The blood circuit, the dialyzer, diffusion and ultrafiltration, the machine’s monitors, and what the common alarms mean.
PublishedNailathalaGipatikPepalwal:ReferencesMga SanggunianMga TinubdanReng Reperensya:10Audience: Physicians who cover hemodialysis shifts · attending nephrologists, medical directors, and head nurses who orient themRead timeOras ng pagbasaOras sa pagbasaOras ning pamamasa:
This module is study material for physicians on duty in hemodialysis units. It explains the reasoning behind common decisions; it is not a protocol. At the chairside, follow your unit’s protocol and ask the attending nephrologist for guidance. Doses and cut-off values are set by your unit. Every case is fictional.
Learning objectives
After this module you should be able to
Trace the path of blood from the patient, through the circuit, and back, naming each component in order.
Explain diffusion, ultrafiltration, and convection, and say which one removes which substance.
Describe how dialyzers differ in membrane, surface area, and flux, and what the trials of flux found.
Explain what the arterial, venous, and transmembrane pressure readings mean for the patient.
Name each safety detector and the harm it is built to prevent.
Read the machine screen and respond to a common alarm by examining the patient first.
Pre-testBefore you read, answer the 8 self-assessment questions and note your score. After the module, select New set of questions and compare. The difference is what you learned; the questions you still miss are what to reread.
Start here
The big picture, in plain words
In plain wordsPut a tea bag in a cup of hot water. Color and flavor move out through the paper into the water, and the leaves stay inside the bag. Nothing pushes them. They spread from where there is more to where there is less. A dialyzer works in a similar way. Blood stays on one side of a thin membrane, clean fluid passes on the other side, and wastes cross from the blood into the fluid.
The picture has three limits. A cup of tea soon stops changing, because the water fills with color. The machine keeps fresh fluid flowing, so the movement continues. The dialyzer also removes water by pressure, which a tea bag cannot do. And the blood is outside the body while this happens. Most of the machine exists to make that safe.
Figure 1. The dialyzer does a small part of the kidney’s work, for a few hours at a time.
The nurses and technicians operate the machine. You need to know what each number and each alarm says about the patient in the chair.
Figure 2 · Key terms for this module
Extracorporeal circuitThe tubing and dialyzer that carry blood outside the body and back.
DialyzerThe filter. Blood and dialysis fluid pass on opposite sides of its membrane.
Dialysis fluidThe prepared solution of water and salts that flows through the dialyzer, also called dialysate.
DiffusionMovement of a dissolved substance from a higher concentration to a lower one.
UltrafiltrationMovement of water across the membrane, driven by pressure.
ConvectionRemoval of dissolved substances that are carried along with the water.
Transmembrane pressure (TMP)The pressure difference between the blood side and the fluid side of the membrane.
BypassA safety state in which dialysis fluid goes to the drain instead of to the dialyzer.
Opening case · fictional
Case: The venous pressure alarm
Mr. Ernesto Bautista, 63, started hemodialysis two months ago. His left forearm fistula was first needled three weeks ago. Today he is ninety minutes into a four-hour session, lying under a blanket, half asleep.
His machine sounds a high venous pressure alarm, and the blood pump stops. Liza, his nurse, straightens his arm, clears the alarm, and restarts the pump. It alarms again four minutes later, and again after that. A technician walking past says, “His needle is small. Widen the venous limit so he can finish.” Liza looks at you. Mr. Bautista opens his eyes and says that his arm feels tight.
Before reading on, decide:
What is a high venous pressure alarm telling you about the circuit?
What do you check in the patient before anyone changes an alarm limit?
If the alarm had stayed silent, would that prove that the venous needle was in place?
Keep your answers in mind. The lesson gives you what you need, and the case review near the end shows one sound response.
Lesson · part 1
The blood circuit, step by step
Every pressure reading and every alarm refers to a point on one loop of tubing.
Figure 3. The blood circuit. Blood leaves the access, is pushed by the blood pump through the dialyzer, passes the air trap, and returns to the patient.
