Nursing Education · Advanced module 13

Convective Therapies: CRRT and Hemodiafiltration

The physiology shared by continuous kidney replacement in the intensive care unit and hemodiafiltration in the dialysis unit, and the nursing that keeps each one safe: modes, prescription, anticoagulation, fluid balance, circuits, complications, and quality.

PublishedNailathalaGipatikPepalwal: ReferencesMga SanggunianMga TinubdanReng Reperensya: 18 Audience: Registered nurses entering nephrology or hemodialysis practice · nurse educators and preceptors Read timeOras ng pagbasaOras sa pagbasaOras ning pamamasa:
An ICU nurse checks the hanging fluid bags on a continuous kidney replacement machine at a critically ill patient's bedside.
Renal Care Academy› Nephrology Nursing & Hemodialysis›Advanced module 13 Nephrology Nursing path: advanced, after the capstone Hemodialysis Nursing path: advanced, after the capstone
🩺

Self-study for registered nurses

This module is self-study for registered nurses. It explains the reasoning behind safe practice; it does not replace the physician’s orders, your unit’s approved procedures, or the manufacturer’s instructions for the equipment you use. Every case is fictional. A quiz score records self-assessment only; practice competencies are observed and signed off by a clinical preceptor or training provider.

After this module you should be able to

A sieve as well as a tea bag

Module 6 pictured standard hemodialysis (HD) as a tea bag. Small solutes such as urea and potassium drift out of the blood and into the dialysate, because there is more of them on the blood side. That is diffusion. It works best for small molecules that move quickly.

Convective therapies add a second picture: a sieve. Pressure pushes plasma water through the membrane. Anything dissolved in that water and small enough to fit the pores is carried along with it, whatever its size up to the pore limit. Larger “middle molecules,” which diffuse slowly, leave much better this way. The water pushed out is far more than the patient can afford to lose, so sterile replacement fluid is given back into the blood line. The net effect is a washing of the plasma, not dehydration.

The picture has a limit. A kitchen sieve holds things back by size alone. A dialysis membrane also carries charge and a coating of protein on its surface. So its real behavior is described by how freely each substance crosses, its sieving coefficient, and not by a simple yes or no.

Two very different treatments use this idea. Continuous renal replacement therapy (CRRT) runs around the clock in the intensive care unit (ICU) for a critically ill patient with acute kidney injury (AKI). Hemodiafiltration (HDF) adds large-volume convection to high-flux HD in an ordinary chronic session. The machines, settings, and risks differ. The physiology and many of the safety rules are the same, which is why this module teaches them together.

Most of what goes wrong in either therapy is predictable: a wrong bag, a fluid number nobody reconciled, a clotting circuit, an alarm cleared without a cause. Your work is to understand the physiology well enough to see these coming, and to keep the record that lets the prescriber act in time.

Convection, replacement fluid, and the filtration fraction

Three ways to move things across a membrane.

ProcessDriving forceWhat it removes bestWhere it is used
DiffusionA concentration difference between blood and dialysateSmall solutes: urea, creatinine, potassiumStandard HD; continuous venovenous hemodialysis
UltrafiltrationA pressure difference across the membraneWater (with whatever is dissolved in it)Every treatment that removes fluid
ConvectionUltrafiltration large enough that solutes are dragged along with the waterSmall and middle molecules alike, up to the pore sizeHemofiltration, CRRT with replacement fluid, HDF

Why size matters. Diffusion slows sharply as molecules get larger. Beta-2 microglobulin, about 11,800 daltons, builds up in long-term dialysis patients, and diffusion alone clears it poorly. Convection works differently. A solute crosses in proportion to the volume of water moved and to its sieving coefficient. That coefficient is close to 1 for urea and potassium, which cross as freely as water. It is lower for larger molecules. It is close to 0 for albumin and blood cells, which a healthy membrane keeps in the blood.

Replacement fluid. To remove many liters of plasma water without draining the circulation, an almost equal volume of sterile, buffered electrolyte solution goes back into the blood path. In CRRT it comes from prescribed bags. In online HDF the machine makes it from ultrapure dialysate. Either way it enters the bloodstream directly. Over hours, its sodium, potassium, calcium, bicarbonate, and glucose become the patient’s plasma composition. Net fluid removal is everything removed minus everything given back.

Where the replacement fluid enters.

Pre-dilution (before the filter)Post-dilution (after the filter)
What happens to the blood in the filterIt is diluted before it reaches the fibers, so it stays thinnerIt is concentrated as water leaves, then diluted again after the filter
Clearance per liter exchangedLower, because part of each liter removed is replacement fluid that was just addedHigher: every liter removed is the patient’s own plasma water
Filter lifeLongerShorter if too much water is removed, because the blood thickens and clots
Typical useCRRT when clotting limits filter life; HDF when blood flow or hematocrit limits post-dilutionUsual mode for HDF; common in CRRT when filter life allows

The filtration fraction. This is the share of the plasma flowing through the filter that is removed as ultrafiltrate. Plasma flow is the blood flow multiplied by (1 minus the hematocrit). For example, a blood flow of 200 mL/min with a hematocrit of 30% gives a plasma flow of 140 mL/min. Removing 35 mL/min of ultrafiltrate gives a filtration fraction of 35 ÷ 140 = 25%. In post-dilution, the higher the fraction, the thicker the blood leaving the fibers and the faster the filter clots. Units set a ceiling by protocol. When it is reached, the prescriber can raise blood flow, lower the exchange, or move some replacement fluid to pre-dilution.

What each number tells you. During post-dilution, a rising transmembrane pressure or falling filter performance usually means the blood in the filter is concentrating and clotting. It does not usually mean the machine is faulty. That link between physiology and alarms runs through both halves of this module.

When a pressure climbs, ask first what is happening to the blood inside the fibers.

Diffusion, ultrafiltration, and convection across a filter fiber wall, with sieving coefficients near one for urea, lower for larger molecules, near zero for albumin.
Diffusion moves small solutes down a concentration difference; convection drags solutes of any size up to the pore limit along with water pushed through by pressure. Each solute's sieving coefficient says how freely it crosses.
Pre-dilution versus post-dilution replacement fluid compared, with the worked filtration fraction: 35 mL/min removed from 140 mL/min plasma flow equals 25%.
Replacement fluid given before the filter keeps blood thinner and the filter alive longer, at some cost in clearance; given after, every liter removed is the patient's own plasma water, but blood thickens in the fibers. The filtration fraction measures how hard the filter is working.

Where each therapy fits, and what is still unknown

Continuous therapy in the ICU. Slow removal over 24 hours causes smaller swings in blood pressure and in brain water than a 4-hour session. On that basis, the Kidney Disease: Improving Global Outcomes (KDIGO) 2012 AKI guideline suggests continuous therapy rather than standard intermittent dialysis for patients who are hemodynamically unstable (graded 2B, a weak suggestion on moderate-quality evidence). It makes the same 2B suggestion for patients with acute brain injury or raised intracranial pressure.1 Randomized comparisons have not shown that any one form of kidney support improves survival. The choice is made for tolerance, availability, and the team’s experience.1,2 Module 4 explains how and when support is started. This module covers how continuous therapy is run.

Hemodiafiltration in the dialysis unit. HDF is a chronic, outpatient therapy. It aims to clear middle molecules that standard HD removes poorly, and it is often better tolerated hemodynamically, partly because the large exchange cools the blood. Whether it improves survival depends on delivering a high convection volume. The HDF sections below set out the randomized evidence, including the largest trial.

Continuous therapy (CRRT)Hemodiafiltration (HDF)
SettingICU, around the clockDialysis unit, about 4 hours, three times a week
PatientCritically ill, usually with AKI, often in shockStable patient with kidney failure
AccessTemporary dialysis catheterFistula, graft, or tunneled catheter with good flow
Replacement fluidCommercial sterile bags hung by the nurseMade online by the machine from ultrapure dialysate
Main nursing risksFluid-balance error, anticoagulation complications, circuit clotting, hypothermia, electrolyte lossWater and dialysate contamination, clotting at high filtration fraction, not achieving the prescribed convection volume

What remains uncertain. For continuous therapy, the best anticoagulant for each patient, the right moment to stop, and how to adjust medicines that the circuit removes are still debated. For HDF, the convection volume needed for benefit, which patients gain most, and how results from European trials apply in Asian units are open questions. Where evidence is incomplete, the unit protocol and the physician’s order decide. Your accurate observations are what make those decisions possible.

Safety rules shared by every convective therapy

Replacement fluid is an intravenous medicine. Liters of it enter the bloodstream every hour. A wrong bag or a wrong additive acts like a large intravenous error. Before you hang or start it, check the prescription, the product and its concentrations (above all potassium, phosphate, calcium, and bicarbonate), the expiry date, the bag’s integrity, and any additive label. Use a second person where the unit requires it. In HDF the machine makes the fluid, so its safety rests on the water system, the ultrafilters, and the disinfection record instead (Module 8).

Three fluid numbers that must agree. The prescribed net removal, the machine’s recorded net removal, and the patient’s total balance are different numbers. The total balance counts every infusion, feed, drug, and drain. Record all three as the protocol requires. Report when they drift apart, or when the patient cannot tolerate the removal.

Common errorWhat it causesHow it is prevented
Wrong replacement or dialysate bag (for example, potassium-free instead of potassium-containing)Dangerous potassium, calcium, or acid–base changes over hoursIndependent check of every bag against the order; labeled additives
Bag or scale disturbed, bag hung but clamp closedFluid removed or given that the machine does not count correctlyFollow the machine’s bag-change steps; never lean on or touch the scales
Alarm silenced and treatment restarted without finding the causeClotting, air entry, blood loss, or unrecorded fluid shiftsLook at the patient and the circuit first, then the machine; escalate repeated alarms
Open connection or contaminated portBloodstream infection; air entryAseptic technique, locked connections kept visible, catheter care as in Module 7
Heat loss through the circuit unnoticedHypothermia that also hides feverTemperature checks and warming as ordered

Answer alarms; never bypass them. Pressure, air, blood-leak, and balance alarms protect the patient. Read which alarm it is. Look at the patient and the lines before you touch the machine. Correct the cause, and document it. An air detector or blood-leak detector is never overridden.

Document and hand over. Each shift’s record shows the prescription, the settings, the delivered treatment (effluent dose in CRRT; achieved convection volume in HDF), the fluid balance, anticoagulation monitoring, alarms and actions, circuit changes, and the patient’s response. Hand over with SBAR (Situation, Background, Assessment, Recommendation).

Every bag is a medicine, every alarm has a cause, and every hour has a balance. Check all three before you move on.

The four continuous modes, and the prolonged intermittent alternative

Continuous renal replacement therapy (CRRT) is the form of kidney replacement therapy (KRT) built for the patient with acute kidney injury (AKI) who is too unstable for a standard 4-hour session. It runs around the clock in the intensive care unit (ICU), usually through a non-tunneled dialysis central venous catheter (CVC). It removes solute and fluid slowly enough that blood pressure (BP) and brain water shift very little from hour to hour. KDIGO suggests continuous therapy rather than standard intermittent therapy for hemodynamically unstable patients and for those with brain injury or raised intracranial pressure (both graded 2B). It describes continuous and intermittent therapy as complementary, not competing.1

Every continuous machine is built from the same parts: a blood pump, a filter, an effluent pump that pulls fluid out of the filter, and, depending on the mode, a dialysate pump, one or two replacement-fluid pumps, and syringe or infusion pumps for anticoagulant. Most current machines weigh each bag on built-in scales, or measure volumes with balancing chambers, and recalculate the fluid balance every few seconds. The name of the mode simply tells you which pumps are running.

