Renal Technician Program · Module 9 of 12

Chemical Analysis and Microbiological Sampling

Chemical, culture, and endotoxin tests; limits and action levels; collecting and sending samples; and responding to each kind of result.

PublishedNailathalaGipatikPepalwal: ReferencesMga SanggunianMga TinubdanReng Reperensya: 4 Audience: Dialysis water treatment technicians and trainees · chief technicians, head nurses, and medical directors who supervise the water system Read timeOras ng pagbasaOras sa pagbasaOras ning pamamasa:
Gloved hands collecting water from a sample port into a sterile container.
Renal Care Academy›Water Treatment System Technician›Module 9 of 12 Study series for dialysis water treatment technicians
🛠️

Study series for dialysis water treatment technicians

This module is study material for dialysis water treatment technicians. It explains the reasons behind safe practice; it does not replace your unit’s approved procedures, the manufacturer’s instructions for your equipment, or the limits your facility has adopted. Limits quoted here come from the international standard and are given so the numbers make sense; your facility’s written limits and action levels are the ones you work to. Every case is fictional.

After this module you should be able to

Pre-testBefore you read, answer the 5 self-assessment questions and note your score. After the module, select New set of questions and compare. The difference is what you learned; the questions you still miss are what to reread.

The big picture, in plain words

In plain wordsThink of a blood test at a clinic. The nurse draws one tube, and the doctor may make decisions on that result for months. If the tube is mislabeled, left in a hot car, or sent to a laboratory that runs the wrong test, the patient has given blood for nothing, and nobody may notice that the answer is wrong.

The picture has a limit. A patient can come back tomorrow for a repeat. A water sample is usually taken once a month, so a careless sample wastes a whole month of protection. Some contaminants cannot be tested for in the water room at all. Metals and other chemicals need a laboratory, and bacteria need days to grow on a plate. Your work in this module is to get a true answer from the laboratory and to know what to do with it.

Figure 1 · Key terms for this module

Chemical analysisA laboratory test for a list of metals and other chemicals.
CultureGrowing bacteria from a sample so they can be counted.
Endotoxin testA laboratory test for fragments of bacterial cell walls.
Action levelA result below the limit at which the unit must already take action.
Aseptic techniqueWorking so that you add no germs to the sample.
Chain of custodyThe record of who handled a sample and when.

Case: A high count from one port

It is Friday afternoon. The water technician, Ms. Jessa Mendoza, 31, opens the monthly laboratory report. It shows 280 colony-forming unit (CFU)/mL at the mid-loop sample port. The loop start and the loop return, collected by her on the same morning, are both under 10. Last month all three points were under 10.

She walks to the mid-loop port. Its cap is cracked, and it sits behind a treatment chair where it is often bumped when the chair is moved.

Before reading on, decide:

Keep your answers in mind. The lesson gives you what you need, and the case review near the end shows one sound response.

Three kinds of laboratory test

Chemical analysisBacterial cultureEndotoxin test
Looks forAluminum, fluoride, nitrate, sulfate, calcium, magnesium, sodium, potassium, heavy metals, and others on the standard’s listLive bacteria that can grow on a plateFragments of bacterial cell walls, from live or dead bacteria
Usual frequencyAt validation, then at least yearly, and after a change of source or membraneCommonly monthly; more often after a high result or a system changeCommonly monthly, with the cultures
Sample taken fromProduct water, after the reverse osmosis unitLoop start, loop return, tank outlet, and machines in rotationThe same points as cultures
Result arrives inDays to weeksAbout a weekHours to days
ContainerAs the laboratory supplies; metals need special bottlesSterileSterile and endotoxin-free

Each test answers a different question, which is why none can replace another. A clean culture does not rule out endotoxin, because endotoxin comes from dead bacteria as well as live ones. A clean chemical analysis says nothing about bacteria.

Frequencies are set by the unit’s monitoring plan and the regulations it follows.1,2 The plan lists every sample point, how often each is sampled, and who reviews the results. If your unit has no written plan, raise it with the medical director.

Limits and action levels

Three rising tiers of water purity: dialysis water, standard dialysis fluid, and ultrapure dialysis fluid.
Figure 2. Three purity levels. The limits for bacteria and endotoxin get stricter as the fluid gets closer to the patient’s blood.
FluidBacteria: limitBacteria: typical action levelEndotoxin: limitEndotoxin: typical action level
Dialysis waterLess than 100 CFU/mL50 CFU/mLLess than 0.25 endotoxin unit (EU)/mL0.125 EU/mL
Standard dialysis fluidLess than 100 CFU/mL50 CFU/mLLess than 0.5 EU/mL0.25 EU/mL
Ultrapure dialysis fluidLess than 0.1 CFU/mL—Less than 0.03 EU/mL—

The limits are those of the international standard. Action levels are typically set at half the limit, and the values written in your facility’s own procedure are the ones that apply.3,4 Many units sample monthly, and more often after a high result, a repair, or a new installation.

Gauge showing the bacteria limit and action level for dialysis water.
Figure 3. Limit and action level. Gauge showing the bacteria limit and action level for dialysis water.