Two words need care. The tubing that carries blood away from the patient is called the arterial line, usually marked red. The tubing that returns blood is the venous line, usually marked blue. These names describe direction only. In a patient with a fistula or graft, both needles sit in the same vessel.1
Figure 4 · Follow the blood
AccessBlood leaves through the arterial needle or catheter limb.
Blood pumpA roller squeezes the tubing and pushes blood forward at the prescribed flow.
Anticoagulant lineIf heparin is prescribed, a syringe pump adds it after the blood pump.
DialyzerWastes and water leave the blood here.
Venous chamberIt traps air bubbles. The venous pressure sensor measures here.
Air detector and clampA sensor checks for air. A clamp below it can shut the line at once.
ReturnBlood goes back through the venous needle or catheter limb.
The blood pump divides the circuit into two zones. Ahead of the pump, the pump is pulling, so the pressure in the tubing is below zero. After the pump, it is pushing, so the pressure is above zero.1 This one fact explains two dangers. A loose connection ahead of the pump does not leak blood. It sucks air in. A loose connection after the pump does the opposite and loses blood.
When the pump stops, the blood in the circuit stands still, and still blood in plastic tubing soon clots. This is why the staff act quickly on an alarm.
RememberBefore you read any number on the screen, look at the patient and then at the circuit. Check that every connection is tight, no line is kinked, the venous chamber has a visible level, and the access is uncovered.
Lesson · part 2
Diffusion, ultrafiltration, and convection
The dialyzer does two jobs. It removes dissolved wastes, and it removes water. The machine controls them separately.
Diffusion: the tea bag
Dissolved particles move at random. When a membrane separates two solutions, more particles cross from the crowded side than from the other, until both sides are equal.2 This is diffusion. It needs no pressure, only a difference in concentration.
Figure 5. Diffusion. Small dissolved substances move through the membrane from where there is more to where there is less. Cells and proteins are too large to pass.
Urea and potassium are higher in the blood than in the dialysis fluid, so they leave the blood. Bicarbonate is higher in the dialysis fluid, so it moves the other way, as Module 2 explained. The composition of the dialysis fluid decides the direction for each substance, which is why it is part of the prescription.
Three things set how fast a substance diffuses: the concentration difference, the size of the molecule, and the flows and membrane area.1 Urea and potassium are small and clear well. Larger wastes clear poorly by diffusion. A blood flow below the prescription delivers less dialysis than planned.
Ultrafiltration: wringing the cloth
Think of squeezing grated coconut in a cloth to make gata. Your hands raise the pressure inside the cloth, and liquid is forced through the weave. The solids stay behind. In the dialyzer, the machine makes the pressure on the blood side higher than on the fluid side, and water is pushed out of the blood.1
Figure 6. Ultrafiltration. A pressure difference across the membrane pushes water out of the blood. The water carries dissolved substances with it, which is convection.
The pressure difference across the membrane is the transmembrane pressure (TMP). The nurse does not set this pressure. The nurse enters the volume to remove and the time. The machine removes the programmed volume and displays the TMP that this required.1
The water comes from the plasma. Module 2 explained plasma refilling. The machine removes whatever volume it is told to remove. It does not know whether the patient’s tissues can refill the vessels that fast.
Convection: what the water carries
The coconut milk that comes through the cloth carries whatever was dissolved in it. In the dialyzer, water pushed through the membrane drags dissolved substances along, up to the size that fits through the pores. This is convection.1
Figure 7. Diffusion moves small molecules down a concentration difference. Convection carries larger ones along with the water.
In standard hemodialysis the volume of water removed is small, so diffusion does most of the cleaning.
RememberThe two jobs are separate. A session can remove the planned fluid and still clean poorly, or the reverse.
Lesson · part 3
The dialyzer: fibers, surface, and flux
A dialyzer is a plastic cylinder packed with thousands of fine hollow fibers. The wall of each fiber is the membrane. Blood flows inside the fibers. Dialysis fluid flows around them.1
Figure 8. Inside a dialyzer. Blood runs through thousands of hair-thin hollow fibers; dialysate runs around them. The fiber wall is the membrane.