ModeWhat runsMain mechanismWhat it removes bestTypical reason it is chosen
Slow continuous ultrafiltration (SCUF)Blood pump and effluent pump only; no dialysate, no replacement fluidUltrafiltration (UF) aloneWater, with only small amounts of solute carried alongFluid overload, for example in heart failure resistant to diuretics, when potassium, acid–base, and urea are acceptable
Continuous venovenous hemofiltration (CVVH)Replacement fluid, pre-filter, post-filter, or both; no dialysateConvectionSmall and middle molecules, in proportion to the volume filteredUnit preference, or when larger molecules are a target; demands careful attention to filtration fraction
Continuous venovenous hemodialysis (CVVHD)Dialysate flowing countercurrent; no replacement fluidDiffusionSmall molecules: potassium, urea, hydrogen ionsCommon with regional citrate protocols; lower filtration fraction, so filters tend to last longer
Continuous venovenous hemodiafiltration (CVVHDF)Both dialysate and replacement fluidDiffusion plus convectionSmall molecules efficiently, plus some middle moleculesA flexible default in many ICUs; the mode used in the two large dose trials discussed below

In every mode the machine also performs net UF. The effluent pump draws slightly more out than the dialysate and replacement pumps put in. The difference is the fluid taken from the patient. The prescriber sets that difference separately from the clearance settings. So a patient can receive a high clearance with zero net removal, or a large fluid removal with modest clearance.

No mode has been shown to improve survival over another. The choice depends on what the patient most needs (fluid, potassium, acid–base, a specific toxin), the anticoagulation plan, the machine and fluids the hospital stocks, and the team’s training. Your job is to know which mode is running, and therefore which fluids, pumps, and checks apply.

Prolonged intermittent therapy. Sustained low-efficiency dialysis (SLED) is one form of prolonged intermittent renal replacement therapy (PIRRT). It uses a standard hemodialysis (HD) machine at reduced blood and dialysate flows, for roughly 6 to 12 hours, often overnight. It sits between the two extremes:

FeatureCRRTSLED or PIRRTIntermittent HD
DurationAround the clockRoughly 6 to 12 hours, per orderAbout 3 to 5 hours
MachineDedicated continuous machine with bagged fluidsStandard HD machine with treated water and online dialysateStandard HD machine
Hemodynamic gentlenessHighestIntermediate to highLowest; needs an otherwise stable patient
StaffingICU nurse, with dialysis team support per hospital modelDialysis nurse present or on call during the run, per hospital modelDialysis nurse
Time off the machine for procedures and mobilizationLittleThe rest of the dayMost of the day
Cost of consumablesHighest: dedicated sets and many liters of bagged fluidLowerLowest

The Philippine picture. Continuous machines and their bagged fluids are found mainly in tertiary hospital ICUs, and the consumables are expensive for families and hospitals alike. In many Philippine hospitals, SLED on a standard HD machine is the practical way to give an unstable patient a gentler treatment. Its water comes from the hemodialysis water system, or from a portable reverse osmosis unit that meets the water standard (Module 8).3 The principles in this module apply to both: slow removal, careful anticoagulation, hourly accounting, and early reporting. When the hospital cannot offer what a patient needs, the honest conversation and the referral belong to the physician. Your part is accurate, timely data.

Reference card comparing SCUF, CVVH, CVVHD, and CVVHDF by fluids used, mechanism, what each removes best, and why chosen, with SLED as the prolonged alternative.
The name of the mode tells you which pumps are running, and therefore which fluids, bags, and checks apply. Every mode can also remove fluid, and no mode has proven survival benefit over another.

Reading the prescription: dose, flows, and fluid removal

A complete continuous therapy order has a separate line for each pump. Read it as a set of independent instructions. A change in one line does not change the others.

Order elementWhat it controlsWhat the nurse checks
ModeWhich pumps and fluids are usedThe machine program matches the order; the correct fluids are hung
Blood flowHow fast blood passes through the filterSet value; whether the catheter actually delivers it without access alarms
Dialysate rateDiffusive clearanceCorrect bag composition and rate
Replacement fluid rate and site (pre-filter, post-filter, or split)Convective clearance and filtration fractionCorrect bag, rate, and line position
Net fluid removalHow much fluid leaves the patient each hour through the machineHourly target, upper and lower limits, and the conditions for stopping or reducing it
AnticoagulationWhether the circuit is protected from clotting, and howAgent, protocol, and laboratory monitoring schedule
Fluid compositionPotassium, phosphate, calcium, and bicarbonate in the bagsThe label of every bag against the order; additives only as ordered and labeled

Dose is effluent. In continuous therapy, small-solute clearance is roughly equal to the volume of effluent. So the dose is written as effluent volume per kilogram of body weight per hour. Effluent is everything the effluent pump removes: spent dialysate, filtered replacement fluid, and the net fluid taken from the patient.

KDIGO recommends delivering an effluent volume of 20 to 25 mL/kg per hour for continuous therapy in AKI. This is a strong recommendation on high-quality evidence (graded 1A). The guideline notes that delivering it usually needs a higher prescription.1 Two large randomized trials explain why more is not better:

A higher dose costs more fluid. It removes more phosphate, potassium, nutrients, and antibiotics. In these trials it did not improve survival. The practical problem in most ICUs is the opposite: delivering less than was prescribed.

Delivered versus prescribed dose. Every minute the blood pump is stopped, no clearance happens. Bag changes, alarms, filter clotting, circuit changes, trips to imaging, and access problems all subtract from the day. Pre-filter replacement fluid also dilutes the blood before the filter, so each liter of effluent carries slightly less solute. This is why the prescription is usually set above the target. It is also why recording downtime is a nursing task with direct consequences for dose.

A worked example (an illustrative calculation, not an order). A 70 kg patient on CVVHDF has dialysate at 1,000 mL per hour, post-filter replacement at 600 mL per hour, and net removal of 100 mL per hour.

  1. Effluent = 1,000 + 600 + 100 = 1,700 mL per hour.
  2. Prescribed dose = 1,700 ÷ 70 = about 24.3 mL/kg per hour.
  3. Over the day, the circuit clots once and the patient goes for a computed tomography scan. The machine ran for 20 of 24 hours.
  4. Delivered dose = 24.3 × 20 ÷ 24 = about 20.2 mL/kg per hour, at the bottom of the target band, before any allowance for dilution.

The site’s continuous therapy dose calculator does the same arithmetic, with a correction for pre-filter replacement. Use it to check your understanding, not to change a prescription.

Blood flow is not dose. In intermittent HD, blood flow drives clearance. In continuous therapy, effluent does. Blood flow matters here for filter life and filtration fraction. If blood flow falls, for example because of repeated access alarms, while the replacement rate stays the same, a larger share of the plasma is filtered. The blood in the fibers thickens, and the filter clots sooner.

Net removal is a rate, not just a daily total. “Two liters negative by tomorrow” is safe only if the hourly rate is one the circulation can refill from the tissues. The prescriber sets the rate and the limits. You watch the patient’s response hour by hour, as the fluid section below describes.

Dose is what reached the patient, not what was typed into the machine. Record every stop.

Keeping the circuit open: no anticoagulant, heparin, or regional citrate

Blood in a plastic circuit clots. A clotted filter stops treatment. If its blood cannot be returned, the patient loses it. A new set costs money, and the delivered dose falls. There are three broad strategies, and the prescriber chooses among them for each patient.

StrategyHow it worksMain riskNursing monitoring, per protocol
No anticoagulantRelies on the patient’s own clotting problem, pre-filter dilution, adequate blood flow, and fewer interruptionsShort filter life and lost bloodPressure trends; filter appearance; prompt response to alarms
Unfractionated heparinInfused into the circuit before the filter; it also enters the patientBleeding in the patient; heparin-induced thrombocytopeniaClotting tests and platelet count as ordered; bleeding from catheter sites, wounds, gut, and airway
Regional citrateInfused before the filter; binds calcium so the blood in the circuit cannot clot; calcium is replaced into the patientLow ionized calcium, citrate accumulation, alkalosis, errors with the calcium infusionPatient ionized calcium and post-filter ionized calcium; total calcium; blood gases and sodium

KDIGO suggests regional citrate rather than heparin for continuous therapy in patients who have no contraindication to citrate (graded 2B). For patients who cannot receive citrate, it suggests unfractionated or low-molecular-weight heparin (graded 2C).1 The RICH trial compared the two directly. Among 596 critically ill patients analyzed, median filter life was 47 hours with citrate and 26 hours with heparin. Bleeding complications occurred in 5.1% with citrate and 16.9% with heparin, about 12 fewer bleeds per 100 patients. New infections were more frequent with citrate, 68.0% against 55.4%, about 13 more per 100 patients. The trial stopped early, so it could not show whether either strategy changes survival.6

How citrate works, step by step.

  1. Citrate solution is infused into the blood line just after the blood leaves the patient. Citrate binds ionized calcium. The clotting cascade cannot run without ionized calcium, so the blood in the circuit does not clot.
  2. In the filter, much of the citrate–calcium complex passes into the effluent and is discarded.
  3. The calcium lost this way is replaced by a separate calcium infusion into the patient, through the return line or a separate central line, according to the machine and protocol.
  4. The citrate that reaches the patient is metabolized, mainly by the liver and also by muscle. Each molecule yields bicarbonate. This is why citrate protocols use fluids with less buffer than heparin protocols.

Two calcium numbers, two different questions. The post-filter ionized calcium, sampled from the circuit after the filter, answers “Is the circuit anticoagulated enough?” It guides the citrate rate. The patient ionized calcium, sampled from an arterial line or the patient’s own vein, answers “Is the patient safe?” It guides the calcium infusion. Mixing up the sample sites leads to the wrong adjustment, so label every sample with its site and time. Target ranges and the response to each result come from the unit’s protocol.

Low ionized calcium in the patient shows as tingling around the mouth or in the fingers, cramps or spasm, a falling BP, arrhythmia, or a prolonged QT interval on the electrocardiogram. Check first that the calcium infusion is running, connected, and not empty. A calcium bag that ran out unnoticed is a common cause. Report at once.

Citrate accumulation happens when citrate enters the patient faster than the body can metabolize it. The patients at risk have severe liver failure, or severe shock with poor tissue perfusion, often signaled by a high or rising lactate. Accumulated citrate keeps binding calcium, so the pattern is:

A rising ratio is reported, not managed alone. The threshold for action, and the response (lowering blood flow and citrate, raising clearance, or changing strategy), are set by the protocol and the prescriber.

Citrate excess with normal metabolism is the opposite problem. The body metabolizes all the citrate it receives, and the bicarbonate produced builds up. The patient develops metabolic alkalosis, sometimes with high sodium, since many citrate solutions carry sodium. Report a rising bicarbonate or pH. The prescriber adjusts the fluids or rates.

Heparin and the platelet count. Heparin-induced thrombocytopenia is possible when the platelet count falls substantially after several days of heparin, when new clots form in the circuit or the patient despite heparin, or when skin lesions appear at injection sites. Report it promptly. In a patient with this condition, KDIGO states that all heparin must be stopped, including heparin catheter locks and flushes (graded 1A), and a non-heparin anticoagulant is chosen by the prescriber.1 Other causes of a falling platelet count in the ICU, including sepsis and the circuit itself, appear under complications.

Safety rules for every strategy. Anticoagulant and calcium infusions are high-alert medicines. Double-check the bag, concentration, rate, and line at every change. Never run the calcium infusion through the same line as the citrate. Never restart a stopped citrate or calcium infusion without knowing why it stopped. No doses appear in this module. Every rate comes from the prescriber’s order and the unit protocol.

On citrate, the calcium bag is as important as the citrate bag. Never let it run dry.

