Using the series to train water room staff
Use the twelve modules as guided study alongside hands-on training on your own system. A practical sequence is one module a week, each followed by supervised practice of its checklist in the water room.
Figure 1 · Suggested sequence
- Walk the systemTrace every pipe from feed water to drain with the drawing in hand.
- Study the moduleRead, work the case, and take the self-assessment.
- DemonstrateThe preceptor performs the skill and explains each step.
- PracticeThe learner performs it under observation, several times.
- Sign offRecord competence on the unit’s own form.
The medical director should approve the limits, action levels, and procedures the learner is taught.1
Proposed completion criteria
- All twelve self-assessments attempted, with missed questions reviewed.
- Every Critical step on every checklist demonstrated satisfactorily on the unit’s own system.
- At least one full disinfection and one microbiological sampling round performed under observation.
- One out-of-range scenario handled in simulation, including the hold, the report, and the record.
- The four capstone stations in Module 12, with all critical steps satisfactory.
These are proposals for the facility to adapt.
Written examination. The end-of-course examination draws 30 questions at random from the question banks of all modules. A proposed pass mark is 80 percent, with missed questions reviewed against their modules before a second sitting. For a supervised sitting, the learner takes it on a unit device with the preceptor present and the preceptor records the score.
Skills checklists for every module
Print this page for a full set. Each checklist starts on a new sheet. Steps marked Critical must all be satisfactory before sign-off; a missed critical step is repeated with the preceptor. Replace any step that does not match your equipment or approved procedure.
Module 1: Why Dialysis Water Is Different
Tracing the system and explaining the risks. Materials: the system drawing, the unit’s list of limits, and the water room itself.
| Observable step | Satisfactory | Needs practice | Not observed |
|---|---|---|---|
| Traces the water path from the feed line to the loop return on the drawing and in the room. | |||
| Names what each stage removes. | |||
| CriticalStates the unit’s limit and action level for total chlorine, bacteria, and endotoxin. | |||
| CriticalExplains why a normal reverse osmosis reading does not prove the absence of chlorine. | |||
| CriticalIdentifies who must be told when water quality is in doubt, and how to reach them. | |||
| Locates the written procedures and the log for the water system. |
Learner ____________________ Preceptor ____________________ Date __________
Module 2: Source Water and System Design
Reading the system drawing and the feed water report. Materials: the system drawing, valve tags, the most recent feed water analysis, and the design-stage analysis.
| Observable step | Satisfactory | Needs practice | Not observed |
|---|---|---|---|
| CriticalNames the unit’s water source or sources and how the unit is told of a change. | |||
| Traces the drawing from feed to drain and matches each numbered valve to its tag. | |||
| CriticalStates the normal position of every bypass valve and shows how each is locked or tagged. | |||
| Reads five lines of the feed water report and says which component each affects. | |||
| Compares the current report with the design-stage report and points out any change. | |||
| Explains what would raise the unit’s peak demand. | |||
| CriticalStates who approves use of the system after a major change. |
Learner ____________________ Preceptor ____________________ Date __________
Module 3: Sediment Filtration and Softening
Filter and softener round with a hardness test. Materials: the log, a hardness test kit, the softener manual, and the system drawing.
| Observable step | Satisfactory | Needs practice | Not observed |
|---|---|---|---|
| Reads both gauges on each filter, calculates the pressure drop, and compares it with the clean value. | |||
| States the pressure drop at which the unit backwashes or changes each filter. | |||
| CriticalChecks the backwash and regeneration timers against the real time. | |||
| Checks the salt level and tests for a salt bridge. | |||
| CriticalPerforms a hardness test at the correct port and records the value with its unit. | |||
| CriticalStates the action for a hardness result above the unit’s limit. | |||
| Finds the point of pressure loss in a simulated low-pressure problem by reading gauges in order. |
Learner ____________________ Preceptor ____________________ Date __________
Module 4: Carbon Tanks and Chlorine Control
Total chlorine test and daily log. Materials: the unit’s total chlorine test kit, a timer, the sample ports, and the log.