The two fluids run in opposite directions, which is called countercurrent flow. Blood that has lost most of its urea meets the freshest dialysis fluid, so a concentration difference remains along the whole fiber.1
Figure 9. Inside the dialyzer. Blood runs inside thousands of hollow fibers and dialysis fluid runs outside them in the opposite direction, which keeps a concentration difference along the whole length.
Dialyzers differ in three ways. The membrane material is usually a synthetic polymer. The surface area, which is the total area of the fiber walls, is chosen to fit the patient. The third is flux.
Flux causes the most confusion. It does not describe how well a dialyzer removes urea. A low-flux dialyzer with a large surface can remove urea well. Flux describes the pores. Larger pores pass more water for a given pressure, and they pass middle-sized molecules such as beta-2 microglobulin, a protein that builds up in kidney failure. Albumin and blood cells stay in the blood with both types.1
Figure 10. Low-flux and high-flux membranes. Larger pores pass more water and larger waste molecules. Albumin and cells stay in the blood with both.
What the trials of flux found
Two randomized trials tested whether high flux helps patients live longer. The Hemodialysis (HEMO) Study assigned 1,846 patients on three sessions a week to low-flux or high-flux dialyzers. Death from any cause did not differ significantly between the groups (relative risk with high flux 0.92, 95% confidence interval 0.81 to 1.05).3 The Membrane Permeability Outcome study assigned 738 patients new to hemodialysis and also found no significant difference overall. Among patients who began with a serum albumin of 4 g per dL or lower, a group defined before the trial started, survival was significantly higher with high-flux membranes.4
The 2015 guideline of the United States Kidney Disease Outcomes Quality Initiative (KDOQI) recommends biocompatible membranes and accepts either high-flux or low-flux types.5 In summary, high-flux dialysis removes more middle-sized wastes, and a survival benefit for all patients has not been shown. The HEMO Study enrolled only patients on three sessions a week, so it does not speak directly to twice-weekly schedules.
Hemodiafiltration (HDF) adds a large volume of convection to diffusion and replaces the removed water with sterile fluid made by the machine. In a randomized trial of 1,360 patients, high-dose HDF resulted in fewer deaths from any cause than high-flux hemodialysis over a median of 30 months (hazard ratio 0.77).6 A pooled analysis of five trials with 4,153 patients found a smaller reduction (hazard ratio 0.84).7 The first trial enrolled patients judged able to reach a high convection volume, and HDF needs machines and water quality that not every unit has.
RememberThe choice of dialyzer belongs to the prescription. Check that the dialyzer on the machine is the one ordered for this patient.
Lesson · part 4
The dialysis fluid side and what guards it
The blood circuit is the half of the machine you can see. The other half, inside the cabinet, makes the dialysis fluid.
Figure 11. Units assign machines by hepatitis status. A technician must know which machines and stations are reserved, and never move them between areas.
The machine mixes treated water with two concentrates in fixed proportions. One holds the acid and the salts, including potassium and calcium. The other holds bicarbonate. The machine warms the mixture to about body temperature and sends it to the dialyzer.1
A mistake in mixing cannot be seen, because dilute fluid looks the same as correct fluid. So the machine tests the fluid continuously before it reaches the dialyzer, using two sensors.
Sensor
What it measures
Harm if it failed
Conductivity
How well the fluid carries an electric current, which reflects its total salt content, mostly sodium
Fluid that is too dilute makes red cells swell and burst. Fluid that is too concentrated raises blood sodium
Temperature
The warmth of the fluid entering the dialyzer
Overheated fluid destroys red cells. Cold fluid chills the patient
Destruction of red cells is called hemolysis. Investigators from the United States Centers for Disease Control and Prevention named its most frequent causes during dialysis: chemical contamination, heat, and mechanical injury from blocked or kinked blood lines.8 The two sensors guard against wrong mixing and heat.
When either sensor reads outside its limits, the machine goes into bypass. A valve sends the dialysis fluid straight to the drain, so none of it reaches the dialyzer. The blood pump keeps running, but no dialysis takes place.1 Bypass is the machine protecting the patient.