CVVHDF circuit from a jugular dialysis catheter: citrate enters the red access line, dialysate runs countercurrent, replacement fluid joins one blood line, calcium returns separately.
Citrate enters the blood line just after it leaves the patient and binds calcium so the circuit cannot clot; much of it leaves in the effluent, and calcium is replaced into the patient by a separate infusion. Dialysate runs outside the fibers against the blood; replacement fluid joins the blood line itself.
Citrate anticoagulation algorithm: how citrate works, two calcium samples, and the signs of low calcium, accumulation, and excess, each ending in report per protocol.
Two calcium samples answer two questions: is the circuit anticoagulated, and is the patient safe. Low calcium, citrate accumulation, and citrate excess each have a recognizable pattern; the nurse checks the infusions and reports, and targets and adjustments come from the protocol.

Hourly fluid balance: three numbers that must agree

Continuous therapy is often started largely to remove fluid. In one large multicenter observational study of critically ill adults with AKI, fluid overload, defined as a gain of more than 10% of body weight, was associated with higher mortality. Among patients who received dialysis, survivors had less fluid accumulated when treatment started.7 The design cannot show that removing fluid saves lives. Sicker patients both gain more fluid and die more often. Still, it is the reason fluid balance is watched so closely.

The three numbers. Confusing them is the most common fluid error in continuous therapy.

NumberWhat it isWhere it comes from
Machine net removalThe fluid the machine took out of the patient, after subtracting what it put back as dialysate and replacement fluidThe machine’s balance display for the hour
Patient total balanceEverything in minus everything out: all infusions, flushes, medicines, nutrition, and blood products, minus urine, drains, gastric losses, stool, and machine net removalThe nurse’s hourly intake and output record
Prescribed goalWhat the prescriber wants the patient’s balance to be per hour or per day, with limitsThe order

The machine knows nothing about the antibiotic infusion, the feed, or the urine bag. A machine that removed exactly what was set can still leave the patient positive if other intake is high.

A worked hour (illustrative). The order asks for a patient total balance of 100 mL negative per hour.

The machine did what it was set to do. The patient is still 50 mL short of the goal this hour. Some units allow nurses to adjust machine removal within written limits to meet a total-balance goal. Others require a call each time. Follow the order and protocol, and document which number you adjusted and why.

Citrate and calcium infusions add volume. Depending on the machine, they may or may not be included in the machine’s balance. Find out how your machine handles them, so that they are counted once: not twice, and not never.

When to call.

Cumulative balance. Total the hourly patient balances across the shift and the day. Carry the running total forward from the start of therapy, and ideally from ICU admission. The site’s continuous therapy flowsheet and fluid-removal calculators follow the same logic: hourly entries, a running total, and comparison with the goal. A daily weight on the same bed scale, with the same equipment on the bed, is a useful cross-check. It does not replace the hourly record.

Handling bags on machines that weigh them. Built-in scales assume that nothing touches the bags except the machine. Common errors:

Each of these creates an error in fluid the machine believes it removed or returned. Unnoticed, such errors can reach liters over a day. That is why most machines stop treatment when cumulative balance errors pass a set limit. Repeated balance alarms are reported and the scales checked, not cleared and restarted.

Document every hour: machine net removal, all intake, all output, patient total balance, cumulative balance, the prescribed goal, vital signs, circuit pressures, any downtime with its cause and duration, and any adjustment made with its authority.

The machine balances only its own bags. You balance the patient.

Machine net removal, patient total balance, and prescribed goal compared, with a worked hour where the machine removes 250 mL yet the patient is 50 mL short.
The machine counts only its own fluids; it knows nothing about the infusions, the feed, or the urine bag. Record all three numbers every hour and report when they drift apart or the patient cannot tolerate removal.

The circuit: priming, pressures, filter life, and alarms

Priming and connection. Load and prime the set exactly as the manufacturer directs, with the prime solution your unit orders. Check that the filter and all lines are free of air and that every clamp is in the right position. Connection to the catheter is an aseptic procedure (Module 7): mask, hand hygiene, clean gloves or sterile technique per policy, scrubbed hubs, and both lumens aspirated and flushed before connection. At the start, blood flow is usually raised in steps while you watch the patient’s BP and the access and return pressures.

The pressure points. Machines name and display them differently. The physiology is the same.

PressureWhere it is measuredWhat a change usually means
Access (arterial) pressureBetween the catheter and the blood pump; normally negativeMore negative: the pump is pulling against a resistance in the catheter or vein
Pre-filter pressureAfter the blood pump, before the filter; positiveRising: clot forming in the filter, or a blockage further downstream
Return (venous) pressureAfter the filter, before blood re-enters the patient; positiveHigher: blockage in the return line, catheter, or vein, or clot in the air trap. Lower or sudden fall: possible disconnection
Effluent pressureOn the fluid side of the filterBecoming more negative: the membrane is clogging and the pump must pull harder
Transmembrane pressureCalculated across the membrane from the other pressuresA steady rise: the membrane pores are clogging, which lowers convective clearance and warns of filter failure
Pressure dropPre-filter pressure minus the pressure at the filter outletA rise: the blood channels inside the filter are clotting

Trends matter more than single numbers. Record pressures hourly. A transmembrane pressure or pressure drop that climbs over several hours predicts clotting long before the circuit stops. That gives you time to report and to plan a controlled change with blood return, instead of losing the circuit.

Troubleshooting. Look at the patient before the machine, every time.

Alarm or trendLikely causeFirst nursing checks
Access pressure too negative, often with coughing or position changeCatheter tip against the vessel wall, kink, low circulating volume, clot in the lumenPosition of head and hip (femoral catheters kink when the patient sits up); external kinks and clamps; dressing tension; a gentle flush per protocol; report persistent alarms. Reversing the lines is done only on a physician order or protocol, and increases recirculation
Return pressure highKinked or clamped return line, clot in the air trap or return lumen, patient positionLine path and clamps; air trap for clot; patient position; report if not resolved
Return pressure suddenly lowDisconnection or line separationLook at every connection at once; stop the blood pump if blood is leaking; this is an emergency
Pre-filter pressure, transmembrane pressure, or pressure drop rising steadilyFilter clotting; post-dilution with high filtration fraction; low blood flow; inadequate anticoagulationAnticoagulant and calcium infusions running and connected; recent interruptions; filter and air trap for dark streaks; report the trend and prepare for a planned change
Air detector alarmAir in the return line, often from a loose connection, an empty bag, or a micro-leakDo not override; inspect the line and air trap; remove air only by the manufacturer’s procedure; check every connection and bag
Blood leak detector alarmRed cells in the effluent from a ruptured fiber, or a dirty or misread sensorLook at the effluent; follow the manufacturer’s procedure; a confirmed leak means the filter is changed
Balance or scale alarmBag touching, wrong bag, clamp closed, empty or full bagBags hanging freely; clamps; bag levels; use the bag-change function

What shortens filter life. A high filtration fraction, especially with post-dilution replacement. Low or interrupted blood flow. Frequent access alarms. A pump left stopped with blood in the circuit. Inadequate or interrupted anticoagulation. Air in the circuit. Most of these are within your reach: a well-positioned catheter, quick answers to alarms, few unnecessary stops, and anticoagulant and calcium bags changed before they run out.

Planned change versus clotted circuit. A circuit is changed when it reaches the manufacturer’s maximum running time, when pressures show clotting is coming, or when a pause will last long enough for the circuit to clot. In a planned change, the blood in the circuit is returned to the patient by the machine’s procedure. That saves a meaningful volume of blood in an anemic ICU patient. When the circuit has already clotted, or the return pressure is very high, returning the blood can push clot toward the patient. Most protocols forbid returning blood from a clotted circuit. Follow yours, document the estimated blood lost, and report it.

Machine-specific steps for priming, bag changes, blood return, and alarm handling come from the manufacturer’s instructions and the unit’s competency program. This module explains the reasons behind them. It does not replace them.

A rising trend is a warning you can act on. A clotted circuit is a loss you can only record.

Continuous therapy alarm algorithm: access, return, pre-filter and transmembrane trends, air and blood leak alarms, each mapped to likely cause, first check, and escalation.
Look at the patient before the machine. Rising transmembrane pressure or pressure drop predicts clotting hours ahead, which gives time to report and plan a change with blood return; air and blood-leak detectors are never overridden.

Complications, and what continuous clearance takes away

Continuous clearance does not choose what it removes. Over days it takes out heat, electrolytes, nutrients, and medicines as steadily as it removes urea and potassium.

Hypothermia. Blood outside the body loses heat, and room-temperature fluids at liters per hour cool it further. Machines have blood or fluid warmers, but patients still cool, particularly at high effluent rates. Cooling hides fever. A patient on continuous therapy may have a normal temperature despite sepsis. Look for other signs of infection: a rising white cell count, new hypotension, a rising lactate, or a warmer setting that keeps having to go up. Report them.

Electrolytes. Low phosphate is common. In the RENAL trial it was more frequent with the higher intensity.4 Low potassium and low magnesium are also common after the first day, especially with potassium-free or low-potassium fluids. Replacement is given as ordered, intravenously or by changing to fluids that contain the electrolyte. Your part is to keep the laboratory schedule and make sure each result is acted on.

Acid–base. Bicarbonate-buffered fluids correct metabolic acidosis over hours to days. KDIGO suggests bicarbonate rather than lactate as the buffer in AKI (graded 2C), and recommends it in circulatory shock (graded 1B).1 Overcorrection can cause metabolic alkalosis, particularly with citrate. Follow blood gases as ordered.

Bleeding. Sources include the catheter site, recent surgical wounds, the gut, and the airway. Heparin raises the risk. Check sites at each assessment. Report new bleeding, a falling hemoglobin, or blood in drains, stool, or the endotracheal tube.

Catheter problems. ICU dialysis catheters are a source of bloodstream infection, and of dysfunction from clot, kinking, or malposition. Reserve the catheter for kidney replacement therapy, handle it aseptically at every connection, and inspect it each shift (Module 7). Report fever, redness, discharge, or pain at the site, and access alarms that keep coming back after repositioning.

Drug removal. Continuous therapy removes many medicines, including many antibiotics. Their dosing differs from both normal kidney function and intermittent HD. Under-dosing an antibiotic in sepsis is as dangerous as overdosing. When a new medicine is ordered, or the effluent rate changes substantially, ask the pharmacist or prescriber whether the dose has been adjusted for continuous therapy. Note the time of any long interruption, because it changes how much drug is removed.

Nutrition. Amino acids, glucose (with some fluids), water-soluble vitamins, and trace elements pass into the effluent. Critically ill patients on continuous therapy have high protein needs. The dietitian and prescriber set the plan. Record feeding interruptions for procedures, so that intake can be compared with the plan.

Low platelets. In an ICU patient on continuous therapy, a falling platelet count may come from sepsis, medicines, consumption in the circuit, or heparin-induced thrombocytopenia. The prescriber determines the cause. Your role is to notice the trend and report it.

Watch forHow you notice itAction
HypothermiaCore temperature trend; cool skin; shiveringCheck the warmer; warming measures per protocol; report; remember that fever may be hidden
Low phosphate, potassium, or magnesiumScheduled laboratory results; weakness; new arrhythmia on the monitorReport; replace as ordered; confirm the bag composition
Low ionized calciumTingling, spasm, hypotension, prolonged QT intervalCheck the calcium infusion; report at once
Rising total-to-ionized calcium ratio or rising calcium requirementLaboratory trend on citrateReport; follow the citrate accumulation protocol
Metabolic alkalosisRising bicarbonate and pH on blood gasesReport for adjustment of fluids or rates
BleedingSite checks; falling hemoglobin; blood in drains, stool, or airwayPressure where appropriate; report; review anticoagulation with the prescriber
Catheter infectionFever (which may be masked), redness, discharge, rising white cell countReport; cultures as ordered; aseptic handling
Inadequate drug levelsNew antibiotic order; change in effluent rateAsk the pharmacist or prescriber about continuous therapy dosing
Falling platelet countDaily blood count trendReport; ask whether heparin-induced thrombocytopenia should be considered
Hemodynamic instabilityFalling BP or rising vasopressor needs during removalReduce or pause removal within protocol; report

Whatever the circuit removes from the blood, it removes without choosing. Ask each shift what it took that the patient needed.

