- Enter access blood flow (Qa) in mL/min and, if available, cardiac output (CO) in L/min, with the date and method of each.
- For a trend, add a baseline Qa with its date, measured by the same method under comparable conditions.
- Record symptoms. The worksheet reports ratios and changes; it does not label the result normal or abnormal.
Do not enter the blood pump flow (Qb) set on the machine as Qa: Qb is the extracorporeal circuit flow, Qa is flow through the access itself.
When to Use
When a patient with an arteriovenous access has breathlessness, edema, or rising cardiac filling pressures without another clear cause, or when a high-flow access is being considered for flow reduction. It organizes measurements for the cardiology and access team.
Access Flow and Cardiac Burden Worksheet
⚕ Qa/CO (%) = (Qa ÷ 1,000) ÷ CO × 100. Change (%) = (current − baseline) ÷ baseline × 100. Descriptive only; no diagnostic category is assigned.
Formula & Evidence
Basile and colleagues (96 patients with fistulas, ultrasound dilution) found higher Qa in upper-arm fistulas (mean 1.58 versus 0.95 L/min); Qa of 2.0 L/min or more identified the 10 patients classified with high-output cardiac failure (sensitivity 89%, specificity 100%) and a Qa/CO ratio of 20% or more had sensitivity 100% and specificity 74.7%, in that same small cohort. These are observational signals for assessment, not validated diagnostic thresholds for all dialysis populations. Acceptance cases: Qa 2,000 with CO 6 = 33.3%; Qa 600 to 450 = −25%.