Conductivity cannot tell potassium from sodium. A wrong concentrate with the right total salt content can pass the sensor. This is why staff check the concentrate label against the prescription at the start of each session.
Stop and callStop and call for a conductivity or temperature alarm with the patient connected that does not clear at once. Do not allow anyone to override it. Examine the patient for chest or back pain, breathlessness, and a change in alertness, follow your unit’s protocol, and call the attending nephrologist. If several machines alarm together, suspect the water or concentrate supply, and call the head nurse and the medical director as well.
Lesson · part 5
What the pressure readings say about the patient
A pressure reading reports how hard the machine must work to move blood through one part of the circuit. Pressures are shown in millimeters of mercury (mm Hg).
Figure 12. Where the two pressures are measured. Arterial pressure is suction ahead of the pump. Venous pressure is the push needed to return blood to the patient.
Reading
Where it is measured
What a change suggests
Arterial pressure
Between the patient and the blood pump. Normally below zero
More negative: the access is not delivering blood easily
Venous pressure
At the venous chamber, after the dialyzer. Normally above zero
High: something resists the return of blood. Low: resistance is lost or flow has dropped
Transmembrane pressure
Calculated from the blood side and fluid side pressures
Rising: fibers are clotting, or fluid removal is fast for this dialyzer
Alarm limits are set around each reading, following your unit’s protocol. Watch the trend as well. A reading that drifts during a session, or differs from the patient’s previous sessions, deserves a look even without an alarm.
Arterial pressure and the blood flow you think you have
The machine calculates blood flow from how fast the pump turns. When the arterial pressure is very negative, the soft tubing in the pump partly collapses, and each turn moves less blood. The screen then shows a higher flow than the patient receives.1 A persistently poor arterial pressure is information about the access, and the attending nephrologist needs to hear it.
Venous pressure and the needle that slips out
It is natural to assume that a venous needle that comes out will trigger the low venous pressure alarm. Often it does not. Most of the venous pressure is created by the narrow needle, not by the patient’s vein. When the needle slips out, blood still flows through it, and the pressure may fall very little.
A study of 99 patients measured the pressure inside the access. With the lower alarm limit set 30 to 40 mm Hg below the running pressure, as was usual practice, the investigators estimated that the alarm could be expected to detect a dislodged needle in nearly all grafts and in only 29% of fistulas.9 In a laboratory simulation, the machine alarm sounded in 8 of 19 simulated dislodgements.10 One study was an estimate from pressure measurements and the other a bench model, so neither gives a true clinical rate. Both point the same way.
A dislodged venous needle can lose a dangerous volume of blood within minutes while the pump keeps running. The defense is visual. Units keep the access and connections uncovered and taped according to protocol. This practice rests on expert recommendation, not on trials.
Stop and callStop and call when a patient is bleeding from the access or the lines, or when the arm is swelling at a needle site. The nurse stops the blood pump and clamps the lines. Apply direct pressure to a bleeding site, assess the patient’s circulation, follow your unit’s protocol, and call the attending nephrologist. A wider alarm limit is never the answer to a repeating pressure alarm.
Lesson · part 6
The safety detectors and what each one prevents
Each detector exists because of one way that blood outside the body can injure a patient.
Figure 13 · What each detector protects against
Arterial pressure sensorA starved pump and unnoticed loss of access flow.
Venous pressure sensorA blocked return path and, less reliably, a disconnected one.
Air detector and venous clampAir entering the patient’s vein.
Blood leak detectorBlood loss into the dialysis fluid through torn fibers.
Conductivity sensorDialysis fluid with the wrong salt content.
Temperature sensorDialysis fluid that is too hot or too cold.
Air
Air enters where the circuit is below zero pressure, ahead of the blood pump. Common entry points are a loose arterial connection and an empty saline bag left open. The venous chamber traps small bubbles. Below it, a sensor checks the tubing. When it detects air, the machine stops the blood pump and shuts the venous clamp.1
The detector protects only a line that is seated in it. Air in the venous line is never pushed into the patient. The staff remove it by the unit’s procedure, and blood is not returned through that line until the air is gone.