Watch table for continuous therapy: hypothermia, electrolyte losses, acid–base, bleeding, catheter problems, drug removal, nutrition, low platelets, each with signs and action.
Continuous clearance does not choose what it removes: heat, electrolytes, nutrients, and medicines leave as steadily as urea and potassium. Notice the trend and report it; replacement and dosing come from the prescriber, pharmacist, and dietitian.

Interruptions, transport, recovery, step-down, and handoff

Planned interruptions. Scans, operations, and procedures often need the patient off the machine. Agree the plan with the team before the time comes. The usual options are:

Either way, stop or adjust the anticoagulant and calcium infusions exactly as the protocol directs when the circuit stops. A calcium infusion that keeps running after citrate has stopped will raise the patient’s calcium. A citrate infusion running without blood flow is a hazard. Record the time off and the reason, because it affects delivered dose and drug dosing.

Transport while still connected is unusual, and done only where the machine and protocol allow it. More often the circuit is disconnected first. Either way, secure and cap the catheter aseptically, and make sure the team carries the plan for reconnection.

Recognizing recovery. A rising urine output without diuretics is often the first sign that the kidneys are recovering. Report it, with the trend in creatinine and potassium. The prescriber may order a trial off therapy, with urine output, fluid balance, potassium, and acid–base watched closely. Your hourly record is the evidence on which the decision to stay off or restart is made.

Step-down to intermittent therapy. Once the patient no longer needs vasopressors and can tolerate fluid removal over a few hours, the team may move from continuous therapy to SLED or intermittent HD, often to allow mobilization and to reduce cost. The ICU nurse’s handover to the dialysis nurse includes the catheter’s function, the cumulative fluid balance, the last laboratory values, the anticoagulation history, and any concern about heparin-induced thrombocytopenia.

Stopping. Therapy stops on an order: when the kidneys have recovered enough, when the patient moves to intermittent therapy, or when goals of care change. The ending is planned. Return the blood per protocol, stop the infusions in the correct order, lock or remove the catheter as ordered, and record the final balances.

Handoff. A structured handoff in the Situation–Background–Assessment–Recommendation format is associated with fewer omissions.8 For a patient on continuous therapy it includes:

When continuous therapy is a trial. For some patients, continuous therapy starts as a time-limited trial with agreed goals and a date to review them, as Modules 4 and 5 describe. Your record of how the patient is responding is part of that review. Bring the family’s questions to the team rather than answering them alone.

Every stop, start, and handover is a moment when a number can be lost. Write it down before you move on.

Online hemodiafiltration: how the machine makes fluid safe to infuse

Earlier in this module you built the picture of convection. Water pushed across the membrane drags dissolved solutes with it, and the more water moves, the more of the larger solutes leave. Hemodiafiltration (HDF) brings that picture into the ordinary outpatient session. It keeps everything hemodialysis (HD) already does by diffusion, and adds a large convective volume on top. Far more fluid is filtered out of the blood than the patient could safely lose, so most of it is given back as substitution fluid during the same session.

Where the substitution fluid comes from. A session needs many liters, far more than anyone could hang as bags. In online HDF the machine makes the fluid itself, continuously, from the dialysate it is already preparing. The chain runs like this:

  1. Treated water from the unit’s reverse osmosis (RO) loop is mixed with acid and bicarbonate concentrates inside the machine, exactly as for HD (Module 8).
  2. The fresh dialysate passes through a first ultrafilter, a filter fine enough to hold back bacteria and endotoxin. The result is ultrapure dialysate, used on the dialysate side of the dialyzer.
  3. Part of that ultrapure dialysate is drawn off through a second ultrafilter (on some machines, a further stage as well). This cascade makes the substitution fluid, which the substitution pump infuses directly into the blood line.

Why it can be safe to infuse. Each ultrafilter stage removes most of what reaches it, so filters in series cut bacteria and endotoxin far below what any single stage achieves. The international standard for dialysis fluid quality is Part 5 of the International Organization for Standardization (ISO) 23500 series, ISO 23500-5:2024. It requires online-prepared substitution fluid to be sterile and non-pyrogenic, and the online process to be validated to produce it. Sterility at that level cannot be proved by culturing a sample at the bedside. It depends on a validated machine and filter system used exactly as the manufacturer intends, on clean water and dialysate upstream, and on disciplined disinfection. That is why every requirement below exists, and why substitution fluid is handled as an intravenous medicine, as the shared safety rules earlier in this module state.

The membrane. HDF uses a high-flux dialyzer. Only a membrane with high water permeability can pass 20 or more liters of plasma water in a session at a tolerable transmembrane pressure. The dialyzer model is part of the order. It is not interchangeable with a low-flux dialyzer.

FeatureHigh-flux HDExpanded HD (medium cut-off membrane)Online HDF
Main clearance mechanismDiffusion, with a small amount of internal convectionDiffusion plus more internal convection within the dialyzerDiffusion plus deliberate, high-volume external convection
MembraneHigh-fluxMedium cut-off, with larger effective poresHigh-flux
Substitution fluid infusedNoneNoneMany liters per session, made online
MachineStandard HD machineStandard HD machineHDF-capable machine with an online fluid module
Middle-molecule removalSomeMore than high-flux HDMore than high-flux HD, rising with convection volume
Main concern to watchDialysate quality (Module 8)Albumin loss into the dialysateSterile substitution fluid, hemoconcentration, achieving the volume

Do not confuse expanded HD with HDF. Expanded HD aims at a similar goal, more removal of larger solutes, without infusing any fluid. The survival trials described later tested online HDF. Their results cannot simply be transferred to medium cut-off membranes, which have not yet been shown in a large randomized trial to change survival.

What convection adds. Diffusion already clears small solutes such as urea and potassium well, so HDF changes them little. The gain is in “middle molecules,” which diffuse too slowly to be removed well. The best-known marker is beta-2 microglobulin, a protein of about 11,800 daltons. Its long-term buildup is linked to dialysis-related amyloidosis, with carpal tunnel syndrome and joint pain after many years on dialysis. HDF lowers pre-dialysis beta-2 microglobulin more than HD does. That is a laboratory marker, a surrogate. A lower level is not by itself proof that the patient will feel or do better. Phosphate is also removed somewhat more. But phosphate leaves the tissues slowly, so treatment time and diet remain the main tools for it. HDF does not replace phosphate binders or dietary teaching.

Albumin. Albumin is larger than the pores of a high-flux membrane and is largely kept in the blood. Small losses do occur in HDF, and they rise with very high transmembrane pressures and with some membranes. In most patients the loss is minor. It matters more in a patient who is already malnourished. Review serum albumin trends with the physician and dietitian like any other nutritional marker. Report a falling albumin; do not attribute it to HDF yourself.

Substitution fluid goes straight into the blood. Its safety is built upstream, in the water, the filters, and the disinfection log.

Online HDF fluid path: treated water and concentrates become dialysate, pass two ultrafilters, and enter the venous blood line after a high-flux dialyzer.
In online HDF the machine filters its own dialysate twice: once to make ultrapure dialysate for the dialyzer, and again to make sterile substitution fluid that runs straight into the blood. That is why the substitution line is handled like an intravenous line.

The HDF prescription: convection volume and what drives it

An HDF order contains everything an HD order contains (Module 6), plus the settings for convection. The single number that best sums up the convective dose is the convection volume: the total volume of plasma water filtered across the membrane during the session.

Convection volume = substitution volume + net ultrafiltration (UF) volume.

Net UF is the fluid removed to bring the patient toward target weight. It is filtered out and not replaced, so it counts as convection too. Substitution volume is the fluid filtered out and then given back. Most machines display the substitution and UF volumes, and many also display the convection volume itself.

The high-volume target. The large recent trial, CONVINCE, enrolled only patients judged able to receive a convection volume of at least 23 liters per session. The mean volume delivered in its HDF group was 25.3 liters.9 Earlier pooled analyses of trial data had found the lowest mortality among patients receiving the highest volumes, above 23 liters per 1.73 m² of body surface area.10 Many units therefore use 23 liters per session as a working target. Each patient’s target comes from the physician’s order and unit protocol, not from you.

Post-dilution versus pre-dilution in practice. In post-dilution, substitution fluid enters the venous line after the dialyzer. Every liter filtered carries plasma at full solute concentration, so each liter does the most work. That is why the major trials used post-dilution. The limit sits inside the fiber: as water is pulled out, the blood thickens toward the dialyzer exit. In pre-dilution, fluid enters before the dialyzer. The blood passing the membrane is thinner, so very high filtration is possible. But each liter filtered is partly the substitution fluid itself, so a much larger volume is needed for the same clearance. Mixed (or mid-) dilution modes split the fluid between the two points. Pre-dilution or mixed modes are used, by order, when hemoconcentration keeps limiting post-dilution: a high hematocrit, high plasma protein, or a circuit that keeps clotting. Their volume targets differ from post-dilution targets. Never compare the two kinds of numbers directly.

The determinants. In post-dilution, the safe convective flow is limited to a fraction of the blood flow. HDF machines and studies usually express this as convective flow divided by blood flow, a simpler bedside cousin of the plasma-based filtration fraction defined earlier in this module. The two rise and fall together. Holding that fraction steady, convection volume is set by:

Worked illustration, arithmetic for teaching only; not a prescription. A four-hour session lasts 240 minutes. To reach 23 liters (23,000 mL), convective flow must average 23,000 ÷ 240 = about 96 mL/min. With a blood flow of 350 mL/min, that is 96 ÷ 350 = about 27% of the blood flow. Keep the same 27%, but let the access deliver only 300 mL/min. Convective flow falls to about 81 mL/min, and over 240 minutes that is about 19.4 liters, below target, though nothing else changed. If that 300 mL/min session also ends 20 minutes early, the volume falls to about 17.8 liters. For the same patient, a session record showing 21.0 liters of substitution and 2.5 liters of net UF gives a convection volume of 21.0 + 2.5 = 23.5 liters.

The illustration shows that blood flow and time, not the substitution setting alone, decide whether the target is reached. It also shows a trap. Net UF counts toward the convection volume, but it is set by the patient’s fluid status. It is never raised to make the convection number look better.

Auto-substitution features. Some HDF machines can set the substitution rate automatically. The machine reads pressures across the dialyzer continuously, and on some models other measurements, and adjusts the rate to keep filtration near the highest level the circuit tolerates without excessive hemoconcentration. These features usually raise achieved volumes and reduce manual adjustment. They work within limits the unit sets, and they cannot create blood flow that the access does not deliver. How the feature works, what it measures, and when it is used follow the manufacturer’s instructions and unit protocol.

What you verify against the order before starting

ItemWhat to check
Modality and modeHDF selected on the machine; post-, pre-, or mixed dilution as ordered
DialyzerThe high-flux model ordered; correct patient label if dialyzers are reprocessed and unit policy permits reprocessing for HDF
Blood flow and timeAs ordered; needle size matched to the blood flow by the unit chart (Module 7)
Substitution settingThe ordered rate or volume, or auto-substitution if ordered
Convection volume targetRecorded on the treatment sheet so that the achieved volume can be compared at the end
Net UF goalCalculated from today’s pre-dialysis weight and the current target weight, as for HD
Dialysate and anticoagulationAs for HD: concentrates, temperature, and anticoagulant all match the current order

Blood flow and minutes make the convection volume. Protect both from the first minute of the session.