The method matters as much as the number. Water bacteria are used to living on very little food, so they grow slowly. Laboratories therefore use low-nutrient culture plates incubated at room temperature for about seven days. A hospital blood-culture method is designed for a different purpose and will undercount them.

RememberA result between the action level and the limit is not a pass. It is the signal to act while there is still a margin.

Taking a sample that can be trusted

A sample is a small cup of water asked to speak for the whole loop. A dirty sample port or a careless hand can add more bacteria than the water itself contains. A falsely high result leads to wasted disinfections and lost trust in the numbers. A falsely low result is worse, because it tells everyone the water is safe when it is not.

Six steps for collecting a water sample aseptically.
Figure 4. Collecting a sample. Six steps for collecting a water sample aseptically.

Figure 5 · Collecting a culture and endotoxin sample

  1. PrepareUse the sterile, endotoxin-free containers the laboratory supplies. Label them first: point, date, time, and your name.
  2. Clean hands, glovesPerform hand hygiene and put on clean gloves.
  3. Flush the portOpen it and let water run for the stated time, commonly at least a minute.
  4. Do not wipe with chemicals unless told toIf the procedure calls for alcohol, let it dry fully. Never use bleach on a sample port.
  5. CollectFill without touching the inside of the cap or the container. Close it at once.
  6. Keep cold and sendRefrigerate the sample and deliver it within the time the laboratory states.

Sample points usually include the start of the loop, the end of the loop (the return, where counts are highest), any storage tank outlet, and a rotating selection of machines for dialysis fluid. Take samples before a scheduled disinfection, when counts are at their highest. The result then shows the worst case that patients were exposed to, which is the number you need to know.

Choosing sample points and timing

Figure 6 · Where to sample, and why

Loop startShows what the reverse osmosis unit and any tank deliver.
Loop returnShows the worst the loop does; it usually has the highest count.
Storage tank outletIf there is a tank, sample it directly.
Machines in rotationDialysis fluid from a different set of machines each month, so that every machine is covered over time.
Bicarbonate systemWhere concentrate is mixed or distributed centrally.
After any repairThe repaired section, before release.

Timing matters as much as place. Collect samples before a scheduled disinfection, not after it, and at a time when the system is running normally. A sample taken the morning after disinfection will look excellent. It will tell you nothing about what patients were exposed to the day before.

System outline marking the usual microbiological sample points.
Figure 7. Where to sample. System outline marking the usual microbiological sample points.
Handling stepRuleWhy
LabelingBefore filling: point, date, time, and nameUnlabeled samples get mixed up
StorageRefrigerated, not frozenWarmth lets bacteria multiply; freezing kills them
Transport timeWithin the laboratory’s stated window, commonly 24 hoursThe longer a sample waits, the further its result drifts
Request formStates that the sample is dialysis water, and names the method requiredA general hospital method undercounts water bacteria
RecordWho collected it, who received it, and whenSo that a strange result can be traced

Water bacteria grow slowly and prefer poor conditions, because they live in water with almost no food in it. The standard therefore calls for low-nutrient culture media incubated at room temperature for about seven days. A laboratory that uses a blood-culture method at body temperature for two days will report falsely low counts. Ask your laboratory which method it uses.3

Reading the report and responding

ResultWhat it meansResponse
Below the action level at all pointsThe system is under controlFile it, chart it, and continue the schedule
Rising month to month but below the action levelBiofilm is developingReport the trend; consider bringing disinfection forward; look for dead legs
At or above the action level, below the limitAct now, while there is a marginReport to the medical director; disinfect; resample; investigate the cause
Above the limitThe water or fluid is out of specificationReport immediately; the medical director decides about treatment; disinfect, find the cause, and resample until the results are acceptable
One point high, its neighbors lowPossibly a contaminated sample or portReport; resample that point promptly with careful technique; inspect the port; do not dismiss it without a repeat
Endotoxin high, culture lowDead bacteria or their fragments; biofilm upstream, or a recent disinfectionTreat it as a real finding; check the ultrafilters and the upstream stages
A chemical above its limitThe membrane or a pretreatment stage is not removing it, or the source has changedReport immediately; review rejection, the feed water, and recent changes
RememberEvery laboratory report is read, signed, and dated by a named person on the day it arrives. A report filed unread is the same as no test at all.

Real situations: problems and solutions

These are fictional but typical. For each one, cover the answer, decide what you would do, then compare. Where a step depends on local rules, your unit’s procedure and your supervisor’s direction come first.

A count that makes no sense

What you see
The laboratory report shows 300 CFU/mL at one sample point, while the points before and after it on the loop are under 10.
What is likely going on
The sample was probably contaminated at the port or while it was being collected.
What to do
Report it, resample that point promptly with careful technique, and inspect the port. Do not dismiss the result until the repeat is back.
Why it matters
A single high result may be technique. A repeat high result is a real problem.

Samples left in the car

What you see
The culture samples taken at 8:00 in the morning sat in a parked car in the midday heat for five hours before they reached the laboratory.
What is likely going on
Bacteria multiply in warm samples, so the results may be falsely high. Some bacteria may die instead, giving falsely low results.
What to do
Tell the laboratory and the person responsible for the system, and take new samples.
Why it matters
A result is only as good as the sample’s journey to the laboratory.