Blood leak
If some fibers tear, blood enters the dialysis fluid. A light sensor on the used fluid line detects it and stops the treatment.1 The blood is often too little to see. The staff confirm it by testing the used fluid for blood, following the unit’s procedure.
A blood leak matters for two reasons. The patient is losing blood. And a hole that lets blood out also lets dialysis fluid, which is clean but not sterile, touch the blood. Whether the blood in the circuit is returned depends on the size of the leak and on your unit’s protocol.
Stop and callStop and call for any blood-leak alarm. Do not allow the alarm to be overridden. The session does not continue on the same dialyzer unless the used fluid tests negative for blood and your unit’s protocol allows it. Check the patient’s blood pressure, heart rate, and temperature, follow your unit’s protocol on whether blood is returned, and call the attending nephrologist.
What no detector sees
The machine watches the circuit. It does not watch the patient. It has no sensor for chest pain, drowsiness, or blood running under a blanket. A quiet machine means the circuit is within its limits. It does not mean the patient is well.
RememberThree habits defeat the detectors: overriding an alarm without finding the cause, widening a limit to stop a repeating alarm, and covering the access.
Lesson · part 7
Reading the screen and responding to alarms
You do not need to learn the menus. Find a few items on the screen and compare each with the prescription and with the patient.
Figure 14. The front of the machine and the readings a physician looks for. The layout differs between manufacturers, but the same items appear on every machine.
On the screen
Compare it with
Blood flow rate
The prescribed flow, and the arterial pressure
Arterial, venous, and transmembrane pressure
Earlier readings today and in past sessions
Fluid removal goal, rate, and volume removed
The patient’s blood pressure, symptoms, and weight gain
Time remaining
The prescribed time. Pauses for alarms shorten real treatment
Conductivity and temperature
The values set at the start
Blood pressure and pulse
The time of the reading. It may be old
What the machine does when it alarms
Alarms fall into two families. For a blood-side alarm, the machine stops the blood pump and closes the venous clamp, so the cause must be found quickly. For a dialysis fluid alarm, the machine goes into bypass. Blood keeps moving, but no cleaning takes place.1
Figure 15 · When any alarm sounds
Look at the patientAwake, comfortable, breathing easily, and no blood on the bed or floor.
Look at the accessUncover it. Check the needle sites or catheter connections.
Trace the circuitLook for kinks, clamps, clots, air, and loose connections.
Read the alarmWhich sensor, and in which direction.
Let the nurse correct the causeThe alarm is cleared only after the cause is found.
Call if it repeatsAn alarm that returns is a finding. Report it to the attending nephrologist.
Alarm
Likely causes
What you check in the patient
Arterial pressure too negative
Needle against the vessel wall; kinked line; poor fistula inflow; blocked catheter; low blood pressure
Blood pressure. The thrill (the buzz you feel) and the bruit (the whoosh you hear) over the access. Arm or neck position
Venous pressure high
Kinked line; clot in the venous chamber; needle out of position or through the vessel wall; narrowed vein beyond the needle
Swelling, pain, or tightness near the venous needle. Compare both arms
Venous pressure low
Needle dislodged; line separated; pump slowed; clot ahead of the sensor
Look for blood at once. Blood pressure, pulse, and color
Transmembrane pressure high
Fibers clotting; fast fluid removal
Dark streaks in the dialyzer. Blood pressure and cramps
Air detected
Loose connection ahead of the pump; empty infusion bag; low chamber level
Torn fibers. Air bubbles in the fluid can give a false alarm
Blood pressure, pulse, temperature. Fever or chills later
Conductivity or temperature
Empty or wrong concentrate; water supply problem; mixing or heater fault
Chest or back pain, breathlessness, confusion. The color of the blood in the venous line
When hemolysis is suspected
Hemolysis during dialysis is rare. In an outbreak traced to narrowed blood tubing, 20 of 30 affected patients had high blood pressure and 18 had abdominal pain. Ten needed intensive care, and two died.8 Chest or back pain, breathlessness, and translucent blood in the venous line, described as port-wine or cherry red, are other described signs.1
Burst red cells release potassium. When hemolysis is suspected, dialysis is stopped, the lines are clamped, and the blood in the circuit is not returned to the patient.1 Follow your unit’s protocol and call the attending nephrologist at once.