Convection volume equals substitution plus net ultrafiltration; worked example shows how lower blood flow or a shorter session drops 23 liters to 19.4 or 17.8.
Convection volume is everything filtered across the membrane: the substitution volume plus the net ultrafiltration. With the same fraction of blood flow, a weaker access or a shorter session misses the target even though the substitution setting never changed.

What safe HDF requires: patient, machine, water, and people

HDF is not simply a machine setting. Four things must be right at the same time: a patient whose access can deliver the blood flow, a machine built for it, water and dialysate of a higher microbiological quality than many units need for HD, and staff trained in its specific steps.

The patient and the access. High convection volumes need a high, steady blood flow. A well-functioning arteriovenous fistula or graft usually provides it. A tunneled catheter may or may not, depending on its position and function. Convection volume often falls on catheter sessions, and a catheter that cannot sustain the flow limits the volume whatever else is set. HDF may suit some patients less, or its benefit may be less certain: those whose access cannot provide the flow, those with recurrent circuit clotting at high filtration fractions, and those whose goals of care favor the simplest effective treatment. Selection is the physician’s decision. You supply the evidence: achievable blood flow, arterial and venous pressure trends, how often the target was reached, and how sessions are tolerated.

The machine. Online HDF needs a machine designed and validated for it, with an online substitution module, its own ultrafilters, a substitution pump, and the software and alarms that control filtration. A standard HD machine cannot be converted at the bedside. The machine also runs its own pre-treatment tests. On many models these include an ultrafilter integrity test. A failed test means the machine is not used for HDF until the fault is fixed.

Water and dialysate quality. Dialysis water must meet the water standard described in Module 8.3 For HDF, the dialysate itself must also be ultrapure, and the substitution fluid sterile. The dialysis fluid standard, ISO 23500-5:2024, sets these limits:11

FluidTotal viable microbial countEndotoxin
Standard dialysis fluidLess than 100 colony-forming units (CFU) per mLLess than 0.5 endotoxin units per mL
Ultrapure dialysis fluidLess than 0.1 CFU per mLLess than 0.03 endotoxin units per mL
Online-prepared substitution fluidSterile, produced by a process validated to achieve sterility; no culture limit is set, because sterility cannot be shown by routine samplingNon-pyrogenic

Your facility’s adopted standard, including the edition in force, governs. The values are shown so that you can see the size of the step from HD to HDF. Ultrapure dialysate has a bacterial limit one thousand times stricter than standard dialysate. Reaching it every time depends on the whole chain: a well-maintained RO loop without stagnant branches, clean concentrate handling (bicarbonate in particular supports bacterial growth), machine fluid paths disinfected on schedule, and intact ultrafilters.

Staff training and documentation. Every nurse who sets up, runs, or ends HDF needs documented training on the specific machine model, observed practice, and periodic competency review. Training covers the substitution line connection, online priming and rinse-back where the machine uses them, the HDF-specific alarms, and the logs. Dialysis technicians who maintain the machines and water system need parallel training on ultrafilter changes, integrity tests, and disinfection.

The Philippine context. Online HDF is available in a limited but growing number of Philippine units, mostly in larger urban centers. It needs HDF-capable machines, a water system that reliably supports ultrapure dialysate, and recurring costs for ultrafilters, testing, and training. Where it is offered, explain the cost to the patient, and what any insurer or government program will pay, before the patient chooses. Use the unit’s current written information, not figures remembered from elsewhere. Patients who cannot reach high volumes, or cannot afford HDF, continue to receive good care with well-delivered high-flux HD.

If any link in the chain is in doubt (water, filter, disinfection, or training), the machine does not run HDF today.

What the trials show, and what they do not

For two decades, observational studies suggested that patients on HDF lived longer than patients on HD. Observational studies cannot separate the effect of a treatment from the effect of who receives it. Healthier patients with better access are more likely to be placed on HDF and to reach high volumes. Randomized trials were needed, and at first their results seemed to disagree.

TrialPatients and comparisonPrimary outcome and resultNote on convection volume
CONTRAST (2012)714 patients; post-dilution online HDF versus low-flux HDAll-cause death: no significant difference (hazard ratio 0.95; 95% confidence interval 0.75 to 1.20)In an on-treatment analysis, patients who received higher volumes had lower mortality; the authors stated that this subgroup finding required confirmation
Turkish online HDF study (2013)782 patients; post-dilution online HDF versus high-flux HDComposite of death and nonfatal cardiovascular events: no significant difference (event-free survival 77.6% versus 74.8%)Mean filtration volume about 17.2 liters; in a post hoc analysis, patients above 17.4 liters had better survival
ESHOL (2013)906 patients; high-efficiency post-dilution online HDF versus continued HDAll-cause death 30% lower with HDF (hazard ratio 0.70; 95% confidence interval 0.53 to 0.92)Designed around high-efficiency HDF; sessions complicated by low blood pressure and all-cause hospitalization were also less frequent with HDF
CONVINCE (2023)1,360 patients on high-flux HD for at least 3 months, all judged able to receive at least 23 liters; high-volume HDF versus continued high-flux HDAll-cause death 17.3% (118 of 683) with HDF versus 21.9% (148 of 677) with HD over a median of 30 months (hazard ratio 0.77; 95% confidence interval 0.65 to 0.93)Mean delivered convection volume 25.3 liters per session

Sources for the table: CONTRAST,12 the Turkish study,13 ESHOL,14 and CONVINCE.9

CONVINCE in absolute terms. The 23% relative reduction sounds large. In absolute terms, 4.6 fewer patients per 100 died over a median of 30 months. That means about 22 patients treated with high-volume HDF for that period to prevent one death. This is a real benefit on the outcome that matters most, in patients selected for their ability to reach high volumes.9

Why the earlier trials differed. The most consistent explanation is the convective dose actually delivered. The trials that found no overall benefit delivered lower volumes on average. In both, the patients who happened to receive higher volumes did better. Those were post hoc and on-treatment analyses, so they show an association, not proof. Patients who reach high volumes usually have better access and better health, and those advantages could explain part of their survival. CONVINCE addressed this directly. It enrolled only patients able to receive high volumes, randomized them, and found lower mortality with HDF.9 The comparison group matters too. CONTRAST compared HDF with low-flux HD. The Turkish study and CONVINCE compared it with high-flux HD, the usual comparator today.

Pooled analyses. An individual-participant analysis of four earlier trials, with 2,793 participants, estimated a 14% lower risk of all-cause death and a 23% lower risk of cardiovascular death with HDF. The largest benefit was among those receiving more than 23 liters per 1.73 m² of body surface area per session. The analysis reported these as relative effects.10 A later individual-patient meta-analysis added CONVINCE: five trials and 4,153 patients. All-cause death occurred in 23.3% on HDF and 27.0% on HD over a median of 30 months (hazard ratio 0.84; 95% confidence interval 0.74 to 0.95), about 3.7 fewer deaths per 100 patients. The effect did not clearly differ across patient subgroups, and risk fell in a graded pattern as convection volume rose.15 That dose–response pattern still comes from volumes patients achieved, not volumes assigned at random. It supports, but does not prove, the idea that the volume itself is what matters.

Quality of life. In CONVINCE, both groups reported a gradual decline in health-related quality of life over time. The decline was slower with HDF in physical function, cognitive function, pain interference, and social participation. The authors described the effect as moderate, most pronounced for cognitive function.16 These are outcomes patients report and feel, and patients often value them as much as survival. Still, do not present this as “HDF makes you feel better.” A patient switched to HDF may notice no difference from one session to the next.

What remains uncertain.

How to talk about it. A fair summary for a colleague or a patient: in a large randomized trial of patients able to receive high volumes, high-volume online HDF reduced the risk of death compared with high-flux HD, and pooled trial data point the same way. It is not accurate to say that HDF benefits every patient, that any HDF session is as good as a high-volume one, or that HD has become unsafe. The decision to offer HDF belongs to the physician and the patient. Your role is to make sure that, where HDF is chosen, the dose actually delivered resembles the dose that was tested.

The benefit was shown at high volume. A session that misses the volume is not the treatment that was tested.

Six HDF studies compared: CONTRAST, Turkish, ESHOL, CONVINCE, and two pooled analyses, with patients, comparator, and result; volume link shown as association.
The trials that delivered lower convection volumes found no overall survival difference; CONVINCE randomized only patients able to receive high volumes and found lower mortality. The link between higher achieved volume and better survival is an association, not proof.

Running an HDF session: before, during, and after

An HDF session follows the same sequence as an HD session (Module 9), with additions at each stage. The machine’s operator’s manual and the unit’s written procedure govern the exact steps, which differ between manufacturers.

Before the session

During the session

Alarm or problemLikely meaningFirst nursing checkEscalate when
Transmembrane pressure highToo much filtration for the blood flow; hemoconcentration; fibers clottingConfirm the actual blood flow (a very negative arterial pressure means the pump is not filling); look for dark streaks or clots; check lines for kinks; reduce substitution rate as the protocol allowsIt recurs after correction, clotting is visible, or the protocol does not cover the next step
Substitution or infusion faultProblem with the substitution line, pump segment, connection, or online fluid supplyCheck the line for kinks, clamps, and secure connection; check for air in the line; follow the machine’s promptsThe fault persists, air is found, or contamination of the connection is suspected
Blood leakBlood detected in the dialysate side, usually a fiber rupture; at high transmembrane pressures, ruptures are more likelyStop as the machine directs; inspect the dialysate outflow for pink or red colorAlways: a confirmed leak means the dialyzer is changed per protocol, and blood is not returned unless the protocol allows it
ConductivityDialysate composition wrong; both dialysate and substitution fluid are affectedCheck concentrate containers, connections, and that they match the order; the machine diverts fluid until correctedIt does not correct, or the patient has received fluid of uncertain composition
Filtration fraction limitConvective flow too high relative to blood flowCheck that blood flow is at the ordered level and the access is delivering it; reduce substitution rate as the protocol allowsThe ordered convection target cannot be reached at a safe filtration fraction; report for prescription review
Ultrafilter integrity test failedThe barrier that makes substitution fluid safe may be breachedDo not start or continue HDF on that machineAlways: report to the technician and charge nurse; HD on another machine only by order

As with every alarm in this course, find the cause before you reset the alarm. Never widen limits to keep a session running.

Anticoagulation. High filtration concentrates the blood inside the fibers toward the dialyzer exit. So post-dilution HDF circuits tend to clot more readily than HD circuits at the same anticoagulant dose. The anticoagulant and its dose remain the physician’s order. Report repeated clotting, streaked dialyzers, or sessions shortened by clotting. The remedy may be a different anticoagulation plan, dilution mode, or substitution setting, and each of those is a prescribing decision.

Ending the session. Follow the machine’s procedure for ending treatment and returning blood. Some machines return blood with online-prepared fluid instead of bagged saline. Either way, the return volume counts in the fluid balance. Disconnect the substitution line as the procedure directs and dispose of single-use parts. Never reuse a substitution line.

Documentation. Besides the usual HD record, document for every HDF session: the dilution mode; substitution volume; net UF volume; achieved convection volume; whether the ordered target was reached; the effective blood flow and effective treatment time; and, if the target was missed, the reason (access, catheter, clotting, alarms, shortened session, patient request). Record HDF-specific alarms and the action taken. A missed target is not a failure to hide. It is information the physician and the unit audit need.

