Using the series to train biomedical staff
Use the twelve modules as guided study alongside the manufacturer’s training for each machine model and supervised bench work. A practical sequence is one module a week, each followed by its checklist on a machine that is out of clinical service.
Figure 1 · Suggested sequence
- Open the manualFind the flow diagram and the maintenance schedule for the model.
- Study the moduleRead, work the case, and take the self-assessment.
- DemonstrateA senior technician performs the task and explains each step.
- PracticeThe learner performs it under observation on an out-of-service machine.
- Sign offRecord competence per model on the facility’s form.
Proposed completion criteria
- All twelve self-assessments attempted, with missed questions reviewed.
- Every Critical step on every checklist demonstrated satisfactorily.
- Manufacturer training completed for each machine model the technician will service.
- One complete preventive maintenance visit and one fault repair documented and reviewed by a senior technician.
- The four capstone stations in Module 12, with all critical steps satisfactory.
These are proposals for the facility to adapt.
Written examination. The end-of-course examination draws 30 questions at random from the question banks of all modules. A proposed pass mark is 80 percent, with missed questions reviewed against their modules before a second sitting. For a supervised sitting, the learner takes it on a unit device with the preceptor present and the preceptor records the score.
Skills checklists for every module
Print this page for a full set. Each checklist starts on a new sheet. Steps marked Critical must all be satisfactory before sign-off; a missed critical step is repeated with the preceptor. Replace any step that does not match your equipment or approved procedure.
Module 1: The Biomedical Technician’s Role and Safe Working
Safe preparation for service and formal release. Materials: an out-of-clinical-use machine, the service manual, tags, a lockout device, protective equipment, and blank service and release forms.
| Observable step | Satisfactory | Needs practice | Not observed |
|---|---|---|---|
| Confirms with the charge nurse that the machine is free and records its identity and assigned area. | |||
| CriticalApplies a visible out-of-service tag with date, reason, and name. | |||
| CriticalConfirms the machine has completed disinfection before opening it. | |||
| CriticalSwitches off, unplugs, closes the water supply, and applies lockout. | |||
| Wears gloves and eye protection and uses static protection when handling boards. | |||
| Locates the correct service manual and revision for the model. | |||
| CriticalStates which tests are required after the simulated repair, by the manual. | |||
| CriticalCompletes the service record and obtains acceptance from the charge nurse before removing the tag. |
Learner ____________________ Preceptor ____________________ Date __________
Module 2: Electrical and Electronic Fundamentals
Multimeter basics on a de-energized training board. Materials: a multimeter with a valid calibration label, a training board or an unplugged out-of-service machine, the service manual, a wrist strap, and an antistatic mat.
| Observable step | Satisfactory | Needs practice | Not observed |
|---|---|---|---|
| CriticalConfirms the equipment is unplugged and locked out before resistance or continuity checks. | |||
| CriticalSelects function and range before connecting the probes. | |||
| Tests the meter on a known source. | |||
| CriticalChecks a fuse for continuity and confirms its rating against the manual. | |||
| Measures a heater element or valve coil resistance and compares with the manual. | |||
| Inspects the plug, cord, and inlet and states what would reject each. | |||
| Uses a wrist strap and handles a circuit board by its edges. | |||
| Identifies five sensors and five actuators on the machine using the flow diagram. |
Learner ____________________ Preceptor ____________________ Date __________
Module 3: Dialysate Side I: Water Inlet, Heating, and Degassing
Temperature verification against a reference thermometer. Materials: an out-of-service machine, its service manual, a calibrated reference thermometer with test adapter, and a service record.
| Observable step | Satisfactory | Needs practice | Not observed |
|---|---|---|---|
| CriticalChecks the reference thermometer’s identity and calibration due date. | |||
| Brings the machine to steady dialysis conditions. | |||
| Measures at the point the manual specifies and waits for a stable reading. | |||
| CriticalRecords as-found display and reference values. | |||
| CriticalStates the tolerance and decides correctly whether adjustment is needed. | |||
| CriticalDemonstrates the high-temperature alarm and bypass by the manual’s method. | |||
| Inspects the inlet hose, filter, and connections. | |||
| Records the result with the instrument’s identity. |
Learner ____________________ Preceptor ____________________ Date __________
Module 4: Dialysate Side II: Proportioning, Conductivity, and Bypass
Conductivity verification and bypass proof. Materials: an out-of-service machine with the unit’s concentrates, its service manual, a reference conductivity meter, standard solution, and a service record.