RememberAn alarm is a question about the patient. Answer it at the chair, with your eyes and hands, before anyone touches the screen.
On the floor
Real situations: problems and solutions
These are fictional but typical. For each one, cover the answer, decide what you would do, then compare. Where a step depends on local rules, your unit’s procedure and your supervisor’s direction come first.
The blood flow that kept getting lower
What you see
A patient’s arterial pressure alarm has sounded five times in an hour. Each time the blood flow was lowered a little and the alarm cleared. The session is on schedule and nobody has mentioned it to you.
What is likely going on
The access is not supplying the prescribed blood flow. The cause may be needle position, a narrowed fistula, a failing catheter, or a low blood pressure.
What to do
Check the patient’s blood pressure and examine the access. Ask the nurse what flow was prescribed and what is running now. Report the gap and the access findings to the attending nephrologist, and record them.
Why it matters
A lower blood flow delivers less dialysis than prescribed, and a repeating arterial alarm is often the first sign of an access that is failing.
Three machines at once
What you see
Conductivity alarms sound on three machines in the same row within a minute. All three have gone into bypass. The patients feel well.
What is likely going on
A shared supply is at fault: the treated water or a central concentrate line. A single machine fault would not affect three stations.
What to do
Do not allow any override. Check each connected patient for chest or back pain, breathlessness, and confusion. Ask the technician to check the water and concentrate supply. Call the head nurse, the medical director, and the attending nephrologists, and follow the unit’s protocol.
Why it matters
Bypass has protected the patients so far. An override would send fluid of unknown composition to several patients at once.
The saline bag that ran dry
What you see
During a saline flush, the bag empties while the nurse is called to another chair. The air detector alarms, the pump stops, and the venous clamp closes. Foam is visible in the venous chamber.
What is likely going on
Air was drawn in through the open saline line, which joins the circuit ahead of the pump where the pressure is below zero.
What to do
Keep the clamp closed. Check the patient for cough, chest tightness, breathlessness, and a change in alertness, and measure oxygen saturation. The staff remove the air by the unit’s procedure before any blood is returned. If the patient has any symptom, or if air passed the clamp, treat it as an emergency: follow the unit’s protocol for suspected air in the bloodstream and call the attending nephrologist at once. Otherwise, report the event to the attending nephrologist and record it.
Why it matters
The detector worked because the line was seated in it. Air that reaches the patient’s vein can be fatal.
A quiet machine and a pale patient
What you see
An elderly man is asleep in the last hour of his session. His machine shows no alarm. The cuff reading on the screen is normal, and it was taken 50 minutes ago. He looks gray.
What is likely going on
The machine monitors the circuit, not the patient. His blood pressure may have fallen since the last reading, or he may be bleeding where it cannot be seen.
What to do
Wake him, uncover the access, and look for blood. Take a blood pressure now. If it is low, follow the unit’s protocol for low blood pressure during dialysis and call the attending nephrologist.
Why it matters
No sensor detects a patient who is becoming unwell. The time stamp on a blood pressure matters as much as the number.
The pressure that crept up
What you see
A patient’s transmembrane pressure has risen steadily since the second hour. The nurse notes that he is on a heparin-free prescription because of a recent tooth extraction. The dialyzer looks darker than at the start.
What is likely going on
The fibers are clotting. With less open membrane, the machine needs more pressure to remove the same volume of water.
What to do
Check the patient and his blood pressure. Ask the nurse what the unit’s protocol provides for a clotting circuit. Call the attending nephrologist before the circuit clots completely, because the choices about anticoagulation and about ending the session early belong to the prescription.