Treat a rigor during HDF as contaminated fluid until proven otherwise. Stop substitution first, then find out why.

Algorithm for six HDF alarms: transmembrane pressure, substitution fault, blood leak, conductivity, filtration fraction limit, failed ultrafilter test, with first checks and escalation.
Find the cause before resetting any alarm, and never widen a limit to keep the session running. A blood leak or a failed ultrafilter test always goes to the charge nurse and the physician or technician.

Introducing and auditing HDF in a dialysis unit

The benefit in the trials came from high volumes delivered session after session. A unit that offers HDF without measuring what it delivers cannot know whether its patients are getting the treatment that was tested. Auditing HDF uses the quality improvement methods from Module 12 and the water and dialysate controls from Module 8.

Before starting an HDF program, a unit typically confirms that its water system and machines can consistently meet the ultrapure dialysate standard, with baseline culture and endotoxin results. It writes procedures for setup, monitoring, alarms, ending, logs, and adverse events. It trains and signs off staff on each machine model. It agrees which patients will be offered HDF and how the choice is explained. And it sets the indicators it will review.

AreaWhat to monitorQuestions the review asks
Delivered doseAchieved convection volume per session; proportion of sessions reaching each patient’s ordered targetWhich patients repeatedly miss the target, and why: blood flow, catheter, time, clotting, alarms?
Treatment timeEffective minutes delivered versus minutes prescribedAre late starts or early sign-offs costing volume?
Water and dialysateCulture and endotoxin results by sampling point, including HDF machines; trend toward action levelsIs any result rising, even if still below the limit? Was the response documented?
Ultrafilters and disinfectionChange log against the manufacturer’s interval; integrity test failures; missed disinfectionsWas any machine used for HDF outside its filter interval or after a missed disinfection?
Safety eventsPyrogenic reactions, bloodstream infections, blood leaks, balancing faults, circuit clottingIs there a cluster by machine, shift, or day? Was each event reported and investigated?
Staff competencyTraining records and periodic sign-off for each nurse and technicianIs everyone who runs HDF currently signed off on that machine model?
Patient experienceTolerance, symptoms, and patient-reported concernsAre patients informed and comfortable with the treatment and its costs?

Incident reporting. Report through the unit’s system any HDF session run with an overdue ultrafilter, a failed or bypassed integrity test, a missed disinfection, a contaminated substitution connection, a chill or fever during treatment, a fluid-balance error, or a water or dialysate result above the action level. Handle a cluster of febrile reactions under the cluster rule from Module 8: stop connecting new patients to the suspected source, assess those affected, and notify the physician, technician, and medical director.

Patient education. Patients often expect a new treatment to feel different. Explain plainly:

Where HDF is offered, it is a choice, not an upgrade every patient must take. A patient who stays on well-delivered high-flux HD, because of access limits, cost, or preference, has not been given second-rate care, and should never be made to feel so.

What a unit does not measure, it cannot claim to deliver. Audit the volume, not just the setting.

Therapeutic plasma exchange: removing what no dialyzer can

Everything in this module so far has used a membrane that keeps proteins in the blood. Continuous renal replacement therapy (CRRT) and hemodiafiltration (HDF) remove water, small solutes, and middle molecules. They are built to hold back albumin and the large proteins of the immune system. Some diseases are driven by exactly those large proteins: an antibody that attacks nerve, muscle, or kidney; an abnormal protein that thickens the blood; or a missing protein that the blood needs. For these, a different idea is used. Therapeutic plasma exchange removes the patient’s plasma itself, with everything dissolved in it, and replaces it with a fluid that does not contain the harmful substance.

The plain model. Picture a pot of soup with something spoiled dissolved in it. Straining the soup keeps the vegetables (the blood cells) but leaves the spoiled taste. Plasma exchange instead ladles out broth, returns the vegetables, and tops up the pot with clean broth.

The picture has a limit. The clean broth mixes with what remains, so each ladle removes a little less of the spoiled substance than the one before. A session removes most of it, not all. And unlike a pot, the body keeps adding to it. Substances that also sit in the tissues move back into the blood between sessions, and the immune system keeps making more antibody. That is why plasma exchange is usually a course of several sessions, combined with medicines that stop the body making the harmful substance. Plasma exchange buys time. The other treatments act on the source.

What it removes. Antibodies and immune complexes, abnormal proteins made by some blood cancers, some lipoproteins, and some toxins bound to proteins. It also removes useful things: clotting factors, some medicines, and, if the replacement fluid lacks them, normal antibodies.

How the plasma is separated, in principle

FeatureCentrifugal separationMembrane separation
How it worksBlood is spun; the heavier cells settle outward and the lighter plasma is drawn off by densityBlood passes through a plasma filter whose pores let proteins through but hold back blood cells
MachineA dedicated apheresis machine, often run by a blood bank or hematology serviceA plasma filter on a machine able to run it, which in some hospitals is a CRRT or dialysis platform
Blood flow and accessCan often work with lower blood flows, sometimes through large peripheral veinsUsually needs higher blood flow, so a central double-lumen catheter is common
Usual anticoagulantCitrate, mixed into the blood as it leaves the patientHeparin or citrate, by order

Neither method has been shown to be better for patients in general. The choice depends on the equipment and service available, the access, and the patient. The current international guideline recognizes both.17

When it is used: the guideline categories. The American Society for Apheresis publishes evidence-based fact sheets for each disease in which apheresis has been studied. The ninth edition, published in 2023, contains 91 fact sheets and 166 graded indications.17 Each indication sits in one of four categories. Learn them, because they explain why plasma exchange is urgent in one patient and debated in another:

CategoryMeaning, in our wordsExamples for plasma exchange in the 2023 edition
IAccepted as first-line treatment, alone or with other treatmentsThrombotic thrombocytopenic purpura; acute Guillain-Barré syndrome; short-term treatment of a myasthenia gravis crisis; anti-glomerular basement membrane disease with lung bleeding or while the kidneys still work; symptomatic hyperviscosity from very high antibody levels
IIAccepted as second-line treatment, alone or with other treatmentsLong-term treatment of myasthenia gravis
IIIBest role not established; the decision is made for the individual patientMicroscopic polyangiitis and granulomatosis with polyangiitis; anti-glomerular basement membrane disease in a patient already dialysis-dependent without lung bleeding
IVEvidence shows or suggests that apheresis is ineffective or harmfulThrombotic microangiopathy caused by the drugs gemcitabine or quinine

Source for the table: the 2023 guideline.17 A category describes how established the indication is. It is the guideline’s own scale, not a KDIGO-style grade, and it changes between editions as new trials appear. The decision to treat, the number of sessions, and the volume exchanged belong to the physician and the apheresis service.

Replacement fluid, in principle. The removed plasma must be replaced, or the patient would lose blood volume and protein. The two main choices are:

Some prescriptions combine the two. The fluid type is part of the order. Check it like a blood product or an intravenous (IV) medicine, because that is what it is.

Plasma exchange removes the harmful protein and the useful ones alike. Know which replacement fluid is running and what it does not contain.

Centrifugal and membrane plasma separation compared: a spinning chamber layers cells outward and draws plasma off, and a plasma filter lets plasma through its pores; removed plasma goes to waste and albumin or donor plasma is returned with the cells.
A dialyzer keeps proteins in the blood; plasma exchange removes the plasma itself, with the antibodies and other large proteins dissolved in it, and replaces it. Neither separation method has been shown to be better for patients in general.

Nursing care during plasma exchange

Plasma exchange happens in an intensive care unit (ICU), a ward, or an apheresis or dialysis unit, depending on how sick the patient is and which service runs the machine. You may be the apheresis nurse running the machine, or the ward or ICU nurse caring for the patient alongside. The same safety rules apply either way: verify the order, watch the patient more than the machine, recognize the predictable complications early, and report per order and protocol.

Before the session

During the session: what to watch for and what to do

ProblemWhy it happensWhat the nurse seesFirst actions
Citrate-related low calciumCitrate, from the anticoagulant and from donor plasma, binds ionized calcium in the patient’s bloodTingling around the mouth or in the fingertips, a strange taste, chills, nausea; later muscle twitching, cramps, tetany, low BP, or rhythm changesTell the operator to slow the return or the citrate rate per protocol; give calcium per order or protocol; notify the physician; check ionized calcium as ordered. Severe signs are an emergency
Low BPBlood in the circuit, slow replacement, fluid shifts, a vasovagal reaction, or an allergic reactionDizziness, pallor, sweating, yawning, a falling BPLay the patient flat, slow or pause per protocol, check that replacement keeps pace with removal, give fluid per order, and notify the physician
Allergic reaction to plasmaReaction to donor proteinsItching, hives, flushing; rarely wheezing, swelling of the lips or throat, and collapseStop the plasma, keep the line open per protocol, assess airway and breathing, notify the physician, and treat per order; anaphylaxis is managed as a medical emergency
BleedingAlbumin replacement removes clotting factors; heparin may be used in the circuitOozing at the catheter or puncture sites, bruising, bleeding gums, blood in urineApply pressure, report, and check that ordered clotting tests are done; procedures such as catheter removal are timed with the physician
OtherCool returning blood; low potassium after albumin; catheter infectionShivering; weakness or rhythm change; fever or redness at the exit siteUse the blood warmer if provided; report; follow catheter-care protocol

Take vital signs at the interval the protocol sets, and more often at the start and during any reaction. Ask the patient to tell you at once if tingling, chills, or itching start. Early citrate symptoms are felt before they are seen, so a patient who has been taught them is your first alarm.

After the session. Record the volume of plasma removed, the replacement fluid by type and amount, the net fluid balance, any reaction and its treatment, and the final vital signs. Check that post-session samples are drawn as ordered. Give any medicines scheduled for after the session, and reassess the findings you recorded at baseline. Hand over the time of the next session, the pending results, and any reaction, using the unit’s structured handoff.

Your place in the team. The physician decides the indication, volume, fluid, number of sessions, and medicine timing. The apheresis service runs the procedure to that order. You watch the patient, recognize complications, and keep the medicine schedule and the record accurate. Prompt reporting and coordination between these roles keep a usually well-tolerated procedure from becoming dangerous.

Tingling lips during plasma exchange mean low calcium until proven otherwise. Slow the citrate, replace calcium per protocol, and call.

Plasma exchange watch algorithm: checks before the session, then early and late signs of citrate-related low calcium, low blood pressure, allergic reaction to plasma, and bleeding, each with first actions and reporting, and the record after the session.
Early citrate symptoms are things the patient feels, so a patient taught to say "tingling" at once is the first alarm. The nurse watches the patient more than the machine and acts per order and protocol; severe signs are an emergency.

Case: Tingling lips at two in the morning

Mrs. Teresita Reyes, 58, a market vendor from Pampanga, is on day two in the ICU with septic shock and AKI after leptospirosis. She is on continuous venovenous hemodiafiltration with regional citrate anticoagulation and a calcium infusion, both run by the unit protocol. Her noradrenaline dose has come down since yesterday.

It is 2 a.m. You are her nurse. She tells you her lips and fingertips are tingling. The machine shows no alarm. Over the last 3 hours her hourly balance has run 150 mL more negative than prescribed. The calcium infusion bag ran out 40 minutes ago. Nobody replaced it, because the unit was busy with an admission.

Before reading on, decide what you would assess, what you would communicate and to whom, what you would teach, and what would make you escalate. Then compare with the outline below.

Assess

Tingling around the mouth and fingers, with an empty calcium bag, points to low ionized calcium from citrate. Check her blood pressure, heart rhythm, and QT interval on the monitor. Look for twitching or spasm. Send the ionized calcium sample the protocol requires. Then review the hourly balance and confirm what was actually infused and removed.

