response.

Sediment filters and the pressure drop

Left-to-right flow diagram of the dialysis water treatment train, from feed water through filters, softener, carbon tanks, and reverse osmosis to the distribution loop.
Figure 2. Find the pretreatment stages on the left of the treatment train. The order of the softener and the carbon tanks differs between installations; follow your own system drawing.

A sediment filter is a tank of layered sand and gravel, or a cartridge in a housing, that traps particles. As it fills with dirt, water has a harder time getting through, so the pressure after the filter falls below the pressure before it. That is why every filter has two gauges, one before and one after.

Cutaway filter with gauges before and after showing an 8 psi pressure drop.
Figure 3. Reading the pressure drop. Cutaway filter with gauges before and after showing an 8 psi pressure drop.
Pressure drop = pressure before the filter − pressure after the filter.
Example: 60 psi before and 52 psi after gives a drop of 8 psi. If the drop was 3 psi when the filter was new, it is clogging.

Your unit’s procedure states the pressure drop at which a tank is backwashed or a cartridge is changed. Record both gauges every day. One reading tells you little. A row of readings shows you the filter loading up, day by day, before it starves the system of water.

Backwashing a media filter

Most media filters and carbon tanks backwash on their own, on a timer or when the pressure drop reaches a set point. You still need to know what the control valve is doing, because you are the one who will notice when it does not do it.

Figure 4 · An automatic backwash cycle

  1. Service endsThe control valve takes the tank off line.
  2. BackwashWater flows upward, lifts the bed, and carries the trapped dirt to drain.
  3. SettleFlow stops, and the layers settle again with the heaviest at the bottom.
  4. RinseDownward flow to drain clears the loosened particles.
  5. Return to serviceWater flows forward again.
Check during or after a backwashWhat you want to seeIf not
Timer settingOutside treatment hoursReset and report; a backwash during treatment hours starves the system of water
Drain flowStrong flow, dirty at first, clearing by the endWeak flow suggests a blocked drain line or a failing valve
Media in the drainNoneSand or carbon in the drain means the flow is too high or an internal part has broken
Pressure drop afterwardBack near the clean valueIf it stays high, the bed is fouled or channeled and needs service

The softener and its brine tank

Hard water carries calcium and magnesium. On a reverse osmosis membrane they form scale, like the white crust inside an old kettle, and the scale blocks the membrane. A softener is a tank of resin beads loaded with sodium. As hard water flows past, the beads take up calcium and magnesium and release sodium in their place.

Five stages of the softener cycle with resin tank and brine tank.
Figure 5. The softener cycle. Five stages of the softener cycle with resin tank and brine tank.
Two brine tank cutaways comparing normal salt with a salt bridge.
Figure 6. Inside the brine tank. Two brine tank cutaways comparing normal salt with a salt bridge.

Figure 7 · The softener cycle

  1. ServiceHard water in, soft water out. The resin slowly fills with calcium and magnesium.
  2. BackwashWater runs upward to loosen and rinse the bed.
  3. Brine drawStrong salt water from the brine tank pushes calcium and magnesium off the resin.
  4. RinseThe extra salt is flushed to drain.
  5. Back in serviceThe resin is recharged with sodium.

During regeneration the softener is off line and salty water is flowing to drain. If the valve fails or the timer is wrong, that brine can travel downstream instead. For this reason a softener is set to regenerate only when no patient is on dialysis, and its timer is checked every day.

You check the softener’s work with a hardness test on the water leaving it, usually at the end of the treatment day, when the resin is most used up. Many units use a limit of about 1 grain per gallon (17 mg/L as calcium carbonate); work to the limit in your own procedure. Check also that the brine tank holds enough salt, and that the salt has not formed a hard crust, a bridge, above the water.

RememberCheck the softener clock every morning, and after every power interruption.

The hardness test and what the result means

Test hardness on the water leaving the softener, usually at the end of the treatment day, when the resin has done the most work. If the softener is running short, that is when it will show. Most kits are either drop-count titrations, in which you count drops until the color changes, or dip strips.

Figure 8 · Performing a drop-count hardness test

  1. Flush the portRun water first, so that you test what the softener is producing now.
  2. Measure the sampleFill the vial exactly to the mark.
  3. Add the indicatorThe sample turns one color if hardness is present.
  4. Add reagent drop by dropSwirl after each drop, and count.
  5. Read the end pointThe color change marks the end; the drop count gives the hardness from the kit’s table.
  6. RecordValue, unit, port, time, and your name.
ResultMeaningAction
Within the unit’s limitThe softener has capacityRecord
Rising over several days but still within the limitCapacity is being used up sooner than beforeReport the trend; check the salt, how often it regenerates, and whether the feed water has become harder
Above the limitResin exhausted, or regeneration failedReport; check the salt, a bridge, the clock, and the brine draw; regenerate outside treatment hours; retest before the next shift

A softener starts regeneration in one of two ways: by the clock, every so many days, or by a meter, after so many liters. A meter-based softener adjusts when water use rises. A clock-based one does not, so it can run out early during a busy week.

Troubleshooting filters and softeners

Work through a fault in order, from the cause you can check fastest to the one that needs a service visit.

ProblemLikely causesWhat to check, in order
Pressure drop rises within days of a backwashDirtier feed water; backwash not finishing; bed fouledFeed water appearance, backwash drain flow, the timer, then call for service
Low pressure everywhere downstreamClogged filter; booster pump fault; low feed pressure from the streetThe feed gauge first, then each gauge in turn, to find where the pressure is lost
Hard water after the softenerNo salt; salt bridge; clock wrong; brine line blocked; resin exhausted or fouled with ironSalt level, bridge, clock, the brine draw during a regeneration, then the resin
Salty water downstream, high conductivity alarmsRegeneration during treatment; rinse too short; control valve stuckStop or bypass by procedure, report, complete the rinse, test hardness and conductivity
Brine tank overflowing or not refillingFault in the float or the refill valveReport for service; do not leave the softener without regeneration
Resin beads in the downstream filterBroken internal screenReport; the softener needs repair before it damages the membrane
RememberTo find where pressure is being lost, read the gauges in order from the feed line. The fault lies between the last good gauge and the first bad one.