Why it matters
A circuit that clots fully cannot be returned, and the patient loses the blood inside it. An early call keeps more options open.
Field notes
When an alarm sounds, look at the patient’s face and the access before you look at the screen.
Ask the nurse two questions about any repeating alarm: what was found, and what was changed.
Read the time printed beside the blood pressure on the screen. An old normal reading is not reassurance.
Learn where the venous chamber, the air detector, and the venous clamp are on your unit’s machines before your first shift.
If you see an access under a blanket, uncover it and explain why to the patient. Most patients agree at once.
Case review
Back to the case: The venous pressure alarm
What is a high venous pressure alarm telling you? The machine is meeting resistance when it returns blood to Mr. Bautista. The venous sensor sits after the dialyzer, so the obstruction lies between the venous chamber and his vein. It could be a kinked line, a clot in the chamber, or a needle that has passed through the wall of a new, fragile fistula.
What do you check before anyone changes a limit? The patient. Ask Liza to keep the pump stopped, lift the blanket, and look at the arm. A new fistula, an arm that feels tight, and a repeating high venous pressure together suggest infiltration, which means that blood is being pumped into the tissue around the vein. Look and feel for swelling above the venous needle, then trace the venous line. Widening the limit would have told the machine to keep pushing blood into his forearm.
Would a silent alarm prove the needle was in place? No. Most of the venous pressure comes from the needle itself. If a needle slips out, the pressure may fall too little to cross the lower limit.9 The reliable monitor is a visible access, and Mr. Bautista’s arm was under a blanket.
What happens next. The nurse manages the needle according to the unit’s protocol. You examine the arm and call the attending nephrologist. Whether the session continues through a new needle site, and how the fistula is used next time, are the nephrologist’s decisions. Record the alarm and what you found.
Skills lab · with your preceptor
Skills checklist: Circuit and alarm walk-through on a training machine
On a primed training machine with no patient connected, identify each component and alarm aloud while a senior dialysis nurse or technician observes. Your preceptor rates each step as satisfactory, needs practice, or not observed. Steps marked Critical must all be satisfactory before sign-off. This checklist supports supervised learning; your unit’s own procedure and competency form are the authority. A printable version of every checklist is on the trainer page.
CriticalTraces the blood path from the arterial connector to the venous connector, naming each component in order.
CriticalPoints to the segment of the circuit that is below zero pressure and explains why air can enter there.
CriticalIdentifies the arterial pressure sensor and states what a more negative reading suggests about the access.
CriticalIdentifies the venous chamber and venous pressure sensor and gives two causes each for a high and a low reading.
CriticalIdentifies the air detector and venous clamp, and confirms that the venous line is seated in both.
Shows the blood inlet, blood outlet, and the two dialysis fluid ports of the dialyzer, and explains countercurrent flow.
Reads the dialyzer label and states its membrane type and whether it is low-flux or high-flux.
CriticalLocates the acid and bicarbonate concentrates and checks the concentrate label against a sample prescription.
CriticalFinds on the screen: blood flow, arterial pressure, venous pressure, transmembrane pressure, fluid removal goal and rate, time remaining, conductivity, and temperature.
Explains what the machine does on a blood-side alarm and on a dialysis fluid alarm, including bypass.
Locates the blood leak detector and states the two reasons a blood leak matters.
CriticalWith the observer triggering three alarms, states for each the likely causes and what to check in the patient before the alarm is cleared.
States which alarms require a call to the attending nephrologist in this unit, and where the protocol is kept.
Placement activity. During your supervised shifts, stand with the nurse for ten alarms. For each, write down the sensor, the cause that was found, and what was checked in the patient. Review the list with your preceptor.
Self-assessment
Check your understanding
Each visit draws 8 questions at random from this module’s 15-question bank. Choose the single best answer for each, then select Check answers; an explanation appears under the best answer. Select New set of questions for a different mix. A good target is 7 of 8 on two different sets. Nothing you select is sent anywhere; scoring happens in your browser.
Go deeper
Read next on this site
Hemodialysis ModalitiesHow standard hemodialysis, high-flux hemodialysis, and hemodiafiltration compare.