Communicate

Call the physician now with SBAR: patient on citrate CRRT, calcium infusion interrupted for 40 minutes, new perioral and finger tingling, current vital signs and rhythm, ionized calcium sent; request orders for calcium replacement and the citrate and fluid settings. Report the balance running more negative than prescribed.

Teach

Tell Mrs. Reyes, calmly, that the machine uses a medicine that lowers calcium in the tubing, and that her calcium drip is being restarted. Ask her to tell you at once about tingling, cramps, or a racing heart.

Escalate

Act immediately per protocol if she develops spasm, seizure, arrhythmia, a long QT interval, or hypotension. Report the interrupted infusion as an incident so that bag changes for protocol-driven infusions are never deferred again.
Documentation task. Write the 2 a.m. progress note: the symptom, what you found (infusion status, vital signs, rhythm, fluid balance), whom you called and when, the orders received, the time the calcium infusion was restarted, the result of the ionized calcium, and the patient’s response.

Check your understanding

Each visit draws 5 questions at random from this module’s 17-question bank, in a shuffled order. Choose the single best answer for each, then select Check answers; a rationale appears under every option. Reload the page or select New set of questions for a different mix. Nothing you select is sent anywhere; scoring happens in your browser.

Question 1 of 17A patient on standard high-flux HD for many years develops joint pain and carpal tunnel symptoms attributed to the build-up of beta-2 microglobulin. Why would adding large-volume convection be expected to remove more of this molecule?
Question 2 of 17During post-dilution continuous hemofiltration, the filter has clotted three times in 24 hours. Blood flow is limited by the catheter, and the filtration fraction is near the unit’s ceiling. Which change, if ordered, most directly addresses the cause?
Question 3 of 17Blood flow is 200 mL/min, the hematocrit is 30%, and the ultrafiltration rate in post-dilution is 35 mL/min. What is the filtration fraction?
Question 4 of 17You are about to hang a new replacement-fluid bag for a patient on continuous therapy. The order specifies a solution containing potassium; the bag on the shelf in the bay is potassium-free and the night is busy. What is the safest action?
Question 5 of 17Which patient is the most typical candidate for continuous therapy rather than standard intermittent dialysis?
Question 6 of 17A patient in the intensive care unit has severe fluid overload from heart failure that has not responded to diuretics. Potassium, bicarbonate, and urea are within acceptable ranges. Which continuous mode is designed for exactly this situation?
Question 7 of 17An 80 kg patient on continuous venovenous hemodiafiltration has dialysate at 1,200 mL per hour, post-filter replacement at 600 mL per hour, and net removal of 200 mL per hour. The circuit ran for 18 of the last 24 hours because of two clotted filters. What best describes the delivered dose?
Question 8 of 17A patient with liver failure and septic shock has been on continuous therapy with regional citrate for 30 hours. The calcium infusion has been increased three times to keep the patient's ionized calcium in range, total calcium is rising, and the metabolic acidosis is worsening. What is the most likely explanation?
Question 9 of 17The order asks for a patient total balance of 100 mL negative each hour. In the last hour, the continuous machine removed exactly the 150 mL net it was set to remove. Intravenous infusions and enteral feed totaled 230 mL, and urine was 30 mL. What is the patient's balance for the hour?
Question 10 of 17Over 4 hours, the transmembrane pressure and the pressure drop across the filter of a patient on continuous therapy with heparin have risen steadily. Access and return pressures are stable, and the patient is comfortable. What is the most appropriate nursing response?
Question 11 of 17A patient on post-dilution HDF has an ordered convection volume target of at least 23 liters per session. At the end of today's session the machine shows a substitution volume of 19.8 liters and a net ultrafiltration volume of 2.4 liters. What should the nurse conclude and do?
Question 12 of 17A patient starting online HDF asks, worried, how fluid made by the machine from tap water can be safe to run straight into her bloodstream. Which explanation is accurate?
Question 13 of 17Two hours into a post-dilution HDF session, transmembrane pressure has risen steadily, a filtration fraction alarm sounds, and the blood in the venous line looks darker than at the start. The arterial pressure is more negative than earlier. Blood pressure is stable and the patient feels well. What is the best first action?
Question 14 of 17During a unit meeting, a colleague says: "CONVINCE proved that HDF lowers deaths for every dialysis patient, so any HDF session, whatever the volume, is better than hemodialysis." Which response best reflects the evidence?
Question 15 of 17At the start of a shift, a nurse reviewing the machine log sees that the substitution-fluid ultrafilter on machine 4 is past the manufacturer's change interval. The machine has just passed its routine self-test, and a patient is scheduled for HDF on it in 15 minutes. What should the nurse do?
Question 16 of 17A patient with severe acute Guillain-Barré syndrome is starting plasma exchange. His wife asks why he cannot simply have dialysis, which she has seen used for other patients in the ICU. Which explanation is accurate?
Question 17 of 17Thirty minutes into a centrifugal plasma exchange using citrate and plasma as replacement fluid, the patient says her lips and fingertips are tingling and she feels a little nauseated. Her BP is unchanged. What should the nurse do first?
Reflect. Think of the last convective treatment you saw or ran. Which of the three fluid numbers (prescribed removal, machine removal, total patient balance) did the team actually reconcile, and who noticed when they disagreed?

What this module builds

Milestone. After this module, the nurse can assist safely with continuous renal replacement therapy (CRRT) or HDF under supervision: verifying fluids against the order, reconciling hourly fluid balance, answering alarms, and recording the delivered dose.

Each competency names an observable ability, the level this module aims for, and how that level is confirmed. Levels follow Miller’s pyramid. Self-study reaches knows; the higher levels are reached only when an educator or preceptor verifies them, and they are recorded on the competency checklist.

CompetencyThreadTarget levelVerified by
Explains convection, the replacement-fluid site, and the filtration fraction for the running treatment, and predicts how the dilution mode affects clearance and filter life.Delivering kidney replacement therapyDoes, supervisedObserved practice
Compares the continuous modes, SLED, and intermittent HD by mechanism and use, and identifies which patients and settings suit continuous therapy and which suit HDF.Delivering kidney replacement therapyKnowsSelf-assessment, Nursing case
Calculates prescribed and delivered CRRT effluent dose and HDF convection volume, and explains how downtime, blood flow, and treatment time reduce the delivered dose.Interpreting results and trendsKnowsSelf-assessment, Nursing case
Verifies each replacement-fluid or dialysate bag against the order with a second person where required, treating replacement fluid as an intravenous medicine.Medicines and their safetyDoes, supervisedObserved practice
Performs a CRRT bag change using the machine’s bag-change function and resolves a balance or scale alarm in a skills-station drill.Delivering kidney replacement therapyShows howSkills station, Simulation
Reconciles machine net removal, patient total balance, and the prescribed goal each hour, and reports a discrepancy or signs that removal is not tolerated.Clinical assessment and fluid statusDoes, supervisedObserved practice, Record review
Responds to a pressure or balance alarm by assessing the patient and circuit before the machine, never overriding air or blood-leak detectors, and documents the cause.Recognizing deterioration and emergency responseDoes, supervisedObserved practice
Monitors citrate or heparin anticoagulation as ordered, labeling patient and post-filter ionized calcium samples by site, and reports results outside the protocol range.Medicines and their safetyDoes, supervisedObserved practice
Recognizes low ionized calcium, citrate accumulation, and citrate excess in a simulation, checks that the calcium infusion is running, and escalates using SBAR.Recognizing deterioration and emergency responseShows howSimulation
Records the delivered dose as effluent dose or achieved convection volume, with downtime or missed-target reasons, and hands over using SBAR.Communication, handoff, and documentationDoes, supervisedObserved practice, Record review
Recognizes CRRT complications such as hypothermia, electrolyte loss, bleeding, and drug removal, and describes how to interrupt, step down, or stop continuous therapy safely.Clinical assessment and fluid statusKnowsSelf-assessment, Nursing case
Explains how online HDF makes substitution fluid from ultrapure dialysate, states its access, machine, water, and dialysate requirements, and describes how a unit audits achieved convection volume.Water, dialysate, and equipment safetyKnowsSelf-assessment, Nursing case
Distinguishes HDF from high-flux and expanded HD and summarizes the randomized evidence, including CONVINCE, using calibrated language about what remains uncertain.Patient and family teaching, nutrition, and shared decisionsKnowsSelf-assessment, Nursing case
Responds in simulation to HDF-specific alarms, including high transmembrane pressure, a failed ultrafilter integrity test, a blood leak, and a rigor during substitution.Recognizing deterioration and emergency responseShows howSimulation, Skills station
Explains what plasma exchange removes, how centrifugal and membrane separation differ, what the four guideline categories mean, and why albumin or plasma is chosen as replacement fluid.Delivering kidney replacement therapyKnowsSelf-assessment, Nursing case
Verifies replacement fluid against the order, checking plasma units with a second person by the transfusion procedure, and confirms the timing of medicines that plasma exchange removes with the physician or pharmacist.Medicines and their safetyDoes, supervisedObserved practice
Recognizes a citrate reaction, low blood pressure, or an allergic reaction to plasma in a simulation, takes the first actions per protocol, and escalates to the physician.Recognizing deterioration and emergency responseShows howSimulation

With your preceptor

With a preceptor experienced in continuous therapy or hemodiafiltration, observe and then assist with one treatment from setup to end. These are prompts for supervised learning, not a procedure. Your preceptor observes and signs off each item on the unit’s own competency form; the educator page has a printable version.

Read next on this site

Glossary & abbreviationsTalahulugan at mga daglatTalaan sa mga pulong ug daglatTalatinigan ampo reng daglat terms used on this page

Abbreviations

AKI
Acute kidney injury.
CFU
Colony-forming unit — The count of viable bacteria in a water sample.
CRRT
Continuous renal replacement therapy.
CVC
Central venous catheter.
CVVH
Continuous venovenous hemofiltration — Continuous therapy that clears solutes by convection, with replacement fluid.
CVVHD
Continuous venovenous hemodialysis — Continuous therapy that clears solutes by diffusion into dialysate.
CVVHDF
Continuous venovenous hemodiafiltration — Continuous therapy that combines diffusion and convection.
DOH
Department of Health — Philippines.
HD
Hemodialysis.
HDF
Hemodiafiltration — Hemodialysis combined with large-volume convection and online replacement fluid.
ICU
Intensive care unit.
ISO
International Organization for Standardization.
IV
Intravenous.
KDIGO
Kidney Disease: Improving Global Outcomes — The international nephrology guideline body.
KRT
Kidney replacement therapy — Dialysis or transplantation.
PIRRT
Prolonged intermittent renal replacement therapy.
PSN
Philippine Society of Nephrology.
RO
Reverse osmosis — The core water-purification stage of a dialysis water system.
SBAR
Situation, Background, Assessment, Recommendation — A structured handoff format.
SCUF
Slow continuous ultrafiltration — Continuous fluid removal without replacement fluid or dialysate.
SLED
Sustained low-efficiency dialysis — A prolonged intermittent therapy run over many hours with a standard machine.
UF
Ultrafiltration — Removal of plasma water across the dialyzer membrane by a pressure gradient.