The laboratory used the wrong method

What you see
The monthly culture report from a new laboratory reads “no growth at 48 hours, 37 °C.”
What is likely going on
This is a clinical method. Water bacteria may not have had the time or the right conditions to grow.
What to do
Tell the medical director the result may be falsely reassuring, and arrange with the laboratory for the dialysis water method.
Why it matters
A string of perfect results from the wrong method hides a real problem.

A report nobody signed

What you see
Filing last month’s papers, you find a chemical analysis showing aluminum above its limit. It has no reviewer’s signature.
What is likely going on
The result was never acted on.
What to do
Take it to the medical director today, together with the reverse osmosis rejection trend for that period.
Why it matters
A laboratory result protects no one until someone reads it.
Field notes
  • Label the container before you fill it.
  • Take samples before disinfection, never just after it.
  • Keep every laboratory report in date order, and plot the counts so the trend is visible.
  • After a chemical disinfection, test every outlet for residual, not only the nearest one.

Back to the case: A high count from one port

Is a true loop count of 280 likely? No. Water that reads under 10 at the start of the loop and under 10 at the return, on the same morning, is unlikely to carry 280 in between. A cracked cap on a port that is often bumped makes contamination of the sample, or of the port itself, the probable explanation.

Can you set the result aside? No. A dirty port is still part of the loop, and it can seed the water that flows past it. A probable explanation is not yet a proven one.

What to do and report this week. Report the result and your reasoning to the chief technician and the medical director. Replace the cap, and clean and disinfect the port by procedure. Resample all three points with careful technique. Move or guard the port so it is not bumped again. If the repeat is clean, Ms. Mendoza records the investigation and closes it. If it is high again, treat it as a real loop problem and disinfect.

Skills checklist: Aseptic sampling and a documented disinfection

Materials: laboratory sample containers, gloves, the sampling plan, the disinfection procedure, the safety data sheet, and residual test strips. Your preceptor rates each step as satisfactory, needs practice, or not observed. Steps marked Critical must all be satisfactory before sign-off. This checklist supports supervised learning; your unit’s own procedure and competency form are the authority. A printable version of every checklist is on the trainer page.

Placement activity. Take part in one full sampling round and one full loop disinfection under observation, and file the laboratory report when it returns.

Check your understanding

Each visit draws 5 questions at random from this module’s 8-question bank. Choose the single best answer for each, then select Check answers; an explanation appears under the best answer. Select New set of questions for a different mix. A good target is 4 of 5 on two different sets. Nothing you select is sent anywhere; scoring happens in your browser.

Question 1 of 8In the international standard, what is the bacteria limit for dialysis water?
Question 2 of 8A culture result lies between the action level and the limit. What does this mean?
Question 3 of 8When are routine culture samples best taken?
Question 4 of 8Why is the sample port flushed before collecting?
Question 5 of 8How often is chemical analysis of product water done, at a minimum, in most monitoring plans?
Question 6 of 8A laboratory cultures dialysis water at body temperature for 48 hours. What is the likely effect?
Question 7 of 8Endotoxin is high but the culture is low. How should this be treated?
Question 8 of 8How should culture samples be stored before transport?

Read next on this site

Glossary & abbreviationsTalahulugan at mga daglatTalaan sa mga pulong ug daglatTalatinigan ampo reng daglat terms used on this page

Abbreviations

CFU
Colony-forming unit — The count of viable bacteria in a water sample.
EU
Endotoxin unit — The measure of bacterial endotoxin in a water sample.

Terms in plain words

Monitoring plan
The unit’s written list of what is tested, where, how often, and who reviews it.
Low-nutrient medium
A culture plate with little food, suited to slow-growing water bacteria.
Out of specification
Outside the limits your unit works to.
ReferencesMga SanggunianMga TinubdanReng Reperensya 4 sources
  1. International Organization for Standardization (ISO). (2024). Preparation and quality management of fluids for haemodialysis and related therapies. Part 1: General requirements (ISO 23500-1:2024).
  2. Department of Health (DOH). (n.d.). Guidelines on the monitoring of water used for dialysis. https://elibrary.judiciary.gov.ph/thebookshelf/showdocs/10/69036
  3. International Organization for Standardization (ISO). (2024). Preparation and quality management of fluids for haemodialysis and related therapies. Part 3: Water for haemodialysis and related therapies (ISO 23500-3:2024).
  4. International Organization for Standardization (ISO). (2024). Preparation and quality management of fluids for haemodialysis and related therapies. Part 5: Quality of dialysis fluid for haemodialysis and related therapies (ISO 23500-5:2024).
Dr. William Gregory M. Rivero, MD

William Gregory Rivero, MD, FPCP, DPSN

Internal Medicine · Nephrology · Nutrition · Philippines · PRC 0105184

Educational study material for dialysis water treatment staff. Does not replace the medical director’s direction, approved unit procedures, manufacturer instructions, the standard edition your facility follows, DOH licensing requirements, or supervised training.

QR code — scan to save Dr. Rivero's contact info

Scan and save