| Observable step | Satisfactory | Needs practice | Not observed |
|---|---|---|---|
| CriticalChecks the reference meter’s calibration date and verifies it against standard solution. | |||
| CriticalConfirms the meter and the machine use the same reference temperature and units. | |||
| CriticalConfirms the machine’s configured concentrate system matches the concentrates connected. | |||
| Samples by the manual’s method at steady state. | |||
| Records as-found readings for each conductivity cell. | |||
| CriticalStates the tolerance and decides correctly whether calibration is needed. | |||
| CriticalDemonstrates alarm and bypass for an out-of-limit condition by the manual’s method. | |||
| Records as-left readings with the meter’s identity. |
Learner ____________________ Preceptor ____________________ Date __________
Module 5: Fluid Balance and Ultrafiltration Control
Ultrafiltration accuracy check with a measuring cylinder. Materials: an out-of-service machine, its service manual, the test connector, a measuring cylinder or calibrated scale, a timer, and a service record.
| Observable step | Satisfactory | Needs practice | Not observed |
|---|---|---|---|
| States which fluid balance design the machine uses. | |||
| Removes air from the loop before testing. | |||
| CriticalPerforms the tightness test by the manual and interprets the result. | |||
| Sets the ultrafiltration rate and time the manual specifies. | |||
| CriticalCollects and reads the volume correctly at eye level. | |||
| CriticalCalculates the percent error and compares it with the tolerance. | |||
| Calibrates or reports for repair if outside tolerance, then repeats the check. | |||
| Records as-found and as-left values and the measuring device used. | |||
| Inspects and replaces worn dialyzer connector seals. |
Learner ____________________ Preceptor ____________________ Date __________
Module 6: The Blood Side: Pump, Pressures, and Heparin Pump
Pump occlusion and pressure verification. Materials: an out-of-service machine, its service manual, the unit’s tubing set, a calibrated pressure meter, a measuring cylinder, a timer, and the unit’s heparin syringes.
| Observable step | Satisfactory | Needs practice | Not observed |
|---|---|---|---|
| CriticalConfirms the tubing size setting matches the unit’s tubing set. | |||
| CriticalPerforms the occlusion test by the manual and adjusts if required. | |||
| Verifies rotor speed or flow delivery at the points the manual specifies. | |||
| Inspects rollers, springs, and guides for wear. | |||
| CriticalConfirms the cover switch stops the pump. | |||
| CriticalVerifies arterial and venous pressure readings against the reference at zero and across the range. | |||
| CriticalConfirms the heparin pump’s syringe configuration and checks delivery accuracy. | |||
| Confirms the hand crank is present and works. | |||
| Records as-found and as-left values with instrument identities. |
Learner ____________________ Preceptor ____________________ Date __________
Module 7: Protective Systems and Alarm Testing
Functional tests of the air detector, clamp, and blood leak detector. Materials: an out-of-service machine, its service manual, a training tubing set, the manufacturer’s test fixtures or solutions, a calibrated pressure meter, and a service record.
| Observable step | Satisfactory | Needs practice | Not observed |
|---|---|---|---|
| Reads the test method and pass criteria in the manual before starting. | |||
| CriticalSets up with training tubing and confirms no patient connection. | |||
| CriticalPerforms the air detector test and confirms pump stop, clamp closure, and alarm. | |||
| CriticalPerforms the venous clamp holding test and records the pressure held. | |||
| CriticalCleans and tests the blood leak detector by the manual. | |||
| Confirms cover and door switches stop the pump. | |||
| Confirms the audible alarm and the power-failure alarm. | |||
| CriticalKeeps the machine out of service if any test fails, and states the next steps. | |||
| Records each result with instruments used. |
Learner ____________________ Preceptor ____________________ Date __________
Module 8: The Preventive Maintenance Program
Building a maintenance calendar and running a visit. Materials: the equipment inventory, service manuals, a year planner, one maintenance kit, and a blank visit checklist.