| Observable step | Satisfactory | Needs practice | Not observed |
|---|---|---|---|
| Confirms the system has been running for the stated time before sampling. | |||
| CriticalChecks that the kit is a total chlorine test and is in date. | |||
| CriticalSelects the correct sample port and flushes it for the stated time. | |||
| CriticalFollows the kit instructions exactly, including the reading time. | |||
| Records the value, unit, port, time, and name. | |||
| CriticalStates the correct action for a result above the limit after the worker tank and after the polisher. | |||
| Completes the remaining daily readings with numbers and units and signs the release. | |||
| Points out one trend in the past month’s log and says whom to tell. |
Learner ____________________ Preceptor ____________________ Date __________
Module 5: Reverse Osmosis
Reverse osmosis readings, calculations, and alarm response. Materials: the unit’s display and gauges, the log with at least one month of readings, a calculator, and the alarm list from the manual.
| Observable step | Satisfactory | Needs practice | Not observed |
|---|---|---|---|
| Confirms pretreatment status and valve positions before start-up. | |||
| Waits for steady running before recording. | |||
| Records all pressures, flows, conductivities, and temperature with units. | |||
| CriticalCalculates percent rejection and percent recovery correctly. | |||
| Compares with last month and describes any pattern and its likely cause. | |||
| CriticalStates the action for each alarm on the unit’s alarm list. | |||
| CriticalReports a simulated high product conductivity alarm to the charge nurse with stations affected. | |||
| CriticalExplains why conductivity does not replace chlorine testing or cultures. |
Learner ____________________ Preceptor ____________________ Date __________
Module 6: The Distribution Loop, Storage, and Final Filters
Loop walk-down and dead-leg hunt. Materials: the loop drawing, a flashlight, the log, and the lamp and filter records.
| Observable step | Satisfactory | Needs practice | Not observed |
|---|---|---|---|
| States whether the system is direct feed or tank-fed and traces the loop from supply to return. | |||
| Records loop supply and return pressures and compares with previous readings. | |||
| CriticalConfirms the loop runs continuously, including overnight. | |||
| CriticalIdentifies every unused outlet, capped branch, and spare hose coil and reports them. | |||
| Checks ultraviolet lamp hours or intensity and the date of the last sleeve cleaning. | |||
| Reads the pressure drop across the ultrafilter and its replacement date. | |||
| Inspects the storage tank lid and vent filter, if a tank is fitted. | |||
| CriticalExplains how machine hoses are disinfected and when they are replaced in this unit. |
Learner ____________________ Preceptor ____________________ Date __________
Module 7: Concentrates and Central Delivery
Mixing, labeling, and verifying a bicarbonate batch. Materials: bicarbonate powder, a cleaned mixing container, dialysis water, the unit’s verification instrument, labels, and the batch log.
| Observable step | Satisfactory | Needs practice | Not observed |
|---|---|---|---|
| CriticalConfirms the container has been cleaned and disinfected by procedure and holds no old solution. | |||
| CriticalUses dialysis water and fills to the correct mark. | |||
| Checks powder type, lot, and expiry and adds the stated quantity. | |||
| Mixes gently until dissolved without over-mixing. | |||
| CriticalVerifies the batch by the unit’s method and states the acceptable range. | |||
| CriticalLabels with contents, time mixed, discard time, lot numbers, and name. | |||
| Records the batch in the log. | |||
| Reads an acid concentrate label and states formula, proportioning system, lot, and expiry. | |||
| CriticalStates the action for a batch that fails verification. |
Learner ____________________ Preceptor ____________________ Date __________
Module 8: Daily Monitoring, Logs, and Trends
Start-of-day round and release. Materials: the log sheet with baselines and action points, test kits, and the system itself.
| Observable step | Satisfactory | Needs practice | Not observed |
|---|---|---|---|
| Allows the system to reach steady running before testing. | |||
| CriticalPerforms the total chlorine test correctly at the correct port. | |||
| Records pretreatment readings and checks timers and salt. | |||
| Records reverse osmosis readings and calculates rejection. | |||
| Records loop readings and checks lamp and filter status. | |||
| Checks concentrate batches and tank levels. | |||
| CriticalCompares each reading with its baseline and action point and reports any drift. | |||
| CriticalEnters numbers with units, time, and name, with no blanks. | |||
| CriticalStates the release to the charge nurse in words and signs it. |
Learner ____________________ Preceptor ____________________ Date __________
Module 9: Chemical Analysis and Microbiological Sampling
Aseptic sampling and a documented disinfection. Materials: laboratory sample containers, gloves, the sampling plan, the disinfection procedure, the safety data sheet, and residual test strips.