Real situations: problems and solutions

These are fictional but typical. For each one, cover the answer, decide what you would do, then compare. Where a step depends on local rules, your unit’s procedure and your supervisor’s direction come first.

Pressure drop climbing fast

What you see
After four days of heavy rain, the log shows the sediment filter’s pressure drop has doubled, from 4 psi on Monday to 8 psi on Friday.
What is likely going on
The feed water is carrying much more silt than usual, and the filter is loading up with it.
What to do
Backwash or change the filter by procedure, record it, and report. Check the cartridge filter downstream as well.
Why it matters
A clogged filter starves the reverse osmosis unit of water and can shut down the shift.

Hardness after the softener

What you see
At 5:30 in the afternoon, the end-of-day hardness test on the water leaving the softener reads above the unit’s limit.
What is likely going on
The resin is exhausted, the brine tank is empty or bridged, or the last regeneration did not run.
What to do
Check the salt and the clock, and report. Regenerate outside treatment hours by procedure, and retest before the next shift starts.
Why it matters
Hard water scales the membrane within days and lowers the quality of the product water.

A salt bridge

What you see
The brine tank looks full to the top, but the morning hardness test after the softener is high for the second day running.
What is likely going on
The salt has formed a hard crust with an empty space underneath. The water never reaches the salt, so no brine is being made.
What to do
Break the bridge carefully by the unit’s method, check the water level in the tank, and run a regeneration at the proper time.
Why it matters
A tank that looks full is exactly why this fault is missed.

Backwash at ten in the morning

What you see
At ten in the morning, with the first shift on, the sediment filter starts to backwash. Machines across the floor alarm for low water pressure.
What is likely going on
The backwash timer has drifted, or it was reset by a power cut.
What to do
Tell the charge nurse why the pressure dropped. Let the cycle finish or stop it by procedure, then correct the timer and record what you found.
Why it matters
Every automatic valve in the water room runs on a clock, and every clock is suspect after a power interruption.
Field notes
  • Check the softener clock every morning and after every power interruption.
  • Two gauge readings give you a pressure drop. One gauge reading gives you almost nothing.
  • Backwashing does not renew exhausted carbon. Only replacement does.
  • Keep spare cartridges, salt, and in-date test kits above a reorder point.

Back to the case: The softener that regenerated at noon

What the softener is sending downstream. Brine. During regeneration the softener flushes strong salt water to drain, and with the timer wrong it is doing so while the system is supplying patients. Salt that slips downstream raises the load on the reverse osmosis membrane and the conductivity of the product water, which is what the machines at stations 4 and 7 detected.

What to do at the softener, and whom to tell. Follow the unit’s procedure to stop the regeneration or bypass the softener. Tell Mr. Aquino at once which machines alarmed and why. He decides what happens with the connected patients; that decision is not yours. Do not return the softener to service until it has finished a full rinse and the hardness and conductivity results are acceptable. Record the times, the readings, and whom you told.

The check that prevents it. Compare the softener clock with the real time every morning and after every power interruption, and write it in the log. A two-minute check would have kept this regeneration in the night where it belongs.

Skills checklist: Filter and softener round with a hardness test

Materials: the log, a hardness test kit, the softener manual, and the system drawing. Your preceptor rates each step as satisfactory, needs practice, or not observed. Steps marked Critical must all be satisfactory before sign-off. This checklist supports supervised learning; your unit’s own procedure and competency form are the authority. A printable version of every checklist is on the trainer page.

Placement activity. Observe one full softener regeneration from start to finish with your preceptor and note what the drain flow looks like at each stage.

Check your understanding

Each visit draws 5 questions at random from this module’s 8-question bank. Choose the single best answer for each, then select Check answers; an explanation appears under the best answer. Select New set of questions for a different mix. A good target is 4 of 5 on two different sets. Nothing you select is sent anywhere; scoring happens in your browser.

Question 1 of 8A filter reads 58 psi before and 46 psi after. What is the pressure drop?
Question 2 of 8What does a softener remove?
Question 3 of 8Why are softeners set to regenerate outside treatment hours?
Question 4 of 8After a power interruption, which softener check comes first?
Question 5 of 8During a backwash, in which direction does water flow through the bed?
Question 6 of 8Sand appears in the drain during backwash. What does this suggest?
Question 7 of 8A clock-based softener runs out early during an unusually busy week. Why?
Question 8 of 8Pressure is normal at the feed gauge and after the sediment filter, but low after the softener. Where is the problem?

Read next on this site

Glossary & abbreviationsTalahulugan at mga daglatTalaan sa mga pulong ug daglatTalatinigan ampo reng daglat terms used on this page

Terms in plain words

Titration
Adding a reagent drop by drop until a color change shows the end point.
Control valve
The automatic valve on top of a tank that directs service, backwash, and rinse flows.
Channeling
Water cutting a path through a bed so most of the bed is bypassed.
Dr. William Gregory M. Rivero, MD

William Gregory Rivero, MD, FPCP, DPSN

Internal Medicine · Nephrology · Nutrition · Philippines · PRC 0105184

Educational study material for dialysis water treatment staff. Does not replace the medical director’s direction, approved unit procedures, manufacturer instructions, the standard edition your facility follows, DOH licensing requirements, or supervised training.