Renal Technician Program · Module 7 of 12

Concentrates and Central Delivery

What acid and bicarbonate concentrates do, reading labels, mixing and verifying bicarbonate, and central delivery systems.

PublishedNailathalaGipatikPepalwal: ReferencesMga SanggunianMga TinubdanReng Reperensya: 1 Audience: Dialysis water treatment technicians and trainees · chief technicians, head nurses, and medical directors who supervise the water system Read timeOras ng pagbasaOras sa pagbasaOras ning pamamasa:
A technician reading the label on a concentrate container against an order sheet.
Renal Care Academy›Water Treatment System Technician›Module 7 of 12 Study series for dialysis water treatment technicians
🛠️

Study series for dialysis water treatment technicians

This module is study material for dialysis water treatment technicians. It explains the reasons behind safe practice; it does not replace your unit’s approved procedures, the manufacturer’s instructions for your equipment, or the limits your facility has adopted. Limits quoted here come from the international standard and are given so the numbers make sense; your facility’s written limits and action levels are the ones you work to. Every case is fictional.

After this module you should be able to

Pre-testBefore you read, answer the 5 self-assessment questions and note your score. After the module, select New set of questions and compare. The difference is what you learned; the questions you still miss are what to reread.

The big picture, in plain words

In plain wordsThink of a carinderia that makes its own iced tea. The water comes from a clean filter, but the drink the customer tastes depends on which powder was scooped in, and how much. If two powders sit in identical tubs, the cook can pick up the wrong one and the drink will still look the same in the glass.

The picture has a limit. A wrong iced tea only tastes wrong. Clean water is only half of dialysate. The other half is two concentrates that the machine dilutes with water: an acid concentrate carrying most of the salts, and a bicarbonate concentrate carrying the buffer. The patient’s blood moves toward whatever the dialysate contains, so a mistake here is a medication error. In many units the water technician mixes and distributes these concentrates, and that is why the work belongs in this track.

Figure 1 · Key terms for this module

Acid concentrateThe concentrate holding sodium, potassium, calcium, magnesium, chloride, a small amount of acid, and usually glucose.
Bicarbonate concentrateThe concentrate holding sodium bicarbonate, the buffer that corrects acid in the blood.
Proportioning ratioThe fixed proportions in which a machine mixes acid concentrate, bicarbonate concentrate, and water.
BatchOne mixed quantity, with its own label and record.
Central deliveryTanks and pipes that carry concentrate to wall outlets at each station.
Use timeHow long a mixed concentrate may be kept before it is discarded.

Case: Two acid formulas on one shelf

The unit stocks two acid concentrates from the same supplier, in identical white containers. They differ only in one number on the label: one gives a dialysate potassium of 2, the other 3. On a busy Tuesday morning a new aide, Mr. Kevin Manalo, 22, restocks the store, and containers of both kinds end up in one row.

At 09:30 the nurse, Ms. Aileen Torres, checks the container at station 5 against the order for her patient, Mrs. Felicidad Ocampo, 70. It does not match. The machine’s conductivity has been normal the whole time, and no alarm has sounded. She calls you to the station.

Before reading on, decide:

Keep your answers in mind. The lesson gives you what you need, and the case review near the end shows one sound response.

What each concentrate does

Why two concentrates instead of one? If all the salts were put into one concentrate with bicarbonate, calcium and magnesium would react with the bicarbonate and form a solid, like the white scale inside a kettle. So the ingredients are kept apart until the last moment, when the machine mixes them into a large volume of water.

Acid and bicarbonate concentrates combining with treated water in the machine.
Figure 2. Two concentrates, one dialysate. Acid and bicarbonate concentrates combining with treated water in the machine.
Acid concentrateBicarbonate concentrate
ContainsSodium, potassium, calcium, magnesium, chloride, a small amount of acetic or citric acid, usually glucoseSodium bicarbonate, sometimes with sodium chloride
Varies between patientsYes: potassium and calcium come in several formulasUsually one formula; the machine can vary the amount delivered
Supports bacterial growthNo; it is too acidic and saltyYes, readily, once mixed with water
Supplied asReady-made liquid in containers or bulk; or powder mixed on sitePowder cartridges on the machine; or powder mixed with dialysis water on site; or ready-made liquid
Main riskWrong formula connectedContamination; loss of strength if kept too long or mixed too hard

Machines mix the two concentrates and water in a fixed proportion, and several different proportioning systems are in use worldwide. Each concentrate is made for one system only, and that system is stated on the label. A concentrate made for a different system will give the wrong dialysate, even though each container is correct on its own. The biomedical technician configures the machine. Your part is to make sure that only matching concentrates reach it.1

RememberConductivity measures the total amount of salt. Two acid formulas with different potassium give almost the same conductivity, so the machine cannot tell them apart. The label is the only safeguard.