| Observable step | Satisfactory | Needs practice | Not observed |
|---|---|---|---|
| Labels containers before sampling with point, date, time, and name. | |||
| Performs hand hygiene and wears clean gloves. | |||
| CriticalFlushes the sample port for the stated time. | |||
| CriticalCollects the sample without touching the inside of the cap or container. | |||
| Stores and sends the sample as the laboratory requires. | |||
| CriticalConfirms no patient is on dialysis and posts warning signs before a chemical disinfection. | |||
| CriticalWears the protective equipment listed on the safety data sheet. | |||
| Verifies disinfectant at the farthest point and observes the full dwell time. | |||
| CriticalRinses and obtains a negative residual test at every outlet before release. | |||
| Records concentrations, times, results, and names. |
Learner ____________________ Preceptor ____________________ Date __________
Module 10: Disinfection and Biofilm Control
A documented loop disinfection. Materials: the disinfection procedure, the approved disinfectant and its safety data sheet, protective equipment, concentration and residual test strips, warning signs, and the disinfection record.
| Observable step | Satisfactory | Needs practice | Not observed |
|---|---|---|---|
| CriticalConfirms no patient is on dialysis, tells the charge nurse, and posts warning signs. | |||
| CriticalReads the safety data sheet and wears the stated protective equipment. | |||
| CriticalChecks stock strength and expiry and calculates the dilution correctly. | |||
| Doses and circulates by procedure. | |||
| CriticalProves working concentration at the farthest point before starting the dwell time. | |||
| Observes the full dwell time. | |||
| CriticalRinses and obtains a negative residual test at every outlet, including unused ones. | |||
| Records concentrations, times, results, and names, and releases the system in writing. | |||
| CriticalStates the first aid for skin and eye contact with the disinfectant in use. |
Learner ____________________ Preceptor ____________________ Date __________
Module 11: Maintenance, Component Replacement, and Validation
Cartridge change with post-change checks. Materials: a replacement cartridge of the correct type, the housing tool, the drawing and valve list, the log, and a service report form.
| Observable step | Satisfactory | Needs practice | Not observed |
|---|---|---|---|
| CriticalAgrees the timing with the charge nurse and confirms no patient is on dialysis. | |||
| Records valve positions and pressure readings before starting. | |||
| CriticalIsolates and depressurizes the housing safely. | |||
| Fits the correct cartridge and seals the housing without contaminating it. | |||
| Flushes to drain and checks for leaks. | |||
| CriticalReturns every valve to its normal position and confirms against the valve list. | |||
| Records the new clean pressure drop and writes the installation date on the housing. | |||
| CriticalStates the post-change checks for carbon, membrane, and pipe work. | |||
| CriticalCompletes the record and releases the system in writing. |
Learner ____________________ Preceptor ____________________ Date __________
Module 12: Failures, Emergencies, and Capstone
Out-of-range drill. Materials: a simulated log with a chlorine result above the limit, hold tags, the unit’s contingency plan, and an incident form.
| Observable step | Satisfactory | Needs practice | Not observed |
|---|---|---|---|
| CriticalRecognizes the out-of-range result without prompting. | |||
| CriticalPlaces a hold and prevents new treatment starts. | |||
| CriticalReports to the charge nurse within two minutes with the result, ports, time, and whether patients are connected. | |||
| CriticalDoes not replace the first result with a better repeat result. | |||
| Lists possible causes without claiming one as fact. | |||
| States the tests required before release and who may release. | |||
| Completes the incident record with facts, times, and names. | |||
| Locates the contingency plan, the reserve volume figure, and the tanker supplier’s contact. |
Learner ____________________ Preceptor ____________________ Date __________