Daugirdas, J. T., Blake, P. G., & Ing, T. S. (Eds.). (2015). Handbook of dialysis (5th ed.). Wolters Kluwer.
Hall, J. E., & Hall, M. E. (2021). Guyton and Hall textbook of medical physiology (14th ed.). Elsevier.
Eknoyan, G., Beck, G. J., Cheung, A. K., Daugirdas, J. T., Greene, T., Kusek, J. W., Allon, M., Bailey, J., Delmez, J. A., Depner, T. A., Dwyer, J. T., Levey, A. S., Levin, N. W., Milford, E., Ornt, D. B., Rocco, M. V., Schulman, G., Schwab, S. J., Teehan, B. P., & Toto, R., for the Hemodialysis (HEMO) Study Group. (2002). Effect of dialysis dose and membrane flux in maintenance hemodialysis. New England Journal of Medicine, 347(25), 2010–2019. https://doi.org/10.1056/NEJMoa021583
Locatelli, F., Martin-Malo, A., Hannedouche, T., Loureiro, A., Papadimitriou, M., Wizemann, V., Jacobson, S. H., Czekalski, S., Ronco, C., & Vanholder, R., for the Membrane Permeability Outcome (MPO) Study Group. (2009). Effect of membrane permeability on survival of hemodialysis patients. Journal of the American Society of Nephrology, 20(3), 645–654. https://doi.org/10.1681/ASN.2008060590
National Kidney Foundation. (2015). KDOQI clinical practice guideline for hemodialysis adequacy: 2015 update. American Journal of Kidney Diseases, 66(5), 884–930. https://doi.org/10.1053/j.ajkd.2015.07.015
Blankestijn, P. J., Vernooij, R. W. M., Hockham, C., Strippoli, G. F. M., Canaud, B., Hegbrant, J., Barth, C., Covic, A., Cromm, K., Cucui, A., Davenport, A., Rose, M., Török, M., Woodward, M., & Bots, M. L., for the CONVINCE Scientific Committee Investigators. (2023). Effect of hemodiafiltration or hemodialysis on mortality in kidney failure. New England Journal of Medicine, 389(8), 700–709. https://doi.org/10.1056/NEJMoa2304820
Vernooij, R. W. M., Hockham, C., Strippoli, G., Green, S., Hegbrant, J., Davenport, A., Barth, C., Canaud, B., Woodward, M., Blankestijn, P. J., & Bots, M. L., for the CONVINCE Scientific Committee and the HDF Pooling Project Investigators. (2024). Haemodiafiltration versus haemodialysis for kidney failure: An individual patient data meta-analysis of randomised controlled trials. The Lancet, 404(10464), 1742–1749. https://doi.org/10.1016/S0140-6736(24)01859-2
Duffy, R., Tomashek, K., Spangenberg, M., Spry, L., Dwyer, D., Safranek, T. J., Ying, C., Portesi, D., Divan, H., Kobrenski, J., Arduino, M., Tokars, J., & Jarvis, W. (2000). Multistate outbreak of hemolysis in hemodialysis patients traced to faulty blood tubing sets. Kidney International, 57(4), 1668–1674. https://doi.org/10.1046/j.1523-1755.2000.00011.x
Ribitsch, W., Schilcher, G., Hafner-Giessauf, H., Krisper, P., Horina, J. H., Rosenkranz, A. R., & Schneditz, D. (2014). Prevalence of detectable venous pressure drops expected with venous needle dislodgement. Seminars in Dialysis, 27(5), 507–511. https://doi.org/10.1111/sdi.12169
Frinak, S., Kennedy, J., Zasuwa, G., Passalacqua, K. D., & Yee, J. (2023). Detection of hemodialysis venous needle dislodgment using venous access pressure measurements: A simulation study. Kidney360, 4(4), e476–e485. https://doi.org/10.34067/KID.0000000000000093
Educational study material for physicians. Does not replace the attending nephrologist’s orders, the unit’s protocols, required certification courses, or supervised practice.