Terms

Convection
Removal of solutes carried along with water pushed through a membrane by pressure.
Sieving coefficient
How freely a substance crosses a membrane with water: 1 means as freely as water, 0 means not at all.
Replacement fluid
Sterile solution infused into the blood path to replace the plasma water removed by convection.
Pre-dilution
Giving replacement fluid before the filter, which thins the blood and lengthens filter life.
Post-dilution
Giving replacement fluid after the filter, which clears more per liter but concentrates blood in the filter.
Filtration fraction
The share of plasma flow through the filter that is removed as ultrafiltrate.
Middle molecules
Solutes larger than urea, such as beta-2 microglobulin, that diffusion removes poorly.
Effluent
All fluid removed by the effluent pump: spent dialysate, filtered replacement fluid, and the net fluid removed from the patient. In continuous therapy its volume approximates small-solute clearance.
Prescribed dose
The effluent rate the order sets, expressed per kilogram of body weight per hour.
Delivered dose
The dose the patient actually received, after subtracting time the circuit was not running.
Downtime
Time during a continuous therapy day when the blood pump is stopped, for example for bag changes, clotting, or procedures.
Net fluid removal
The fluid the machine removes from the patient each hour beyond what it returns as dialysate and replacement fluid.
Patient total balance
All fluid in minus all fluid out, including machine net removal, for a set period.
Post-filter ionized calcium
Ionized calcium measured in blood sampled after the filter; used to judge circuit anticoagulation with citrate.
Citrate accumulation
Build-up of citrate in a patient who cannot metabolize it, shown by a rising total-to-ionized calcium ratio, a rising calcium requirement, and worsening acidosis.
Transmembrane pressure
The pressure difference across the filter membrane; a steady rise indicates clogging of the membrane.
Pressure drop
The difference between the pressure entering and leaving the filter on the blood side; a rise indicates clotting in the blood channels.
Recirculation (circuit)
Connecting the ends of a disconnected circuit so that blood or prime keeps moving during a short interruption, where the protocol and manufacturer permit.
Prolonged intermittent renal replacement therapy
Kidney replacement run on a standard machine for about 6 to 12 hours at reduced flows; sustained low-efficiency dialysis is one form.
Online hemodiafiltration
HDF in which the machine prepares substitution fluid continuously from ultrapure dialysate during the session.
Convection volume
Total plasma water filtered across the membrane in a session: substitution volume plus net ultrafiltration.
High-volume HDF
HDF delivering a large convection volume per session; at least 23 liters was the entry requirement in CONVINCE.
Ultrapure dialysate
Dialysate with a total viable count below 0.1 CFU per mL and endotoxin below 0.03 endotoxin units per mL (ISO 23500-5:2024).
Ultrafilter (endotoxin-retentive filter)
A filter on the dialysate path fine enough to hold back bacteria and endotoxin; used in series to make substitution fluid.
Post-dilution
Infusing substitution fluid into the blood line after the dialyzer; the most efficient mode, limited by hemoconcentration.
Medium cut-off membrane
A dialyzer membrane with larger effective pores used in expanded HD on a standard machine, without substitution fluid.
Beta-2 microglobulin
A middle molecule of about 11,800 daltons used as a marker of convective removal; its accumulation is linked to dialysis-related amyloidosis.
Auto-substitution
A machine feature that adjusts the substitution rate automatically from pressure and other measurements, within unit-set limits.
Therapeutic plasma exchange
Removal of a patient's plasma, with the substances dissolved in it, and its replacement with albumin solution or donor plasma.
Centrifugal separation
Separating plasma from blood cells by spinning the blood, so that components separate by density.
Membrane separation
Separating plasma from blood cells with a plasma filter whose pores pass proteins but hold back cells.
Apheresis category
The American Society for Apheresis rating (I to IV) of how well established apheresis is for a given disease indication.
Citrate reaction
Symptoms of low ionized calcium caused by citrate binding calcium during apheresis, starting with tingling around the mouth and fingers.
ReferencesMga SanggunianMga TinubdanReng Reperensya 18 sources
  1. Kidney Disease: Improving Global Outcomes (KDIGO) Acute Kidney Injury Work Group. (2012). KDIGO clinical practice guideline for acute kidney injury. Kidney International Supplements, 2(1), 1–138. https://doi.org/10.1038/kisup.2012.1
  2. Ostermann, M., Bellomo, R., Burdmann, E. A., Doi, K., Endre, Z. H., Goldstein, S. L., Kane-Gill, S. L., Liu, K. D., Prowle, J. R., Shaw, A. D., Srisawat, N., Cheung, M., Jadoul, M., Winkelmayer, W. C., Kellum, J. A., & Conference Participants. (2020). Controversies in acute kidney injury: conclusions from a Kidney Disease: Improving Global Outcomes (KDIGO) Conference. Kidney International, 98(2), 294–309. https://doi.org/10.1016/j.kint.2020.04.020
  3. International Organization for Standardization (ISO). (2024). Preparation and quality management of fluids for haemodialysis and related therapies — Part 3: Water for haemodialysis and related therapies (ISO 23500-3:2024). https://www.iso.org/standard/84370.html
  4. RENAL Replacement Therapy Study Investigators. (2009). Intensity of continuous renal-replacement therapy in critically ill patients. The New England Journal of Medicine, 361(17), 1627–1638. https://doi.org/10.1056/NEJMoa0902413
  5. Palevsky, P. M., Zhang, J. H., O'Connor, T. Z., Chertow, G. M., Crowley, S. T., Choudhury, D., Finkel, K., Kellum, J. A., Paganini, E., Schein, R. M., Smith, M. W., Swanson, K. M., Thompson, B. T., Vijayan, A., Watnick, S., Star, R. A., & Peduzzi, P. (2008). Intensity of renal support in critically ill patients with acute kidney injury. The New England Journal of Medicine, 359(1), 7–20. https://doi.org/10.1056/NEJMoa0802639
  6. Zarbock, A., Küllmar, M., Kindgen-Milles, D., Wempe, C., Gerss, J., Brandenburger, T., Dimski, T., Tyczynski, B., Jahn, M., Mülling, N., Mehrländer, M., Rosenberger, P., Marx, G., Simon, T. P., Jaschinski, U., Deetjen, P., Putensen, C., Schewe, J. C., Kluge, S., ... RICH Investigators and the Sepnet Trial Group. (2020). Effect of regional citrate anticoagulation vs systemic heparin anticoagulation during continuous kidney replacement therapy on dialysis filter life span and mortality among critically ill patients with acute kidney injury: A randomized clinical trial. JAMA, 324(16), 1629–1639. https://doi.org/10.1001/jama.2020.18618
  7. Bouchard, J., Soroko, S. B., Chertow, G. M., Himmelfarb, J., Ikizler, T. A., Paganini, E. P., Mehta, R. L., & Program to Improve Care in Acute Renal Disease (PICARD) Study Group. (2009). Fluid accumulation, survival and recovery of kidney function in critically ill patients with acute kidney injury. Kidney International, 76(4), 422–427. https://doi.org/10.1038/ki.2009.159
  8. Müller, M., Jürgens, J., Redaèlli, M., Klingberg, K., Hautz, W. E., & Stock, S. (2018). Impact of the communication and patient hand-off tool SBAR on patient safety: a systematic review. BMJ Open, 8(8), e022202. https://doi.org/10.1136/bmjopen-2018-022202
  9. Blankestijn, P. J., Vernooij, R. W. M., Hockham, C., Strippoli, G. F. M., Canaud, B., Hegbrant, J., Barth, C., Covic, A., Cromm, K., Cucui, A., Davenport, A., Rose, M., Török, M., Woodward, M., Bots, M. L., & CONVINCE Scientific Committee Investigators. (2023). Effect of Hemodiafiltration or Hemodialysis on Mortality in Kidney Failure. The New England Journal of Medicine, 389(8), 700–709. https://doi.org/10.1056/NEJMoa2304820
  10. Peters, S. A., Bots, M. L., Canaud, B., Davenport, A., Grooteman, M. P., Kircelli, F., Locatelli, F., Maduell, F., Morena, M., Nubé, M. J., Ok, E., Torres, F., Woodward, M., Blankestijn, P. J., & HDF Pooling Project Investigators. (2016). Haemodiafiltration and mortality in end-stage kidney disease patients: a pooled individual participant data analysis from four randomized controlled trials. Nephrology Dialysis Transplantation, 31(6), 978–984. https://doi.org/10.1093/ndt/gfv349
  11. International Organization for Standardization (ISO). (2024). Preparation and quality management of fluids for haemodialysis and related therapies — Part 5: Quality of dialysis fluid for haemodialysis and related therapies (ISO 23500-5:2024). https://www.iso.org/standard/84372.html
  12. Grooteman, M. P., van den Dorpel, M. A., Bots, M. L., Penne, E. L., van der Weerd, N. C., Mazairac, A. H., den Hoedt, C. H., van der Tweel, I., Lévesque, R., Nubé, M. J., ter Wee, P. M., Blankestijn, P. J., & CONTRAST Investigators. (2012). Effect of online hemodiafiltration on all-cause mortality and cardiovascular outcomes. Journal of the American Society of Nephrology, 23(6), 1087–1096. https://doi.org/10.1681/ASN.2011121140
  13. Ok, E., Asci, G., Toz, H., Ok, E. S., Kircelli, F., Yilmaz, M., Hur, E., Demirci, M. S., Demirci, C., Duman, S., Basci, A., Adam, S. M., Isik, I. O., Zengin, M., Suleymanlar, G., Yilmaz, M. E., Ozkahya, M., & Turkish Online Haemodiafiltration Study. (2013). Mortality and cardiovascular events in online haemodiafiltration (OL-HDF) compared with high-flux dialysis: results from the Turkish OL-HDF Study. Nephrology Dialysis Transplantation, 28(1), 192–202. https://doi.org/10.1093/ndt/gfs407
  14. Maduell, F., Moreso, F., Pons, M., Ramos, R., Mora-Macià, J., Carreras, J., Soler, J., Torres, F., Campistol, J. M., Martinez-Castelao, A., & ESHOL Study Group. (2013). High-efficiency postdilution online hemodiafiltration reduces all-cause mortality in hemodialysis patients. Journal of the American Society of Nephrology, 24(3), 487–497. https://doi.org/10.1681/ASN.2012080875
  15. Vernooij, R. W. M., Hockham, C., Strippoli, G., Green, S., Hegbrant, J., Davenport, A., Barth, C., Canaud, B., Woodward, M., Blankestijn, P. J., Bots, M. L., CONVINCE Scientific Committee, & HDF Pooling Project Investigators. (2024). Haemodiafiltration versus haemodialysis for kidney failure: an individual patient data meta-analysis of randomised controlled trials. The Lancet, 404(10464), 1742–1749. https://doi.org/10.1016/S0140-6736(24)01859-2
  16. Rose, M., Fischer, F. H., Liegl, G., Strippoli, G. F. M., Hockham, C., Vernooij, R. W. M., Barth, C., Canaud, B., Covic, A., Cromm, K., Cucui, A. M., Davenport, A., Fischer, K. I., Hegbrant, J., Jaha, H., Schappert, A., Török, M., Woodward, M., Bots, M. L., ... CONVINCE Scientific Committee and CONVINCE Investigators. (2024). The CONVINCE randomized trial found positive effects on quality of life for patients with chronic kidney disease treated with hemodiafiltration. Kidney International, 106(5), 961–971. https://doi.org/10.1016/j.kint.2024.07.014
  17. Connelly-Smith, L., Alquist, C. R., Aqui, N. A., Hofmann, J. C., Klingel, R., Onwuemene, O. A., Patriquin, C. J., Pham, H. P., Sanchez, A. P., Schneiderman, J., Witt, V., Zantek, N. D., & Dunbar, N. M. (2023). Guidelines on the use of therapeutic apheresis in clinical practice: Evidence-based approach from the Writing Committee of the American Society for Apheresis: The ninth special issue. Journal of Clinical Apheresis, 38(2), 77–278. https://doi.org/10.1002/jca.22043
  18. Counts, C. S. (Ed.). (2020). Core curriculum for nephrology nursing (7th ed.). American Nephrology Nurses Association. https://old.annanurse.org/core-curriculum-nephrology-nursing