Reading labels and organizing the store

On the labelWhy you check it
Product type: acid or bicarbonateThe two connect to different ports
Formula: potassium and calcium levelsIt must match the physician’s order for that patient
Proportioning systemIt must match the machines in your unit
Lot number and expiry dateSo a recalled batch can be traced; expired product is never used
Seal and container conditionA broken seal or a damaged container is rejected

Figure 3 · A store that prevents mix-ups

One formula per shelfWith a large shelf label showing the potassium and calcium.
Color or tag codingA unit-wide code for each formula, applied on receipt.
Look-alikes apartNever side by side.
First in, first outOldest stock at the front.
Trained restockingOnly staff who know the formulas put stock away.
Quarantine areaFor damaged, expired, or unidentified containers.

Mix-ups happen in the store long before they happen at the station, so the store is where you prevent them. By international convention, acid concentrate connectors and containers are marked red and bicarbonate blue, and the connectors are shaped so that one will not fit the other’s port. Treat any connector that has been forced or modified as a hazard, and report it.

Mixing bicarbonate from powder

Steps for mixing, verifying, and labeling bicarbonate concentrate.
Figure 4. Mixing bicarbonate safely. Steps for mixing, verifying, and labeling bicarbonate concentrate.

Figure 5 · Preparing a batch

  1. PrepareClean hands and clean gloves. Use a container or mixer that has been cleaned and disinfected by procedure.
  2. WaterAdd dialysis water to the mark. Never tap water.
  3. PowderAdd the whole contents of the stated number of bags. Check the lot and expiry first.
  4. MixMix gently until fully dissolved. Do not whip air into it.
  5. VerifyTest the batch by the unit’s method.
  6. LabelWrite the contents, date and time mixed, discard time, lot numbers, and your name.
  7. RecordEnter the batch in the log.
MistakeWhat it causes
Mixing too hard or for too longCarbon dioxide escapes, the solution becomes more alkaline, and it may behave wrongly in the machine
Wrong amount of water or powderA concentrate that is too strong or too weak; conductivity alarms at every machine
Powder not fully dissolvedA weak concentrate at first and a strong one later; blocked pick-up filters
Topping up an old batch with newBacteria from the old batch seed the new one
Keeping a batch past its use timeBacteria grow and the concentrate loses strength
A dirty container, or the lid left offContamination

Bicarbonate in water behaves differently from the acid concentrate. Once it is mixed with water, bacteria grow in it readily, and it gives up carbon dioxide if it is mixed too hard. Both facts shape the rules below.

Verify each batch before you release it. Units use conductivity, a density check, or pH, as the concentrate manufacturer advises. The acceptable range is in the manufacturer’s instructions and in your procedure. A batch outside the range is discarded. It is not corrected by adding a bit more powder or water, because you no longer know what is in the container.

The use time for mixed bicarbonate is set by the manufacturer and is short, commonly within a day. Write the discard time on the label when you mix. Empty, rinse, and disinfect the container at the frequency your procedure states.

Stop and reportNever add fresh bicarbonate to a container that still holds old solution, and never use a batch with no label.

Central concentrate systems

Larger units pipe concentrates from tanks to wall outlets at each station. This removes heavy lifting and the handling of single containers. It also means that one error in the tank reaches every station at the same moment, so the safeguards move from the station to the tank room.

Central acid and bicarbonate concentrate tanks feeding separate lines to station outlets.
Figure 6. Central concentrate delivery. Central acid and bicarbonate concentrate tanks feeding separate lines to station outlets.

Figure 7 · Safeguards in a central system

Separate tanks and linesAcid and bicarbonate never share pipework.
One formula per acid lineEach line and each outlet is labeled with its formula.
Different connectorsAcid and bicarbonate outlets do not accept each other’s hoses.
Filling controlsTrained staff fill the tanks, with a second person checking the product against the tank label.
Level and low alarmsSo that a tank does not run dry in the middle of a shift.
Scheduled disinfectionThe bicarbonate tank and loop are disinfected and rinsed on a schedule; acid lines are not disinfected with bleach.
ProblemHow it showsFirst response
Acid tank filled with the wrong formulaUsually nothing at all on the machinesStop use of that line and tell the charge nurse immediately; the clinical team reviews the patients
Bicarbonate tank running low or drawing airLow conductivity alarms on several machines at onceCheck the tank level and the pick-up; tell the nurse the cause
Bicarbonate batch too weak or too strongConductivity alarms across the unit soon after a new batch goes on lineTake the batch off line, verify it, and discard it if out of range
Contaminated bicarbonate systemHigh cultures in dialysis fluid from several machines; slime or cloudiness in the tank; possibly reactions in patientsReport; disinfect the tank and lines; review the cleaning frequency
White crystals at outlets and connectorsDried bicarbonateClean them off; check for leaks; crystals can block connectors

Real situations: problems and solutions

These are fictional but typical. For each one, cover the answer, decide what you would do, then compare. Where a step depends on local rules, your unit’s procedure and your supervisor’s direction come first.

Conductivity alarms after the new batch

What you see
At 07:10, ten minutes after you put a fresh bicarbonate batch on line, machines across the unit alarm for low conductivity.
What is likely going on
The batch is probably too weak: too much water, or too little powder.
What to do
Take the batch off line and tell the charge nurse the likely cause. Verify the batch, and discard it if it is out of range. Prepare a new one.
Why it matters
A shared supply error reaches every station at the same moment.