| Observable step | Satisfactory | Needs practice | Not observed |
|---|---|---|---|
| Lists every machine with model, serial number, interval, and last visit date. | |||
| Calculates the number of visits due per year and spreads them across the calendar. | |||
| CriticalIdentifies any machine currently overdue and reports it. | |||
| CriticalChecks for open safety notices affecting each model. | |||
| Follows the visit order: prepare, receive, decontaminate, inspect, replace, verify, test, finish. | |||
| Asks the operators about problems before starting. | |||
| CriticalRecords as-found values before adjusting. | |||
| Updates the inventory with the visit date and next due date. | |||
| CriticalReleases the machine formally to the charge nurse. |
Learner ____________________ Preceptor ____________________ Date __________
Module 9: Calibration and Test Equipment
A calibration with a traceable record. Materials: an out-of-service machine, its service manual, a reference conductivity and temperature meter with a valid certificate, standard solution, and a blank service record.
| Observable step | Satisfactory | Needs practice | Not observed |
|---|---|---|---|
| CriticalChecks the reference instrument’s identity and calibration due date before use. | |||
| Verifies the reference meter against standard solution. | |||
| Allows the machine to reach steady operation before measuring. | |||
| CriticalMeasures and records as-found conductivity and temperature. | |||
| CriticalStates the tolerance from the service manual and decides correctly whether adjustment is needed. | |||
| Adjusts by the manual’s procedure, if needed, and records as-left values. | |||
| CriticalCompletes the service record with machine identity, readings, instrument identity, and signature. | |||
| CriticalReports any as-found value far outside tolerance to the chief technician and charge nurse. |
Learner ____________________ Preceptor ____________________ Date __________
Module 10: Disinfection, Descaling, and Machine Fluid Hygiene
Disinfection with residual test, hose check, and filter change. Materials: an out-of-service machine, its manual, approved agents, residual strips for those agents, protective equipment, a replacement dialysis fluid filter, and the machine’s record.
| Observable step | Satisfactory | Needs practice | Not observed |
|---|---|---|---|
| CriticalSelects the program and agent approved for the stated purpose. | |||
| Wears the protective equipment on the safety data sheet. | |||
| CriticalAllows the full cycle and rinse to complete. | |||
| CriticalPerforms the residual test with the correct strip at the stated outlet and time. | |||
| Inspects the inlet hose and concentrate connectors and states their replacement dates. | |||
| CriticalConfirms the drain hose has an air gap. | |||
| CriticalChecks the dialysis fluid filter’s due date and replaces it by the manual if due. | |||
| Records agent, program, times, residual result, and filter details. |
Learner ____________________ Preceptor ____________________ Date __________
Module 11: Electrical Safety Testing and Facility Utilities
Electrical safety test with an analyzer. Materials: an out-of-service machine, its service manual, an electrical safety analyzer with a valid calibration, an outlet tester, and the test record.
| Observable step | Satisfactory | Needs practice | Not observed |
|---|---|---|---|
| CriticalCarries out the visual inspection in order and corrects or reports defects before testing. | |||
| CriticalChecks the analyzer’s calibration date and leads. | |||
| CriticalMeasures protective earth resistance while flexing the cord and compares with the limit. | |||
| CriticalPerforms the leakage or insulation tests by the manual’s method and compares with the limits. | |||
| Confirms the machine functions normally afterward. | |||
| CriticalTags the machine out of service if any test fails. | |||
| Records values, limits, analyzer identity, and the next due date, and labels the machine. | |||
| Checks the station outlet with an outlet tester and reports any fault to the facility electrician. | |||
| CriticalConfirms no extension leads or adapters are in use at the station. |
Learner ____________________ Preceptor ____________________ Date __________
Module 12: Troubleshooting, Life Cycle, and Capstone
Fault-finding and return to service. Materials: an out-of-service machine with a simulated fault set by the preceptor, the service manual, test instruments, and blank service and release forms.
| Observable step | Satisfactory | Needs practice | Not observed |
|---|---|---|---|
| Collects the exact alarm text and the circumstances before touching the machine. | |||
| CriticalChecks simple and shared causes before opening the cabinet. | |||
| Uses the flow diagram and fault table to narrow the fault area. | |||
| Tests one idea at a time and states the reasoning. | |||
| CriticalRepairs by the manual with approved parts. | |||
| CriticalLists and performs the verification tests the repair requires. | |||
| CriticalDisinfects and passes a residual test if the fluid path was opened. | |||
| CriticalCompletes the service record and hands over in person to the charge nurse before removing the tag. |
Learner ____________________ Preceptor ____________________ Date __________