- Read the RO feed and permeate meters (and reject, if metered) at the start and end of the same interval, on the same boundary.
- Enter completed HD sessions in that interval and your facility's metering tolerance.
- Optionally choose a reuse mode, the engineering-approved recoverable fraction or measured capture, and the eligible nonclinical demand it would replace.
- Read recovery, per-session volumes, the residual, and the modeled displacement. Treat flagged results as prompts to check meters, not as findings.
All computation runs in your browser; no values are stored or transmitted.
When to Use
Use it for baseline water accounting, to track liters per completed treatment over time, and to size a proposed reuse project for engineering and infection-prevention review.
When not to rely on it
Not for setting dialysate flow or treatment time, approving emergency water sources, or judging dialysis-water quality. Reject reuse requires approved nonclinical use, water characterization, cross-connection protection, and engineering oversight.
Pearls & Pitfalls
Reject is not spent dialysate
RO reject is concentrated feed water discarded by the membrane before treatment. Used dialysate is clinical effluent and is excluded from this estimate.
Reuse does not shrink RO feed
Reusing reject replaces fresh water used elsewhere in the facility. The RO still draws the same feed; the saving appears on the building meter, not the RO meter.
Pitfalls
(1) Mixing meter intervals or boundaries. (2) Forcing a mass balance that does not close; recycle loops and bypass flows need their own meters. (3) Claiming displacement without matched demand, storage, and timing. (4) Treating the KDIGO-reported 20%–80% reject range as a default.
Dialysis RO Water & Reject-Reuse Estimator
Enter feed and permeate to begin.
⚕ Water accounting and a modeled reuse estimate only. It does not set treatment parameters, approve emergency water, or replace engineering and infection-prevention review.
Evidence & References
Formula
| Quantity | Formula |
|---|---|
| Recovery | P ÷ F (if F > 0) |
| Reject (estimate) | F − P, only for a reconciled simple boundary |
| Mass-balance residual | F − P − R, flagged beyond the facility tolerance; reuse suppressed if flagged |
| Per session | F ÷ N and R ÷ N (if N > 0) |
| Modeled displacement | min(R × a, D) or min(measured capture, D) |
| Tariff sensitivity | displaced L ÷ 1,000 × PHP per m³ (gross) |
References
- Barraclough KA, Berman-Parks N, Jha V, et al. Green dialysis: environmentally sustainable care, growth, and innovation — conclusions from a KDIGO Controversies Conference. Kidney Int. 2026;110(1):42-60. doi:10.1016/j.kint.2026.01.015
- Department of Health (Philippines). Administrative Order No. 2013-0003: Implementing guidelines in the analysis, monitoring and maintenance of water used in dialysis facilities. 2013.