No label on the jug

What you see
A full container of clear liquid sits in the mixing area with no label. Nobody on the shift remembers filling it.
What is likely going on
It could be bicarbonate, acid, disinfectant, or water. Nobody can know.
What to do
Do not use it, and do not smell or taste it. Quarantine it and discard it by procedure.
Why it matters
Unlabeled containers are how disinfectant ends up in a concentrate line.

Slime in the mixing tank

What you see
When you empty the bicarbonate tank for its scheduled cleaning, the inside walls feel slippery to your gloved hand.
What is likely going on
Biofilm has formed. The cleaning frequency is not keeping up.
What to do
Report it. Clean and disinfect the tank and lines by procedure, and review the schedule. Ask for dialysis fluid cultures from the machines.
Why it matters
Contaminated bicarbonate bypasses every barrier in the water room.

A forced connector

What you see
At station 8, an acid outlet connector has been filed down so that a different hose will fit.
What is likely going on
Someone defeated the feature that prevents acid and bicarbonate from being swapped.
What to do
Remove it from use, report it, and have the correct connector fitted.
Why it matters
Connector shapes are a safety system, not an inconvenience.
Field notes
  • Read the whole label: type, formula, proportioning system, lot, and expiry.
  • Label every batch when you mix it, including the discard time.
  • Never top up old bicarbonate with new.
  • Store look-alike acid formulas on separate shelves.

Back to the case: Two acid formulas on one shelf

Why the machine did not alarm. Conductivity reflects the total salt content, and the two formulas differ by a very small amount of potassium. The machine read both as normal. Only the label tells them apart.

What should happen at station 5, and whose decision it is. Ms. Torres and the physician decide what to do for Mrs. Ocampo. Your part is to state exactly which product was connected and from what time, and to keep the container so it can be checked. Then check every other station and the store, because the same row supplied them.

What makes a repeat harder, and what to report. Separate shelves with large formula labels; a color tag applied to each container on receipt; restocking only by trained staff; look-alike products never in the same row; and a second check of the container against the order at setup. Report the event as an incident even if Mrs. Ocampo is well. The cause is in the store, not in the person at station 5, and only a report will change the store.

Skills checklist: Mixing, labeling, and verifying a bicarbonate batch

Materials: bicarbonate powder, a cleaned mixing container, dialysis water, the unit’s verification instrument, labels, and the batch log. Your preceptor rates each step as satisfactory, needs practice, or not observed. Steps marked Critical must all be satisfactory before sign-off. This checklist supports supervised learning; your unit’s own procedure and competency form are the authority. A printable version of every checklist is on the trainer page.

Placement activity. Audit the concentrate store with your preceptor: look for look-alike products stored together, expired stock, and unlabeled containers.

Check your understanding

Each visit draws 5 questions at random from this module’s 8-question bank. Choose the single best answer for each, then select Check answers; an explanation appears under the best answer. Select New set of questions for a different mix. A good target is 4 of 5 on two different sets. Nothing you select is sent anywhere; scoring happens in your browser.

Question 1 of 8Why are acid and bicarbonate supplied as two separate concentrates?
Question 2 of 8Two acid concentrates differ only in potassium. Will the machine’s conductivity alarm tell them apart?
Question 3 of 8Which water is used to mix bicarbonate concentrate?
Question 4 of 8Why should bicarbonate not be mixed vigorously for a long time?
Question 5 of 8A bicarbonate container is half full of yesterday’s batch. What is correct?
Question 6 of 8Low conductivity alarms occur on six machines at once shortly after a new bicarbonate batch goes on line. What is the likely cause?
Question 7 of 8By convention, which color marks acid concentrate connectors?
Question 8 of 8A mixed batch fails verification. What do you do?

Read next on this site

Glossary & abbreviationsTalahulugan at mga daglatTalaan sa mga pulong ug daglatTalatinigan ampo reng daglat terms used on this page

Terms in plain words

Buffer
A substance that neutralizes acid. In dialysate the buffer is bicarbonate.
Precipitate
A solid that forms and settles out of a liquid.
Lot number
The manufacturer’s code identifying one production batch.
Density check
Measuring how heavy a liquid is for its volume, as a check on how strong it is.
ReferencesMga SanggunianMga TinubdanReng Reperensya 1 source
  1. International Organization for Standardization (ISO). (2024). Preparation and quality management of fluids for haemodialysis and related therapies. Part 5: Quality of dialysis fluid for haemodialysis and related therapies (ISO 23500-5:2024).
Dr. William Gregory M. Rivero, MD

William Gregory Rivero, MD, FPCP, DPSN

Internal Medicine · Nephrology · Nutrition · Philippines · PRC 0105184

Educational study material for dialysis water treatment staff. Does not replace the medical director’s direction, approved unit procedures, manufacturer instructions, the standard edition your facility follows, DOH licensing requirements, or supervised training.

QR code — scan to save Dr. Rivero's contact info

Scan and save